General Information Description The Maury model A034B1 connector gage kit is designed to measure 3.5mm and 2.92mm (K) connectors with either female or male interfaces. CAUTION: Use your A034B1 for 3.5mm and 2.92mm (K) connectors only. Do not use for SMA connectors.
Page 4
- Readings indicate Recession of Center Conductor + Readings indicate Protrusion of Center Conductor Grasp the Gage here for Maximum Stability Male Setting Surface Female Setting Surface Master Gage Figure 1. The A034B1 Connector Gage Kit Operating Instructions September 2022...
U.S.A. National Institute of Standards and Technology (NIST). To verify that your gage kit is performing to traceable specifications, periodically send the kit to Maury Microwave Corporation for calibration. The recommended calibration cycle is one year. The actual need may vary depending on usage.
MAURY MICROWAVE 034-92 A Operation Visual Inspection Inspect all connectors carefully before each use. If a connector shows deep scratches, dents, uneven wear, or par- ticles clinging to the mating plane surfaces, clean it and inspect again. Damaged connectors should be set aside for repair.
Page 8
MAURY MICROWAVE 034-92 A Align your Connector Gage and the Master Gage Zero Set the Connector Gage using the Master Gage. If Dial Hand Does NOT Read Zero, Reset as Shown Below. Adjust Outside Bezel for Proper Zero Reading. Dial Lock Figure 3.
MAURY MICROWAVE 034-92 A Gaging 3.5mm and 2.92mm (K) Connectors The critical mechanical specification in precision 3.5mm and 2.92mm (K) connectors is the recession (setback) of the center conductor relative to the outer conductor mating plane. No protrusion of the center conductor in front of the outer conductor mating plane is allowable.
Page 10
MAURY MICROWAVE 034-92 A Zero Set the Connector Gage using the Master Gage. Align your Connector Gage and the Connector. 0.0015 in. Gently Connect the Device. 0.0 in. Figure 4. Example of Gaging the 3.5mm and 2.92mm (K) Connectors Operating Instructions...
Page 11
Corporation for calibration. The recommended calibration cycle is one year. The actual need may vary depending on usage. Should Additional Information or Service be Required, Address Inquiries To: Maury Microwave Corporation Attention: Customer Service 2900 Inland Empire Boulevard Ontario, California 91764-4804...