Cochlear implant speech processor and its accessories (18 pages)
Summary of Contents for NUROTRON CS-10A
Page 1
CS-10A Operation Manual of Cochlear Implant V 2.4 Zhejiang Nurotron Biotechnology Co.,Ltd. Address: No.17 Longtan Road, Yuhang District, Hangzhou, Zhejiang 311121, China National Service Hotline: +086-4006-333-571 BC040122 Website: http://www.nurotron.com...
Page 2
Part I Surgeons’ Manual 1. Working Principle 2. Product Structure 3. Model 4. Serial Number 5. Performance Index of the Product 5.1 Basic parameters 5.2 Performance index 6. Scope of Application 7. Contraindication 8. Measuring Cochlea size and selecting electrode length 9.
Page 3
5.2 Airport security check 6. Taking Air Plane 7. Other information 8. Warranty 9. Manufacturer's Information Part III Audiologist’ Manual 1. Mapping Time 2. Preparatory Works before fitting 3. Basic Fitting Procedures 4. Manufacturer's Information Part IV Nurotron Cochlear Implant System Components List...
1) Range of electrode impedance: 1KΩ~20KΩ 2) Performance of stimulus signal of electric pulse: Product model: CS-10A. Product model of the implant is printed on the shell of stimulator at Max. stimulus frequency of every current source: 20 kHz the place shown as in Figure 2.
Implantation of Cochlea published by China Medical Association, which mainly includes following procedures: processor produced by Nurotron Company. The part has been certified by national or local SFDA departments. Its commissioning shall be conducted with the aid of NuroSound fitting software.
05 06 Operation Manual of Cochlear Implant Operation-related Risk Prevention and Warning pressure. 3) Do not place the commissure directly over electrode array. 11.1 Risk prevention 4) Incorrect implanting position of electrode array may impair the auditory sense of human body 1) Cochlear implantation operation shall be performed under a general anesthesia.
Page 7
3) Always ensure asepsis of implant. Do not allow patients with a CS-10A cochlear implant to be in the area of an MRI scanner unless 4) Make sure no falling of implant on hard objects.
Marks shall be treated as polluted biologically hazardous substance that shall abide by universal Following marks are printed on CS-10A implant component or package: precautionary measures, placed inside a leakproof brine vessel which has been sterilized (or disinfected), and sent back to Nurotron Company.
Page 9
3T MRI, the implant magnet must be removed by a qualified surgeon. MRI Warnings Do not allow patients with a CS-10A cochlear implant to be in the area of an MRI scanner unless the following conditions have been met: ● The external sound processor and headpiece are removed before entering a room where an MRI scanner is located.
13 14 Operation Manual of Cochlear Implant some discomfort and pain at the implant site may still be experienced. Please consult with your Electromagnetic Interference physician if this is an issue. Wearers of cochlear implant are open to electromagnetic interference of some equipments. But Monopole electro surgical apparatus is strictly prohibited.
3) Production License No.: ZSYJSCX No.20130030 5.2 Airport security check 4) Manufacturer: Zhejiang Nurotron Biotechnology Co., Ltd. Security gates of some airports are provided with magnetic induction facilities, which may 5) Address: No.17 Longtan Road, Yuhang District, Hangzhou, Zhejiang 311121, China cause damage to your speech processor.
Page 12
17 18 Operation Manual of Cochlear Implant This part (for audiologist) describes the starting up and mapping time and specific procedures of cochlear implant after operation. Mapping Time It is suggested to switch on the device and do the first mapping 2 to 4 weeks after the surgery, Part depending on the healing condition.
Page 13
2) Product Standard No.: GXZZ 20173464498 3) Production License No.: ZSYJSCX No.20130030 Components List 4) Manufacturer: Zhejiang Nurotron Biotechnology Co., Ltd. 5) Address: No.17 Longtan Road, Yuhang District, Hangzhou, China Registration address: No.17 Longtan Road, Yuhang District, Hangzhou, China 6) National Service Hotline: +086-4006-333-571 7) Website: http://www.nurotron.com...
Page 14
21 22 Operation Manual of Cochlear Implant Product Reference Product Name Picture Product Reference Product Name Picture CP01.0016 COCHLEAR IMPLANT CS-10A (SD)* CP02.2130 Enduro Charging Stand CP01.0018 COCHLEAR IMPLANT CS-10A (TS) CP01.0019 COCHLEAR IMPLANT CS-10A (TM) CP02.2145 Enduro Remote Control CP01.0020...
Page 15
23 24 Operation Manual of Cochlear Implant Product Reference Product Name Picture Picture Product Reference Product Name Picture CP00.0081 Programming Cable for Enduro CP02.1038 VENUS 60C Rechargeable BTE Battery CP02.1039 VENUS 60C Body-worn Rechargeable Battery CP00.0080 Programming Cable for VENUS (NSP-60C) VENUS 60C Connection wire CP02.1041 CP00.0079...
Page 16
CP02.1034 VENUS 60C BTE Battery Extension Wire (20cm) CP02.1035 VENUS 60C BTE Battery Extension Wire (30cm) CP02.1036 VENUS 60C BTE Battery Extension Wire (40cm) *CS-10A (SD) is also noted as CS-10A in internal documents of Nurotron. Hear more, Live better!