1.2 List of Safety Instructions Proper function of the FMS belt scale is only guaranteed with the recommended application of the components. In case of other arrangement, heavy malfunction can be the result. Therefore, the installation instructions on the following pages must be followed strictly.
Operating manual BMGZ 600A Table of contents Safety Instructions .................. 2 1.1 Description Conditions 1.2 List of Safety Instructions Definitions ....................5 System Components ................5 System Description ................. 7 4.1 Functional Description 4.2 Measuring Roller 4.3 Evaluation Electronic Unit 4.4 External Displays 4.5 Block Diagram Dimensions .....................
Page 4
Operating manual BMGZ 600A 10.1 Wiring of the PROFIBUS Data Cable 10.2 Setting the PROFIBUS Address PROFIBUS Interface Description ............41 11.1 GSD File 11.2 BMGZ610A/611A/620A DP Slave Functional Description 11.3 Initial Parameters 11.4 Configuration 11.5 Function Code Trouble Shooting ................... 43 12.1 Position of the Fuses List of Spare Parts ................
Operating manual BMGZ 600A 2 Definitions Measuring uncertainty: Each measuring device has some inaccuracy, but however, inaccuracy is in a small, defined range of tolerance. This deviation of the measuring value referring to the effective value is called measuring uncertainty. Measuring accuracy: Its behaviour is the contrary to the measuring uncertainty: Increases the uncertainty, decreases the accuracy, and also in reverse mode.
Page 6
Operating manual BMGZ 600A (components in italic letters are option)
Operating manual BMGZ 600A The versions BMGZ610A/611A provide evaluation of 1 measuring roller; version BMGZ620A provide evaluation of 2 measuring rollers. Versions BMGZ611A provide additional digital inputs and outputs which can be used for extended control functions such as proportioning belt scales, etc. All versions support an RS232 interface.
6 Installation and Wiring Caution Proper function of the FMS belt scale is only guaranteed with the recommended application of the components. In case of other arrangement, heavy malfunction can be the result. Therefore, the installation instructions on the following pages must be followed strictly.
Page 13
Operating manual BMGZ 600A fig. 7: The measuring roller has to be placed as far away from the material feed as the bulk material may settle before passing the measuring roller. B600001e fig. 8: To minimize influences of the belt tension, the measuring roller should be located as far away as possible from the drive roller.
Operating manual BMGZ 600A 6.2 Mounting the Measuring Roller fig. 10: The mounting devices are adjustable in many ways to fit best to the channels. B600018e The measuring rollers have a sliding part on each side of the support for easy and flexible adapting to the width of the channels (fig.
Page 15
Operating manual BMGZ 600A Aligning the height of the measuring roller fig. 12: The middle part of the measuring roller must be aligned with the neighbouring support rollers. B600004e After mounting the measuring roller on the channels, the height of the roller must be aligned to the neighbouring support rollers (fig.
Page 16
Operating manual BMGZ 600A Repeat the procedure for the 2 side disk.
Operating manual BMGZ 600A 6.3 Mounting the Electronic Unit The evaluation electronic unit is available in variants as follows: BMGZ6xxA.W Housing for wall mounting (aluminium; IP54) (fig. 6) BMGZ6xxA.S Housing for panel mounting (aluminium; Front IP54, Back IP00) ...
Page 18
Operating manual BMGZ 600A 8. Replace the earth cable and all the cables with connectors back to place.
Operating manual BMGZ 600A 6.5 Wiring of the Connection Cable Wiring between measuring roller and evaluation electronic unit must be done with the shipped 8-wired, shielded twisted-pair cable (4x2x0.75mm ). Length of the cable is done according to customer specification (specify when ordering). The cable must be installed separate from power lines to prevent any inductive disturbances.
Operating manual BMGZ 600A 6.7 Wiring of Main Supply The evaluation electronic unit is available for supply voltage of 24VDC or 230VAC. All electronic units have 6 terminals for wiring of main supply. But only the terminal pair corresponding to the nameplate is internally connected (refer to wiring table).
Operating manual BMGZ 600A Scaling of the signal (how many kg for each pulse) is parametrized with the parameters Pulse output 1 or. Pulse output 2. Analogue display (actual flow rate) According to the wiring table analogue signals which are proportional to the actual flow rate are provided.
Page 24
Operating manual BMGZ 600A The additional digital inputs and outputs can be used for some control functions. Fig. 21 shows a suggestion for a proportioning belt scale: fig. 21: Wiring diagram for proportioning belt scale (The power part is not shown). B611003e In the diagram shown in fig.
Operating manual BMGZ 600A 7 Operation 7.1 View of the Operating Panel fig. 22: The operating panel BMGZ600A B600008e 7.2 State Diagram of Main Operating Menu fig. 23: State diagram BMGZ610A/611A B600021e...
Operating manual BMGZ 600A fig. 24: State diagram BMGZ620A B600022e 7.3 Setting the Installation-specific Parameters For correct calculation of the flow rate, the following parameters must be set or checked during initial operation (refer to „8. Parametrization“): Belt length roller 1 ...
Note FMS recommends running the tare program daily to compensate eventual changes of belt tension, etc. If the belt scale begins to count slowly forwards or backwards with the belt running empty, it is time at the latest to run the tare program again.
