Memory input
When a valid measurement is made, the result (systolic, diastolic, pulse rate,
date and time) will be stored in the memory automatically as soon as the
device is turned off or when another measurement is initiated.
The device is capable to store 2×99 readings, and also compute the average
of the latest 3 readings. When readings stored more than 90, the oldest
ones will be replaced.
Memory recall
By pressing Memory button, AVG will be shown in the middle left hand, which
means you entered into memory function. AVG means the average of all
recorded readings. Now press the Memory button again, 01 will be shown in
the down left hand corner, which is your latest reading in the memory,
when you press it again, 02 will be shown which is the one before the latest
reading. When you press the button repeatedly, the records will be displayed
sequentially
Er Indicators and Troubleshooting
Error indicators
E
rror indicators
Indication:
Incomplete inflation
Correction:
Check the cuff has been connected with device firmly, if same
Er message indicated, request a repair
Possible sources of errors
30
Indication:
cuff over inflated and release automatically
Correction:
Take a break, and reposition the cuff and measure again
Indication:
Could not obtain the pulse rate
Correction:
Reposition the cuff and measure again
Indication:
Strong Electromagnetic interference
(Mobile phone and computer)
Correction:
Relax and take a break, and then measure again.
Indication:
irregular pulse rate or blood pressure,
Correction:
Relax and take a break, and then measure again.
Troubleshooting
Problem
Recommended method
Make sure that the batteries are installed in the proper direction.
No display
(Polarity matches to the indication in the battery case.)
No
Check that the cuff is positioned properly, pgo over the measurement
measurement
procedure again.
occurs
Blood pressure
Check if the arm is positioned properly
readings are too
Posit ion your arm on a table, so the cuff is at the same level as your
high or too low
heart Remain seated during the entire measurement period
Refrain from hand and body movements during measurement
No Cuff Inflates
Bad machine, contact seller