Page 2
The qualified service personnel who get Kaixin written authorization can repair the instrument out of warranty by themselves. But this should be agreed by Xuzhou Kaixin Electronic Instrument Co., Ltd. We will provide circuit diagrams, component part lists or other information to assist service personnel to repair those parts of our equipment that are designated by our company as repairable by service personnel.
Page 3
This user’s manual contains warnings for predictable dangers. Users shall also exercise care at any time to be aware of the dangers unforeseen in this manual. Kaixin shall not be liable for the damages and losses arising out of neglecting to follow the operation instructions herein described.
Page 4
User’s Manual V1.08 Bladder Volume Tester Safety Cautions 1. Warning Symbols and Definitions The following warning symbols are used in this manual to indicate safety level and other important items. Please remember these symbols and understand the meaning as you read this user’s manual. These symbols convey specific meanings as detailed in the table below: Symbols &...
Page 5
User’s Manual V1.08 Bladder Volume Tester 3. Labels Label Description Attention: When using the battery should pay attention to precautions. Symbol for the marking of electrical and electronics devices according to Directive 2012/19/EU. The device, accessories and the packaging have to be disposed of waste correctly at the end of the usage.
User’s Manual V1.08 Bladder Volume Tester Contents Chapter One Overview……………………………………………………………………………1 Chapter Two Technical Specifications……………………………………………………………2 Chapter Three System Outline……………………………………………………………………3 3.1 Structure composition of the instrument……………………………………………………3 3.2 Components name……………………………………………………………………………3 3.3 Parts of the probe………………………………………………………………………..3 3.4 Function keys description…………………………………………………………………..3 Chapter Four System Configuration……………………………………………………………5 Chapter Five Operation Condition………………………………………………………………5 System Installation and Check…………………………………….……………6...
Page 7
User’s Manual V1.08 Bladder Volume Tester 12.2.1 System cleaning and disinfection…………………………………………………………22 12.2.2 Clean the probe socket……………………………………………………………………24 12.2.3 Clean the touch screen……………………………………………………………………24 12.2.4 Clean the foot switch……………………………………………………………………24 12.2.5 Clean the shell and probe bracket………………………………………………………24 12.2.6 Clean the thermal printer…………………………………………………………………24 12.3 Replacement of accessories and consumables……………………………………………24 12.3.1 Replace the thermal paper………………………………………………………………24 12.3.2 Replace the battery………………………………………………………………………25 12.3.3 Replace the probe, foot switch, power adapter and power supply cord, fuse…………25...
Overview 1.1 Introduction BVT01 bladder volume tester is composed of main unit, probe, power adapter, etc., which is used to non-invasively measure bladder volume with ultrasound principle. It is used to assess urinary retention and urinary incontinence, and given the timing of implement an objective clinical catheterization to reduce the catheterizing frequency and reduce the risk of urinary tract infections.
Bladder Volume Tester Chapter Three System Outline 3.1 Structure composition of the instrument BVT01 bladder volume tester is composed of main unit, probe, power adapter, etc. 3.2 Components name Fig. BVT01 sketch map 3.3 Parts of the probe Fig. Parts name of 2.5MHz probe...
Page 11
User’s Manual V1.08 Bladder Volume Tester Read data Touch the key to open data file; Save data Touch the key to save data file; Print Print various measurement images/data; Image In main interface, touch the keys to select the selection keys desired left or right image;...
User’s Manual V1.08 Bladder Volume Tester Chapter Four System Configuration 4.1 Typical configuration 1. Main unit 1 unit 2. 2.5MHz probe 1 pc 3. Power adapter 1 pc 4. Internal battery 1 pc 5. Touch pen 2 pcs 4.2 Optional parts 1.
Warning: 1. When instrument works abnormally, do stop working, turn off the power and check the reason, then contacts the Kaixin Company about it. 2. Turn off power and pull out of the plug from socket after each operation. 3. It is forbidden to drag and press the power and probe cables emphatically; regularly inspect whether there is spilt and bareness, if there is the phenomena like this;...
User’s Manual V1.08 Bladder Volume Tester 6.1 System placement Please carefully read through and fully understand the safety cautions before moving and placing the system. 1. Unpack the instrument case and check the goods for its completeness according to the packing list furnished.