Operating manual BMGZ 600A Run a charge of any weight over the belt scale, for example 10 tons, and then load it into a truck of known tare weight (refer to „7.6 Weighing a charge (manual)“). Check the load on the weighbridge as exact as possible and compare with the value shown in the belt scale display ...
Operating manual BMGZ 600A is reached (parameter Pre detect charge), the silo will be shut automatically. The conveyor belt is still running. If the charge is completely weighed (parameter Ref. charge 1), the conveyor belt stops automatically. With the „stop“ key, the silo can be shut and the conveyor belt can be stopped at any time.
Operating manual BMGZ 600A 8 Serial Interface RS232 A RS232 interface command consists of 6 ASCII characters. It is always terminated with <CR <CR ASCII character determines the > (e.g. XRV001 >). The first module number. A replay telegram consists of 7 ASCII characters including decimal point (XXXXXXX).
Operating manual BMGZ 600A 9.3 State Diagram Parameter Setting fig. 26: State diagram BMGZ610A/620A. The version BMGZ621A has its own but equivalent parameter menu. B600020e...
Operating manual BMGZ 600A 9.4 Description of the Parameters The parameter changing mode will be activated by pressing the key for 3 seconds. Generally, the parameters are settable using the keys as follows: choose switch the selections or increase / decrease numeric values ...
Page 34
5000 Default: Increment: Unit: Diameter Roller 1/2 Use: This parameter stores the diameter of the measuring roller 1 (refer to nameplate of measuring roller). FMS measuring rollers have a standard diameter of 108mm. Range: 10 to 1000 Default: Increment: Unit:...
Page 35
Operating manual BMGZ 600A Pulses roller 1/2 Use: This parameter stores the number of vanes of the pulse generator (refer to nameplate of the measuring roller). The value corresponds to the number of pulses for each roller rotation. Range: Default: Increment: Unit: Distance roller 1/2...
Page 36
Operating manual BMGZ 600A Pulse output roller 1/2 Use: The pulse output of the measuring roller shows a single pulse after a known feeding quantity. This parameter is used to determine how many kg correspond to one pulse. Range: 1, 10, 100, 1000 Default: Unit: [kg]...
Operating manual BMGZ 600A The value stored in this parameter should nearly correspond to the bulk material weight which is still on the conveyor belt after shutting the silo and which still has to pass the belt scale. Range: 0.00 100.0 Default: 0.00...
Page 38
Operating manual BMGZ 600A The service mode 1 contains parameter for checking the AD value of the force sensors which are integrated into the measuring roller. The service mode 1 is activated by pressing the CLR and PAR keys for 3 seconds. AD value roller 1/2 Use: This parameter shows the source signal of the corresponding...
Operating manual BMGZ 600A 10 PROFIBUS Interface 10.1 Wiring of the PROFIBUS Data Cable Wiring of the PROFIBUS cables The standardized PROFIBUS cable type A (STP 2x0.34 ) [AWG] has to be used for the PROFIBUS data cable. The cables are bared referring to fig. 6 and connected to the terminals according to the wiring diagramm.
Operating manual BMGZ 600A 10.2 Setting the PROFIBUS Address The electronic unit requires a unique PROFIBUS address which indicates it definitely in the whole PROFIBUS network. Therefore no other PROFIBUS device in the network may use the same address. The address has to be between 2...125. The PROFIBUS address is set with the system parameter Identifier.
The GSD file can also be supplied on a floppy disk on request. In this case please contact FMS customer service. Read in the GSD file into the PROFIBUS DP Master How to read in the GSD file into the control system (DP Master) is depending on the used control system.
Operating manual BMGZ 600A 11.4 Configuration The configuration defines how many process data (byte and word) are sent during the data communication from the control system to the measuring amplifier and vice versa. To ensure maximum flexibility different modules are provided in a single measuring amplifier.
No message on the Power supply not correct Check / correct power display supply; Check fuses in the supply line Evaluation electronic unit Contact FMS customer defect service 12.1 Position of the Fuses fuse 24VDC 1.0A T ; F100 fuse 230VAC 0.5A T...
Operating manual BMGZ 600A 13.2 Assembly Drawing Troughed Measuring Roller fig. 32: Troughed measuring roller. Rollers with Belt widths greater than 1000mm may have more than 1 pair of bevelled side disks. B400012e...
Operating manual BMGZ 600A 13.3 Part List Pos.# Description Pos. Description Middle roller Support Pan head screw M4x8 Sliding part Pan head screw M4x16 Mounting bracket Set screw M8x6 Cover Hex nut M12, 0.5xD Force measuring bearing Hex stop nut M10 LMGZ203 Cover A Washer M10...
Operating manual BMGZ 600A 14 Technical Data Evaluation electronic unit BMGZ610A BMGZ611A BMGZ620A Evaluation of 1 measuring roller Evaluation of 2 measuring rollers Min and Max limit relays Additional digital in-/ outputs (galvanically isolated) Belt speed monitoring Operation 6 keys, LCD display 2x16 characters Display possibilities total output conveyed [t] daily output / charge [t]...
Page 49
Overload protection 150% of nominal load Breaking load >1000% of nominal force Axial load allowed 20% of nominal load FMS Force Measuring Systems AG FMS Italy FMS USA, Inc. FMS UK Aspstrasse 6 Via Baranzate 67 2155 Stonington Ave. Suite...