0200-NAWI-08815 0,05 kg Geeicht für: Zone 3 S/N 643043 Eichzählerstand: 3 B O S C H E GmbH Co.KG Reselager Rieden 3 49401 Damme www.bosche.eu TYPE PLATE- EXAMPLE WITHOUT CALIBRATION DATA Typ: S/N: Rating: 1000 mA BOSCHE Wägetechnik Reselager Rieden 3 49401 Damme 05491-999689-0 www.bosche.eu...
The manufacturer's contact data is listed on the reverse of the title page. If you have any queries or problems, please contact the manufacturer without delay. Note If you have any questions for Bosche GmbH & Co. KG, please have the serial number to hand. Version 4.0...
3.2 Packaging and Disposal................15 3.3 Installation and Connection................16 3.4 Device Variants ..................... 17 3.4.1 MWI-N / EWI-N in Stainless Steel Housing ......... 17 3.4.2 MWI-T / EWI-T in Aluminium Housing ..........17 3.5 System Configuration..................18 3.6 Interface ......................19 3.6.1 RS485 ....................
Page 7
Weighing Terminal EWI / MWI Foreword 4 Operation ......................25 4.1 Turn On ......................25 4.2 Error Message....................26 4.3 Power Zero Limits ..................26 4.4 Turn Off ......................27 4.5 Auto Off ......................27 4.6 Reseting the Scale to Zero................27 4.7 Zero Tracking ....................
Weighing Terminal EWI / MWI Safety | 1 AFETY This chapter warns against possible risks when handling the device. The information for detection of risks contained in this chapter is intended to allow the save and correct operation. It is important to read and adhere to this operating manual and particularly this chapter prior to operating this device.
Important notes and information about an efficient, economic and environ- mental friendly handling are specified here. NTENDED The Weighing Terminal EWI / MWI exclusively serves to display the weight in combination with suitable load cells. The load cells are intended for use as "non-automatic weighing instruments".
The BOSCHE company offers a restricted warranty for components, which became faulty due to strain or material faults. The warranty starts with the date of delivery. The BOSCHE company retains the right to repair or replace components. Repair work executed during the warranty period will not extend the period of warranty.
Page 12
1 | Safety Weighing Terminal EWI / MWI • Use of an incorrect or non-licensed analogue / digital converter • Non-observance of the specifications in the operating manual • Conversion, modification or opening of the device • Unintentional or mechanical damage and damage caused by media, liquids, natural wear.
ESCRIPTION ENERAL The weighing terminal EWI / MWI is a multifunction scale display and is used to display weighing values obtained in combination with suitable load cells. Up to 4 load cells can be connected. After reaching a stable weight value, the weight value are shown on the display.
2 | Description Weighing Terminal EWI / MWI 2.2.1 ISPLAY VERVIEW Status indication: the weight on Hand tare weighing (Pretara) the scale above the selected 8. 8 . 8 . 8 . 8 . limit value. Status indication: the weight on TARE 8.
2 | Description Weighing Terminal EWI / MWI 2.2.3 EYBOARD VERVIEW Operating level Functional level Opens the "Quick menu" for Opens the function menu when quick settings. Switch button bet- the button is pressed during the ween multiple scales, if connec- self-test of the scale.
Description | 2 Value inputs Note The numeric keypad is not available in the EWI version. Numeric keys for enter parameters (tare weights, etc.) Decimal point / decimal separator Clear input value. Is used as a clear key when setting values for parameters or other functions.
Page 18
2 | Description Weighing Terminal EWI / MWI Version 4.0 BOSCHE Weighing Systems Page 14 2022-11-14...
Weighing Terminal EWI / MWI Transport, Installation, Connection | 3 RANSPORT NSTALLATION ONNECTION ONTROL When the display is delivered, check the packaging, the device and possible accessories for visible damages. ACKAGING AND ISPOSAL Keep all parts of the original packaging for a possible return.
3 | Transport, Installation, Connection Weighing Terminal EWI / MWI NSTALLATION AND ONNECTION Keep the device clean and do not expose to an environment influencing the display accuracy. Protect against draught! Protect against heat, Protect against tilting a sun and frost!
Weighing Terminal EWI / MWI Transport, Installation, Connection | 3 EVICE ARIANTS 3.4.1 MWI-N / EWI-N TAINLESS TEEL OUSING 232 mm 81 mm 84 mm 232 mm 110 mm 3.4.2 MWI-T / EWI-T LUMINIUM OUSING 45 mm 231 mm Version 4.0...
3 | Transport, Installation, Connection Weighing Terminal EWI / MWI YSTEM ONFIGURATION In weighing systems, one or more load cells are connected for evaluating the measuring signal. Several load cells of a scale are connected in parallel in a junction box to deliver a common output signal.
NTERFACE 3.6.1 RS485 The weighing terminals of the EWI / MWI series offer different options to connect load cells / weighing platforms (weighing bridges) to the transmitter. Basically: • Has the weighing terminal on the back one or more 5-pin plugs, these are used to connect 4-wire or 6 wire load cells.
