Tc 1200; Tcp 350; Operation; Operation Mode Backing Pump - Pfeiffer Vacuum HiPace PM 071 282-X Operating Instructions Manual

Backing pump relay box, shielded
Table of Contents

Advertisement

Operation

5.3

TC 1200

5.4

TCP 350

6
Operation
6.1

Operation mode backing pump

10
TC 1200
1a
1c
 Connect the control lead of the relay box directly to a vacant accessory port of the
electronic drive unit or using an Y-Connector.
 Make the settings and control via the interfaces of the electronic drive unit.
TCP 350
 Connect the control lead of the relay box to the "remote" connection of the electronic
drive unit using an adapter cable.
 Make the settings and control via the interfaces of the electronic drive unit.
 Establish the mains supply for the relay box.
– Make sure the supply voltage for the backing pump is valid.
– Observe the rating plates!
Operation of a connected backing pump via the electronic drive unit depends on the
backing pump type.
Operation mode [P:025]
"0" continous operation
"1" Intermittend operation
"2" Delayed switching on
 Adjust the parameter [P:025] to the desired value.
1
PM 061 377 -T
PM 061 376 -T
1a
1c
recommended backing pump
all kinds of backing pumps
diaphragm pumps only
all kinds of backing pumps
V
1b
V
1b