User’s Manual V1.08 Bladder Volume Tester 6.3.1 Connecting the probe Warning: Before connecting or using the probe, make sure that the probe, connecting cable and connector are in normal condition (free of cracks or drop). Use of defective probe may cause electric shock. The red mark of connector should be aligned with the red mark on the socket Attention: when inserting the probe.
EQUIPMENT the place where it is difficult to operate the disconnection device. 2. AC/DC adapter is as a part of the equipment, please only use the AC/DC adapter provided by Kaixin Company. 3. To avoid damaging power adapter or harming people by unexpected fallen, make sure the power adapter is placed on the leveled desk.
User’s Manual V1.08 Bladder Volume Tester Warning: Only use the coupling gel that in accordance with MDD regulations. If the overheat acoustic lens is placed on the patient’s skin, heat injury may occur. Attention: Attention : Thoroughly clean the coupling gel on the probe surface each time after ultrasonic operation, or the coupling gel may become hardened on the acoustic lens of the probe, deteriorating quality of image.
User’s Manual V1.08 Bladder Volume Tester Chapter Seven Work Main Interface 7.1 Work main interface Press the button on the front panel, turn on the machine, the touch screen displays the work main interface, as follows: Fig. Work main interface Work main interface shows the ultrasonic images, various data and the corresponding operation menus, etc.
User’s Manual V1.08 Bladder Volume Tester The image and data region is divided into left part and right part. The upper part of left displays patient type and image selection keys ( ). Patient types are divided into male, female, female with hysterectomy, obesity and child. Touch left or right image to enter manual contour interface, for individual complex images having large error, the doctor may choose manual contour function.
User’s Manual V1.08 Bladder Volume Tester Chapter Eight System Preset 』key, the system preset dialog box will be displayed on the screen, as below. In the Touch『 "About" interface, you can view the full version and release version of the software. System Basic Settings About Hospital:...
Page 21
User’s Manual V1.08 Bladder Volume Tester Enter 4. Touch key on the digital keyboard to complete the entry and the cursor automatically jumps to the next input position; Bspace 5. If needs revise, touch key to delete the previous character and retype; Bspace Enter Fig.
6. If you want to obtain the U disk encryption folder; please contact International Trade Dept of Kaixin. U disk encryption Copy the “KX” folder provided by our company to the root directory of the U disk, and the encryption is completed.
User’s Manual V1.08 Bladder Volume Tester Screen copied to U disk 1. In the boot state, insert U disk to main unit; 』key, the screen displays “Save Data” dialog box; 2. Touch『 3. Touch「U:\Save Img」key, pop up “Success” dialog box after finish saving, touch「OK」, the screen content will be saved to U disk;...
User’s Manual V1.08 Bladder Volume Tester 」to print out the stored data or pictures; (4) Touch「 」to exit “Open Data” interface. (5) Touch「 4. Touch 「Copy to U」 , pop up “Success” dialog box, touch 「OK」 , the selected file will be dumped to “DUMP”...
User’s Manual V1.08 Bladder Volume Tester Chapter Ten Bladder volume measurement 10.1 Scanning and positioning bladder The correct positioning of bladder is the basis of accurate measurement of bladder volume. Bladder is located in the lower abdomen, below the pubic symphysis. Before the examination, place ultrasound coupling gel on the patient's abdomen, place the probe in accordance with the position of probe shown in the figure below: Fig.
User’s Manual V1.08 Bladder Volume Tester Figure A shows: bladder projection obviously deviates from the centerline of the test, which is located in the lower right of the centerline of the probe. The measured data is not accurate, it need to adjust the angle and position of probe, and re-measurement.
User’s Manual V1.08 Bladder Volume Tester Attention To ensure the accurate measurement, please make sure: 1. There is no air gap between the probe and the patient’s skin when scanning, and use appropriate pressure to keep the contact with the patient’s skin; 2.
User’s Manual V1.08 Bladder Volume Tester Chapter Eleven Principle of Acoustic Power 11.1 Biological effect It is generally recognized that ultrasonic diagnosis is safe for human’s health. So far, there has been no report on bodily harm done by ultrasound. Nevertheless it is also believed that not all types of ultrasound are absolutely safe.