Page 25
In case of damage to the weighing terminal, the verification seal is broken and a replacement device can be connected. Recalibration is then mandatory. For further questions please contact the Bosche customer service. 5-pin plug connection for interface RS485 Pin assignment: 1.
3 | Transport, Installation, Connection Weighing Terminal EWI / MWI 3.6.2 RS232 The weighing terminal is connected via this interface (COM 1) to a PC or a printer, it prints the weighing results along with the selected weighing unit. 5-pin plug connection for interface RS232 Pin assignment: 1.
Weighing Terminal EWI / MWI Transport, Installation, Connection | 3 3.6.3 LAN- I NTERFACE The LAN interface is used to exchange and query data in a network. Pin assignment: 1. NC 2. NC 3. NC 4. RX- 5. TX+ 6. RX+ 7.
Page 28
3 | Transport, Installation, Connection Weighing Terminal EWI / MWI Version 4.0 BOSCHE Weighing Systems Page 24 2022-11-14...
Weighing Terminal EWI / MWI Operation | 4 PERATION Note The basic functions of the weighing terminal are similar or identical used for all applications. The signal light shown in some pictures can be integrated as an option. The standard version of this weighing terminal does not have a signal light.
4 | Operation Weighing Terminal EWI / MWI RROR ESSAGE If no digitizing unit is recognized after switching on and the display self-test, the following message appears in the display: Please check if a scale is connected. For further error messages see chapter 5 “Troubleshooting“.
Weighing Terminal EWI / MWI Operation | 4 If you are in the menu, the scale is turned off without that any changes made are stored. In addition, a voltage tester is connected to indicate a low battery voltage by the battery symbol.
4 | Operation Weighing Terminal EWI / MWI Then the scale can be set by Tare button to "zero" and continues to work in the net mode. RACKING The zero point correction automatically corrects minor deviations of the Zero indicator. The zero range must be switched on in the password protected menu The zero tracking: •...
Weighing Terminal EWI / MWI Operation | 4 EIGHING Display Weighing Unstab Dat & Stab a tive T ar 4.9.1 IMPLE WEIGHING After switching on, the self-test and initialization of the scale, "zero" is shown. 30 kg If the weight of the weighing sample will exceed the set weighing range the fol- lowing message is displayed: >...
4 | Operation Weighing Terminal EWI / MWI 4.9.2 EIGHING IN ULTI ANGE ULTI NTERVAL To use the weighing terminal as multi-range or multi-interval scale, the appro- priate settings must be made in the password protected menu. Attention Changes in the password-protected menu of a calibrated scale lead to the expiry of the verification.
Weighing Terminal EWI / MWI Operation | 4 4.9.3 EIGHING WITH EFERENCE CALE When a reference scale is connected it can be changed between the connec- ted scales by pressing this button.. The active scale is shown at the bottom of the display.
4 | Operation Weighing Terminal EWI / MWI 4.10 EIGHING 4.10.1 ETERMINATION OF ARE BY EIGHING The weight of each weighing container can be removed by pressing , so that the net weight is displayed. Tare (with stable weight): 1 kg...
Page 37
Weighing Terminal EWI / MWI Operation | 4 Second tare (e.g. when weighing in several components): 0,6 kg 1 kg 0,6 kg 1 kg Clear Tare Version 4.0 BOSCHE Weighing Systems Page 33 2022-11-14...
Weighing Terminal EWI / MWI Operation | 4 4.11 EMORY The display has several settings to add weights. Weights are added up manually or automatically. The stored weights are removed by turning off the scale or manually by deleting from memory.
4 | Operation Weighing Terminal EWI / MWI 4.11.1 ANUAL UMMATION 11 kg oder 11 kg 5 kg max. max. 9999,99 kg Version 4.0 BOSCHE Weighing Systems Page 36 2022-11-14...
Page 41
Weighing Terminal EWI / MWI Operation | 4 Possible summation errors (with unstable weight): 11 kg „Memory Error“ 11 kg 5 kg > 9999,99 kg Version 4.0 BOSCHE Weighing Systems Page 37 2022-11-14...
Weighing Terminal EWI / MWI Operation | 4 4.12 UICK VERVIEW 2 sec HILoGo MEMory ALIbI Fvnc dISPL Back to weighing mode Version 4.0 BOSCHE Weighing Systems Page 39 2022-11-14...
Page 44
4 | Operation Weighing Terminal EWI / MWI Quick Menu - Short description Note The Quick menu is used for simplified input or reading out of parameters and parameterization of the display. In the Quick Menu parameters to operate at...
Page 45
Weighing Terminal EWI / MWI Operation | 4 Menü Untermenü Optionen Funktion nUmBeR a04-00 on off Consecutive ticket number (for INfo totals weighing in Pr2 or STC mode). Set the print parameters ALIbi on/off If the alibi memory is switched...
