LOGGER S0841
Dual channel temperature logger with two binary inputs and LCD display
Instruction Manual

Advertisement

loading
Need help?

Need help?

Do you have a question about the S0841 and is the answer not in the manual?

Questions and answers

Summary of Contents for Comet S0841

  • Page 1 LOGGER S0841 Dual channel temperature logger with two binary inputs and LCD display Instruction Manual...
  • Page 2 Instruction Manual for use of S0841 logger Instrument is designed for measurement and record of temperature from two external temperature probes and two binary signals from external contacts. Temperature probes and binary signals are connected by means of connectors. Measured value of temperature including calculated temperature difference of both channels and actual state of binary inputs are displayed on dual line LCD display and are stored in adjusted time interval to internal non-volatile memory.
  • Page 3 Limit condition: temperature -40 to +70 °C, humidity 0 to 100 %RH Storing condition: temperature -40 to +85 °C, humidity 0 to 100 %RH Dimensions: 93 x 64 x 29 mm Weight including battery: approximately 120 g Material of the case: ABS IE-LOG-S0841-07...
  • Page 4 If alarms are about to be applied, set both limits for each measured quantity and enable alarm (not relevant to binary inputs) optionally enable permanent alarm indication (alarm with memory) switch ON or switch OFF the display of logger IE-LOG-S0841-07...
  • Page 5 = T1 – T2 difference between channels (lower line "DIF"). T Instrument now displays actual binary state of channel 3 (LCD upper line) and channel 4 (LCD lower line). Symbol On indicates closed contact, symbol OFF indicates open (or not connected) contact. IE-LOG-S0841-07...
  • Page 6 If value of measured quantity gets outside of set limits, alarm of proper quantity is active and it is indicated on the display. It is possible to select "memory alarm mode" when alarm is indicated permanently up to reset from the PC. IE-LOG-S0841-07...
  • Page 7 Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line. After appearance it is necessary immediately (till indication point is displayed) to remove magnet from guide slots and logger switches ON. IE-LOG-S0841-07...
  • Page 8 LCD upper line. After decimal point appearance it is necessary immediately (till indication point is displayed) to remove magnet from guide slots. Reading CLR MIN MAX appears for several seconds and MIN/MAX values will be cleared. IE-LOG-S0841-07...
  • Page 9 EN 55011 Class B immunity: EN 61000-4-2 (levels 4/8 kV, Class A) EN 61000-4-3 (intensity of electric field 3 V/m, Class A) EN 61000-4-4 (levels 1/0.5 kV, Class A) EN 61000-4-6 (intensity of electric field 3 V/m, Class A) IE-LOG-S0841-07...