Electrical connection
Voltage-bearing elements
Danger to life from electric shock as a result of improper installation.
Electrical connection may be carried out only by trained and authorised electricians.
Ensure the system is adequately earthed.
Establish an adequate fuse protection on customer side (depending on the model).
Unscrew and remove the cover of the relay box.
Pay attention to grounding cables!
1
9
L N PE
AC out
6
1b
Fig. 3:
Backing pump relay box, single phase, 7 A
1
Casing
1.1
Casing cover
1a
Control lead for electronic drive unit
1b
Mains connection output, backing pump
1c
Mains connection input
3
Connection terminal
4
Semiconductor relay
Install the power supply and the backing pump connection, according to their technical
data to the terminals of the relay box.
– L/L1: Conductor
– N: Neutral
– PE: Protective earth
Secure cables with cable fittings.
Close the relay box.
DANGER
3
4
+ -
L1 N PE
AC in
6
1c
6
7
9
11
L/L1 Conductor (black)
N
PE
Electrical connection
1.1
11
7
1a
Cable fitting M16
Cable fitting M12
Fuse
Earthing cable
Neutral (blue)
Protective earth (yellow/green)