10.3 WHAT DO THE ERROR MESSAGES ON THE DISPLAY MEAN? ................. 28 11. DESCRIPTION OF THE CONTROLS ........................29 12. TECHNICAL DATA ............................. 31 13. SAMPLING SYSTEM ............................32 Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
The device records all measured values and changes to the device per channel in the background and has a USB interface via which all data can be read out as a logbook in CSV format. Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
( V4A ) or in PVDF, This means that almost all requirements regarding corrosion resistance can be met. Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 5
PTFE pipe is to be used. Other plastic materials may cause a strong diffusion of moisture into the gas. Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
This preparation phase is complete when the value displayed on the device does not change within about 30 minutes. Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
• Scaling of the display area • Alarm/ Limit-configuration • Setting of the automatic sensor test (per channel) • Analog output scaling (per channel) Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 8
2. Engineer - PIN - change for device configuration; Read data; Read device status 3. Service - protected interface - change soft/firmware; Diagnosis For each available channel, the intended parameters can be defined or selected. Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 9
The recorded measurement data can be read out via the "logbook" on a USB stick and made available as a date-based, daily *.csv file. Basic menu structure with header for function keys Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 10
The arrangement of the channel representation can be changed as desired: By touching a channel field, it appears in the foreground: The visible range of the timeline is adjustable and zoomable by touch: Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 11
As long as the adjustable limit value is exceeded, the display flashes red. Regular curve after falling below the limit value. (here 55.0 ppm) : Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 12
A sensor test is carried out to check the measuring cell. By touch, the respective sensor is selected: Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 13
After confirming with "Enter" you will be taken to the test menu. You can now start the test ( yes ), or leave the menu again ( no ): Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 14
In the meantime, the analog output is switched to "sample and hold": The test result is displayed and, as in the example shown, again with "ok" switched to measurement mode: Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 15
(here 50.6 ppm) and the relay "Maintenance" is activated. The sensor is cleaned and the success is checked with "ok". Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 16
If the cleaning result is positive (< 0.1 ppm), the voltage is restored again switched on and the sensor can be coated with phosphoric acid Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 17
(see NAMUR recommendations; Function "Key Switch") The following image will then appear with the selection option: Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 18
After completion of the work, the maintenance mode can be deactivated again via the button "maintenance sensor": Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
For this purpose, the IP address of the AQUATRACE® IV must be adapted according to the conditions at the place of use. When the system is fully loaded, please press "change configuration": Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 20
Next, please navigate to the "digital output": Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
The UltraVNC software is available on the supplied USB stick. This version is portable and does not need to be installed. Please launch the UltraVNCViewerPortable application: Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
The adjustment of the sensor characteristic curve may only be carried out by specially trained personnel authorized by the operator. The same applies to the setting of a desired "final value". Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 23
The following pictures are intended to illustrate this possibility: Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
"no sensor" there is no measuring signal on the corresponding measuring channel = no measuring cell connected; Measuring cable or sensor defective Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Measuring cell see 8.1 Status output see 9.0 Analog output see 8.2 Relay outputs see 8.3 Relay outputs see 8.4 Ethernet Socket Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
Page 30
Relay outputs see 8.3 Relay outputs see 8.4 USB – Connection for - Updates or -Copy of the recorded log. Data Power switch and Mains connection Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...
➢ P2O5-measuring cell ➢ Stainless steel Note: Also available in PVDF and many other configurations. Depends from your needs! *** Please contact us *** Radeberger Str. 21; 01900 Großröhrsdorf Tel. / Fax. 0049 35952- 429465 /- 429457 info@dks-engineering.de E-Mail: ; www.dks-engineering.de...