Mirus for H2 Classic GrainGage
H2 Classic GrainGage Calibration
12. Pour the grain sample into the hopper (to simulate
harvest). Ensure the isolation gate is open so the grain
drops into the GrainGage.
13. Enter the Known Moisture percentage.
14. Tap Click to run a sample.
15. Repeat steps 12–14 with the same sample at least two
more times.
Note: The Last Value button can be used to fill the
Known Moisture field with the previous percentage
when the same sample is run multiple times.
16. Tap the next arrow
17. Tap the check icon
Mirus displays the adjusted moisture curve with its
moisture percentages and voltages.
to complete the recalibration.
.