10 - 2
MOBILETT Plus only
This supplies the turn-plate solenoid coil with a lower voltage of about 8 V DC.
The magnetic force resulting in the turn-plate solenoid is somewhat greater than the
opposing forces in the turn plate's system of levers and springs. This holds the turn plate
in position.
The magnetization current surge in the turn-plate solenoid coil following energizing the
relay MR1 with the signal "PIVOT ON" can be measured with a current probe in cable K
68.