Summary of Contents for ActivePure F170A

  • Page 1 MODEL F170A READ MANUAL CAREFULLY FOR PROPER USE AND OPTIMAL OPERATION. The Aberdeen Bldg. 14841 Dallas Parkway, Suite 500 Dallas, TX 75254...
  • Page 2: Important Safety & Use Instructions

    • Use only as described in this manual • Do not attempt to repair or adjust any electrical or mechanical functions of the ActivePure® Medical Guardian. Contact Customer Service for service support at 1.800.572.0317 • This device will automatically turn off when the door is open •...
  • Page 3: Table Of Contents

    BEFORE YOU BEGIN USING THE ACTIVEPURE® MEDICAL GUARDIAN First, read this user manual in its entirety to ensure that the ActivePure® Medical Guardian device is used as intended. Then plug it in, turn it on to the appropriate fan setting. The ActivePure® Medical Guardian is designed for quiet operation, as not to disturb people in a professional healthcare environment.
  • Page 4: Device Description

    10-watt UVGI bulbs operating at a wavelength of 254nm; two pieces of polycarbonate honeycomb material with a combined surface area of 708 square inches, coated with a TiO2 based catalyst called ActivePure®; power switch; circuit board which controls the device operations and an internal fan to draw air through the device, as shown in the general device design (Figure 1).
  • Page 5: Product Features

    PRODUCT FEATURES It is important to be familiar with the components of the ActivePure® Medical Guardian. If any components are missing, contact Customer Service for support at 1.800.572.0317 or email at medical@activepure.com. Door Pull Back Handle Panel Airflow Input Front...
  • Page 6: Device Specifications

    • Near water, outdoors or in a damp or wet area Place the ActivePure® Medical Guardian in a location that allows air to move freely into, around and out of the unit. Allow minimum of 12" of clearance around intake grill (airflow input).
  • Page 7: Treatment Space

    • Rates of ventilation The device is intended to result in a 4-6 log reduction of airborne microorganisms from initial concentrations in 60 minutes as shown in Table 1. The ActivePure® Medical Guardian, model F170A is not a sterile device and is not intended to create a sterile environment.
  • Page 8: Settings

    Disconnect power before servicing. The ActivePure® Medical Guardian is designed to reduce airborne contaminants with minimal maintenance. Filter: The ActivePure® Medical Guardian HEPA Filter is replaced when the change filter light illuminates after 180 days (approx 6 months) of use.
  • Page 9: Cleaning/Disinfecting And Filter Replacement Instructions

    CLEANING/DISINFECTING AND FILTER REPLACEMENT INSTRUCTIONS The ActivePure® Medical Guardian, model F170A is a device intended for medical purposes that is used for the reduction of staphylococcus epidermidis and erwinia herbicola bacteria, MS2 and Phi-X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from the air in a temperature-controlled professional healthcare environment of 70~71°F, 40~45%RH.
  • Page 10 CLEANING/DISINFECTING AND FILTER REPLACEMENT INSTRUCTIONS (CONT.) Step 10: Remove a Metrex CaviWipe from its package. Beginning by cleaning the surface “D”. Wipe the CaviWipe over the surface area, thoroughly wetting the soil. Wait 1 minute, then using hand pressure rub the wipe over the surface until soil is no longer visible on this part of the device.
  • Page 11: Activepure® Cell Replacement Instructions

    Step 8: Slide the new cell back inside the device and secure with the 4 screws removed in Step 3. Step 9: Plug device in and reset the ActivePure Cell Change Light by pressing and holding the power button for 10 seconds until it goes out.
  • Page 12: Replacement Parts

    10 seconds After you replace the filter or ActivePure Cell you will want to reset the indicator light(s). To do so simply press and hold the power button for 10 seconds. This will reset the filter and/or cell module replacement indicators.
  • Page 13: Warranty Information

    HEPA filters with genuine filters every 180 days (6 months) of FOR U.S. APPLICATION ONLY use, and replacing the ActivePure® Cell with a genuine ActivePure® Cell every This warranty gives you specific legal rights, and you may also have other 365 days (1 year) of use.
  • Page 14: Device Iec 60601-1-2:2014 Emc Information

    DEVICE IEC 60601-1-2:2014 EMC INFORMATION This device passed each of the 11 tests, as indicated. Immunity test levels performed on the device, • Electrostatic Discharge Immunity – Tested at 2,4,8 and 15 kV • Radiated Immunity – Tested 3 V/m (Professional Healthcare) and 10 V/m (Home Healthcare) •...
  • Page 16 ActivePure® Medical, LLC | The Aberdeen Bldg. | 14841 Dallas Parkway, Suite 500 | Dallas, TX 75254 | 1.800.572.0317 ActivePure® Medical, LLC Canada, Inc. 3480 Laird Road, Suite 2-5 Mississauga, ON L5L 5Y4 © ActivePure® Medical, LLC. All Rights Reserved...