Operation
5.3
TC 1200
5.4
TCP 350
6
Operation
6.1
Operation mode backing pump
10
TC 1200
1a
1c
Connect the control lead of the relay box directly to a vacant accessory port of the
electronic drive unit or using an Y-Connector.
Make the settings and control via the interfaces of the electronic drive unit.
TCP 350
Connect the control lead of the relay box to the "remote" connection of the electronic
drive unit using an adapter cable.
Make the settings and control via the interfaces of the electronic drive unit.
Establish the mains supply for the relay box.
– Make sure the supply voltage for the backing pump is valid.
– Observe the rating plates!
Operation of a connected backing pump via the electronic drive unit depends on the
backing pump type.
Operation mode [P:025]
"0" continous operation
"1" Intermittend operation
"2" Delayed switching on
Adjust the parameter [P:025] to the desired value.
1
PM 061 377 -T
PM 061 376 -T
1a
1c
recommended backing pump
all kinds of backing pumps
diaphragm pumps only
all kinds of backing pumps
V
1b
V
1b
1