Page 46
4 | Operation Weighing Terminal EWI / MWI Menü Untermenü Optionen Funktion bL AU a06-00 Automatic activation / switch-off dI5PL LI9HT of the backlight after N seconds. 10/20/30/40/50 --> Set the The backlight is automatically Backlight value with <UP> activated when the weight changes or a key is pressed.
Weighing Terminal EWI / MWI Operation | 4 4.12.1 A01 - O UICK VERVIEW IMIT ALUES Menu navigation HILoGo HiLoGo INFO 2 sec A01-00 beep A01-01 STorE A01-03 LIGHTS Version 4.0 BOSCHE Weighing Systems Page 43 2022-11-14...
Page 48
4 | Operation Weighing Terminal EWI / MWI Quick menu A01-01 - BEEP (Set acoustic signal) A01-00 beep A01-00 8E oK 8E Lo 8e HI 8eHILo DonE Version 4.0 BOSCHE Weighing Systems Page 44 2022-11-14...
Page 49
Weighing Terminal EWI / MWI Operation | 4 Quick menu A01-02 - STORE (Storage of limit values)) A01-01 STorE A01-01 Version 4.0 BOSCHE Weighing Systems Page 45 2022-11-14...
Page 50
4 | Operation Weighing Terminal EWI / MWI Quick menu A01-03 - LIGHTS (Settings for the brightness) A01-03 LIGHtS A01-03 DonE Version 4.0 BOSCHE Weighing Systems Page 46 2022-11-14...
4 | Operation Weighing Terminal EWI / MWI 4.12.3 A04 - P UICK RINT ETTINGS Note If parameters such as the date and time have been turned off on the menu, this can not be printed, even then, if the parameters for date and time have been set to ON.
Weighing Terminal EWI / MWI Operation | 4 4.12.4 A06 - D UICK ISPL ISPLAY D1sPL A04-01 InFo A04-01 A04-00 CoPY nUMbEr ALLbI dATE TIMe Gross MEtto TArE t.Gross A04-01 t.MEtto T.TArE nUMbEr Enter value 1 - 4 DonE or off DonE Version 4.0...
4 | Operation Weighing Terminal EWI / MWI 4.13 HECK EIGHING With the check weighing (only MWI) you can specifically check weights. At the check-weighing a beep sounds when the weight on the scale reaches or exceeds a stored limit.
Page 55
Weighing Terminal EWI / MWI Operation | 4 Example: Set the limit values >HI 60 kg <LO 5 kg <HI >LO 40 kg Version 4.0 BOSCHE Weighing Systems Page 51 2022-11-14...
Weighing Terminal EWI / MWI Operation | 4 4.14 RINTOUT Before a printer can be put into operation, the appropriate parameter settings in the function menu (see separate Technical Manual) and in the quick menu (see chapter 4.12.3 “Quick Menu A04 - Pri (Print Settings)“) must be made.
4 | Operation Weighing Terminal EWI / MWI 4.14.2 „ C“ RINTOUT WITH ELECTED ARAMETER When a weight is placed and it is stable, the indicator sends the following infor- mation to the printer. In addition, the data are stored in the sum memory of the...
Weighing Terminal EWI / MWI Operation | 4 4.14.3 „PR1“ RINTOUT WITH ELECTED ARAMETER When a weight is placed and it is stable, the indicator sends the following infor- mation to the printer if the PRINT button is pressed. Standard...
4 | Operation Weighing Terminal EWI / MWI 4.14.4 „PR2“ USDRUCK MIT GEWÄHLTEM ARAMETER When a weight is placed and it is stable, the indicator sends the following infor- mation to the printer if the PRINT button is pressed. In addition, the data are stored in the sum memory of the weighing terminal.
Weighing Terminal EWI / MWI Troubleshooting | 5 T ROUBLESHOOTING N THE EVENT OF A MALFUNCTION We recommend switching off the scales briefly and disconnecting it from the mains, if a malfunction occurs in the program sequence. Then restart the weighing process.
Weighing Terminal EWI / MWI Troubleshooting | 5 NFORMATION FOR THE CONTACT TO THE CUSTOMER SERVICE Owner/user Specifications Name of your company Name of a contact person Contact data Telephone Email Tabelle 5.1: Your company Product Specifications Model name Serial number...
Page 64
5 | Troubleshooting Weighing Terminal EWI / MWI Version 4.0 BOSCHE Weighing Systems Page 60 2022-11-14...
Weighing Terminal EWI / MWI Technical Data | 6 ECHNICAL EIGHING ERMINAL Feature Value / Unit Dimensions (B x H x T) Stainless steel housing • Table unit 232 x 154 x 90 mm • Wall unit 232 x 154 x 110 mm...
6 | Technical Data Weighing Terminal EWI / MWI COPE OF ELIVERY Component Note Weighing terminal Mains adapter Operating manual Table/wall holder Not included for the version with stand Version 4.0 BOSCHE Weighing Systems Page 62 2022-11-14...