Page 1
Contact:+8618106639615 Email:lluticoeu@outlook.com This manual is for biological microscope Model N125-DE. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope, it is recommended strongly that you study this manual thoroughly before operating the microscope.
Contents N125-DE Contents ......................User Notices ..................... 1. Components Name................... 2. Assembly ....................2.1 Assembly Diagram ................. 2.2 Assembly procedure ................3. Adjustment and Operation..............3.1 Adjustment set diagram ............... 3.2 Adjustment and Operation..............4. Technical Specifications................4.1 Main specifications .................
Page 3
Use Notices N125-DE Ⅰ. Safety note 1. Open the box carefully to avoid the accessories, like lens, dropping to ground or being damaged. 2. Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking environment.
Page 4
1. Components Name N125-DE Eyepiece Binocular viewing head Nosepiece Objective Condenser Stage Illumination Main body Biological Microscope N125-DE...
2.Assembly N125-DE 2.1 Assembly Diagram The following figure shows the installation sequence of the components. The number in the figure shows the assembly steps. ★ Before installing, be sure every components is clean, do not score any parts or glass surface.
N125-DE 2.2 Assembly procedure 2.2.1 Installing binocular viewing head(Fig.1-2) Insert the binocular viewing head into the microscope head and turn it to a proper position, then tight it. Fig.1 2.2.2 Installing the eyepiece (Fig3-4) Insert the eyepiece into the eyepiece tube until they are against each other, as shown in Fig.4.
Page 7
N125-DE 2.2.3 Installing the objective(Fig.5-6) 1. Adjusting the coarse focus knob until the support device of the mechanical stage reaches its low limit position. 2. Screw the lowest magnification objective into the nosepiece from the left or the right side, then revolve the nosepiece clockwise and mount other Fig.5...
Page 8
N125-DE 2.2.5 Connecting the Power Cord (Fig.8-10) ★ The cable and cords are vulnerable when bent or twisted, never subject the power cord to excessive force. ② 1. Set the main switch①o “O” (off ) state before ① connecting the power cord.
Page 9
N125-DE 2.2.7 Installing and Replacing the bulb (Fig.11-12) ◇ Please use the specified halogen Lamp 6V20W. 1. Loosing the screw①。 2. Hold the bulb②after you wrap it with gauze or other protection materials, and then insert its ① pindeeply intothe jack in the lamp holder.
N125-DE 3.2 Adjustment and Operation 3.2.1 Brightness Adjustment(Fig.15-16) 1. Connect the power cord and set the main switch① to “-”state (ON). 2. Turning the brightness adjustment knob ②. ① 3.2.2 Placing the Specimen (Fig.17) 1. Place the specimen the center of the mechanical Fig.15...
Page 13
N125-DE 3.2.4 Condenser Adjustment(Fig.19) Turn the condenser focus knob①to move the condenser up and down.Raise the condenser when using the high magnification objective, and descend ② it when using the low magnification one. ① 1. Focus the specimen with 10× objective.
Page 14
N125-DE 3.2.6 Adjusting the Interpupillary Distance(Fig.21) The interpupillary distance range: 48mm-75mm。 While looking through the eyepieces, move both eyepieces round until the left and right fields of view coincide completely. 3.2.7 Adjusting the Diopter (Fig.22) Fig.21 1. Turn the eyepiece①to adjust the diopter while looking through it.
6.Troubleshooting Guide N125-DE 6.1 Optical System TROUBLE CAUSE SOLUTION The nosepiece is not in the located Locate the nosepiece position (objective and light path not coaxial) properly where it clicks 1. The edge of the field of view is dark or the...
Page 18
N125-DE 6.2 Mechanical System TROUBLE CAUSE SOLUTION 1.Field of view of one eye does not Interpupillary distance is incorrect Adjust interpupillary distance match that of the other The diopter is not proper Adjust the diopter properly 2. Observation is tiring...
7. Warranty N125-DE 90-Day Money Back You May Return Your Product For A Full Refund Within The First 90 Days Of Your Purchase. 1-year parts warranty If you encounter any quality problem with this product within one year after using it, please contact us.