General; Training Of Customer Support Engineers; Document Required; Components Included - Siemens MOBILETT Plus HP Service Manual

Table of Contents

Advertisement

Prerequisites

General

These instructions describe how to replace the inverter with part no. 65 08 654 on the
MOBILETT Plus HP.
The inverter D13 with part no. 65 08 654 is no longer available. It has been replaced by
65 28 306. When replacing the old type with the new one, the existing software has to be
replaced by V0.8H (part no. 65 86 114).
It is also necessary to equip the MOBILETT Plus HP with two extra capacitors. To make
room for the capacitors, the battery boxes and battery shelves must be replaced too.

Training of customer support engineers

Due to the technology used in this equipment, setup, service and maintenance should be
performed only by customer support engineers who have work authorization for
MOBILETT Plus.

Document required

Service Instructions for MOBILETT Plus HP.

Components included

The kit (part no. 65 53 981) contains the following parts:
Software kit: V0.8H
Inverter
Capacitor unit: M11
Battery box
Upper battery shelf
Lower battery shelf
Cable tie
Screw
Label
Repair Instruction (this instruction)

Tools required

Standard tool kit.

Time required

Replacement of inverter, belonging software, battery boxes and battery shelves, and
installation of capacitor unit: approximately 2 1/2 hours for 1 person.
Siemens-Elema AB
Solna, Sweden
Article
SPR8-220.091.01
Rev. 03
Quantity
Part no.
1
65 86 114
1
65 28 306
1
65 53 767
4
65 53 700
1
65 53 692
1
65 53 684
15
65 01 712
2
62 29 178
1
65 53 676
1
66 53 930
Page 1 of 2
03.03
SPS-UD
1 - 1
1
1
1
1
1
1
1
MOBILETT Plus HP
Service

Advertisement

Table of Contents
loading

Table of Contents