Page 1
LOGGER S0141 Four channel temperature logger with display Instruction Manual...
Page 2
Instruction manual for use of temperature logger S0141 Logger is designed for measurement and record of temperature from four temperature probes connected via connectors. Measured values are displayed on the two-line LCD display and are stored in selectable time interval to internal nonvolatile memory. All logger control and setting are performed from the PC and password is applicable.
Page 3
Communication with computer: via RS232 (serial port) by means of COM adapter or USB port by means of USB adapter; data transfer from logger via communication adapter is optical Real time clock: adjustable from computer, integrated calendar including leap years Error of internal RTC: <...
Page 4
After connecting the logger to the computer reading of logger info is enabled by means of the PC software and also setting of the instrument accordingly to the user needs (menu Configuration / Setting of instrument parameters). Before logging start it is necessary: •...
Page 5
Consequently reading of estimated remaining battery capacity for approximately 2 s is displayed (values 0 to 100%). It is valid if logger is operated at temperature from -5 to +35°C. If logger is operated often outside of the above temperature range battery life can be reduced to 75%, i.e. if displayed remaining battery capacity drops below 25%, it is recommended to replace the battery.
Page 6
Alarm indication on the display It is necessary to enable alarm function from the PC and set for each quantity lower and upper limit. If measured value is inside of set limits, alarm of proper quantity is not active. If value of measured quantity gets outside of set limits, alarm of proper quantity is active and it is indicated on the display.
Page 7
Switching the logger ON by magnet Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line. After appearance it is necessary immediately (till indication point is displayed) to remove magnet from guide slots and logger switches ON. Switching the logger OFF by magnet The procedure is identical with the above procedure for switching ON.
Page 8
Replacement procedure: • switch off the logger by the PC program or by magnet (if low battery allows) • unscrew four corner screws and remove the lid • remove old battery by pulling the glued tape • insert new battery respecting the correct polarity (see symbols + and – near battery holder). If you connect new battery up to 30 s, all logger settings remain unchanged.