Download Print this page

Anton APM 145 Operating Instructions page 2

Pressure meter

Advertisement

Test Menu – Diff Pressure
Used for differential pressure measurement where
both pressure spigots are used and the pressure
difference is shown on screen.
Test Menu – Working Pressure
Used to determine correct working pressure is set up
for appliances.
Data can be stored & shared.
Test Menu – Operating Pressure
Used to determine correct operating pressure is set up
for appliances.
Data can be stored & shared.
Test Menu – Flue Draught Pressure
Used to determine correct flue draught pressure.
Data can be stored & shared.
Units of Measurement
Select this menu option to change the pressure units
displayed on the meter.
Settings Menu – Auto off timeout
Changes the period when the unit will switch off if
unattended.
Settings Menu – Backlight
Changes the brightness of the backlight.
Lower settings will prolong battery life.
Settings Menu – Key Click
Enable/disable button click.
Settings Menu – Report
Change report sending option (printer, wireless or
toggle).
Settings Menu – Supervisor Settings
User configurable settings. Allows setting of time and
date, owner/company name and user password.
Stored Logs Menu
Search and retrieve stored logs. Allows log to be
reviewed, re-printed or re-transmitted over the
wireless link.
BATTERY REPLACEMENT
1. Remove the protective rubber boot by slipping it
away from the bottom of the case and sliding it over
the pressure spigots exposing the case back.
2. Remove the battery cover from the bottom of the
case by removing the four screws.
3. Remove the three AA batteries and replace with
good quality AA Alkaline batteries taking care to
install them the right way round.
4. Replace the battery cover carefully making sure it is
alligned properly and replace the screws. Take care
not to over tighten the screws otherwise it may
damage the case.
For best results and longer life always use quality
alkaline batteries MN1500 or equivalent.
CAUTIONS
Exposure to silicone based products, corrosive gases,
water / condensation and freezing should be avoided
at all times as this will cause damage to the sensor
and internal component corrosion.
This instrument should not be stored or operated
outside of its operating temperature range.
Particular care must be taken not to exceed the
maximum pressure ratings for the device as this
may rupture the sensor membrane. This is not
covered by the manufacturer's warranty.
NOTE: The differential/gauge instruments will only
measure positive pressures i.e. positive pressure
applied to positive port, or negative pressure applied
to the negative port.
WARRANTY
This instrument is guaranteed against defects of
workmanship and materials for a period of twelve
months from date of invoice.
During the warranty period a defective instrument will
be repaired or replaced at the discretion of the
manufacturer. This warranty does not cover damage
or failure due to misuse or accident. Modification,
adjustment or any alteration shall void the warranty.
For any warranty claims to be considered the
instrument must be returned to Anton along with proof
of purchase at the senders cost.
STANDARD ACCESSORIES
The APM 145 is supplied complete with:
Three AA Alkaline batteries
Operating instructions
Protective rubber boot with stand, hook and
magnet
SPECIFICATIONS
Power Supply
3 x AA Alkaline Batteries
Battery Life
> 20 hours*
Duty Cycle
Continuous.
Auto switch off after 5 mins.
Operating
-10 to +50 °C
Temperature
Weight
395 grams including
batteries
Dimensions
185 x 75 x 43 mm
Accuracy
±2% of reading
* Using recommended Industrial batteries
Units
APM 145
mbar
-39.99 to +19.99
+20.0 to +130.0
PSI
-0.29 to +1.89
mmHg
-15.00 to +19.99
+20.0 to +98.1
mmH
O
-199.9 to +199.9
2
+200 to +1336
inH
O
-8.03 to +52.63
2
hPa
-39.99 to +19.99
+20.0 to +130.0
Pa
-400 to 9999
kPa
-1.99 to 13.00

Advertisement

loading