2. Clean the probe with sterile water to remove the residual chemicals after disinfection, because the residual chemicals may be harmful for humans. 3. Kaixin Company will not make any guarantee for the efficacy of disinfector. Please contact the appropriate manufacturer for details.
Page 30
User’s Manual V1.08 Bladder Volume Tester Attention: 1. In the process of cleaning and disinfection, avoid probe overheat (exceeding 55℃) as it may be deformed or damaged under excessive heat. 2. In the operation of disinfection, please refer to medical institutions disinfection technical specifications.
User’s Manual V1.08 Bladder Volume Tester Attention 1. It is a normal phenomenon that color of the acoustic lens may change and color of the probe label may fade away. The regular disinfection times should be minimized as it may lead to degrade of the probe safety and performance.
12.3.3 Replace the probe, foot switch, power adapter and power supply cord, fuse If the probe, foot switch, power adapter and power supply cord, fuse are damaged during the specified period, please contact the Kaixin Company. Replacement method: 1. Refer to Section 6.3 for the replacement of probe.
1. The fuse is inside the power adapter. Fuse shall be replaced by qualified service personnel who get Kaixin approval. 2. Before replacing the fuse, please contact Kaixin and replace it under the guidance of Kaixin. 3. Before replacing the fuse, you must disconnect the mains supply from the mains supply.
International Trade Dept of Kaixin for support. If the following problems occur on starting up the machine, try to make corrections following the method in the table. If the problem remains unsolved, contact International Trade Dept of Kaixin for support.
User’s Manual V1.08 Bladder Volume Tester instrument, please consult the International Trade Dept of Kaixin or its dealers, as they need to have professional technology engineers. 1. The following safety checks should be performed at least every 12 months by a qualified person who has adequate training, knowledge, and practical experience to perform these tests.
User’s Manual V1.08 Bladder Volume Tester Chapter Thirteen Storage and Transportation 1. If the instrument is stored over 3 months, take out the instrument from the packing case, connect it to power supply for 4 hours, and then disconnect the power and place it in the case again following the direction indicated by arrows on the package.
User’s Manual V1.08 Bladder Volume Tester Chapter Fifteen Safety Classification 1. Classified according to the type of protection against electric shock: Class I EQUIPMENT + internally power EQUIPMENT 2. Classified according to the degree of protection against electric shock: Type B applied part 3.
User’s Manual V1.08 Bladder Volume Tester Guidance and Manufacturer’s Declaration Chapter Sixteen This product complies with EMC test standard IEC 60601-1-2. Warning: The use of inappropriate accessory will reduce the performance of the product. Attention: 1. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.
Page 39
User’s Manual V1.08 Bladder Volume Tester Table 1: Compliance level for electromagnetic emission Emission test Compliance RF emissions Group 1 CISPR 11 RF emissions Class A CISPR 11 Harmonic emissions Class A IEC 61000-3-2 Voltage fluctuations/flicker emissions Complies IEC 61000-3-3 Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A).
Page 40
User’s Manual V1.08 Bladder Volume Tester 70%U ) (30%dip in U for 25 (50Hz) cycles 0%U (>95%dip in U for 300 (60Hz) cycles Power frequency (50/60 Hz) magnetic field 30(A/m) IEC 61000-4-8 Note 1: The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz;...
In the acoustic output measurement data, the MI uncertainty is 12%, TI uncertainty is 23%. Manufacturer: Xuzhou Kaixin Electronic Instrument Co., Ltd. Product Name: Bladder Volume Tester Table 201.103 Transducer Model: 2.5S120M1 Model: B...
Page 43
KAIXIN ELECTRONIC XUZHOU KAIXIN ELECTRONIC INSTRUMENT CO., LTD. Kaixin Mansion, C-01, Economic Development Zone, Xuzhou, Jiangsu, China Zip Code: 221004 Tel: +86-516-87732932/87733758 Fax: +86-516-87732932/87792848 Website: http://www.kxele.com E-mail: info@kxele.com Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany Information contained in this manual is subject to change without further notice.