Page 1
Hear more, Live better! Cochlear Implant Speech Processor and Its Accessories V 2.3 Zhejiang Nurotron Biotechnology Co., Ltd. No.17 Longtan Road, Yuhang District, Hangzhou, China National Service Hotline: +086-4006-333-571 Website:http://www.nurotron.com BC040108...
Page 2
12.3 Other Preventive Measures 13. Manufacturer's and Authorized representative in the European Community Information 14. After-sales Service Information 14.1 Service content 14.2 Information on service institutions 15. Mark Symbols 16. Guidance and Manufacturer’s Declaration 17. Nurotron cochlear implant system component list...
Implant recipients below 14 years old or persons with an intellectual disability are not allowed to complete all operations in this manual, Enduro cochlear implant speech processor and its accessories are the non-implantable part of the and the operations of the speech processor shall be completed by the cochlear implant system, mainly composed of signal processing unit, transmission wire, guardian/caregiver.
Page 4
03 04 2.1 Signal Processing Unit slowly flickers for three times, and then the indicating lamp is off to enter the normal operating status. When the battery power is less than the specified value, the alarm prompt flickers. Figure 2 is the composition diagram of the signal processing unit. The microphone (7) collects ◆...
Page 5
Warning Chocking hazard-small parts included in this device, please supervise children The Enduro speech processor contains small parts that may cause choking if under 3 years old during use. swallowed or ingested. Please supervise children under 3 years old during use.
Page 6
The remote controller matched with the signal processing the rechargeable lithium battery cell. unit of Enduro series may achieve the following functions: Battery cycle life is at least 300 times. ◆ Operation control function: It is able to be matched with the specific signal processing unit.
Page 7
Marking naming method of the signal processing unit: ED (represent that it belongs to Enduro ◆ Volume “-” key: Press “-” key for once so that the sound volume is reduced by Level 1, after the...
Diameter of coil ≤31.5 mm CS-10A series cochlear implant produced by Nurotron Company; at the time of fitting, it is used Transmission coil weight ≤9 g with NuroSound fitting software and fitting box.
Page 9
13 14 Operation Manual of Cochlear Implant Speech Processor 8.3 Connection of Transmission Wire In daily use, the remote controller can be used to carry out the following operation for the speech processor: Installation of transmission wire: As shown in the Figure 12, one hand holds the signal processing 1) Power on/off the speech processor;...
Page 10
15 16 the other hand holds the pin-header plug for transmission wire. Gently pull out; do not strongly pull processing unit, this indicates that the battery enters the low power mode. At this time, please take down the battery, and connect it to the charging holder for charging. After the completion of the cable or reverse the pin-header plug for transmission wire.
Restart the remote controller, and carry out pairing operation again. damaging the battery contacts. If the related problems still exist after your check, please contact your audiologist or Nurotron At the time of cleaning the surface of the speech processor, gently wipe with a piece of cloth dipped Company.
Page 12
When the dustproof membrane for microphones on the cover plate for keys of the signal processing processing unit, and ensure that the connection distance is unit gets dirty or the quality of hearing the sound is reduced, please contact Nurotron Company to The remote controller is not less than 1.5m.
6) In case of leakage on battery, the skin or eye shall not be in contact with the leaked fluid. Once All buyers of Nurotron Cochlear Speech Processor have a right to the after-sale service provided by exposed, immediately wash with clear water, and go to hospital for treatment. After the battery is Nurotron Company, which covers: disposed, please immediately wash hands.
23 24 Mark Symbols Guidance and Manufacturer’s Declaration Below cables information are provided for EMC reference. The following symbols are printed on the assembly or package of Enduro cochlear implant speech processor: Max. cable length, Cable Number Cable classification Shielded/unshielded...
Page 15
25 26 Operation Manual of Cochlear Implant Speech Processor EMI Compliance Table (Table 1) Table 3 – Proximity fields from RF wireless communications equipment Immunity test levels Table 1 - Emission Test frequency Band (MHz) (MHz) Home healthcare environment Phenomenon Compliance Electromagnetic environment 380-390...
Page 16
CP02.2138 Enduro Transmission Wire (20cm) CP02.2139 CP02.2131 Enduro Signal Processing Unit CP02.2140 Enduro Transmission Wire (28cm) CP02.2132 Enduro Rechargeable BTE Battery Enduro Waterproof case CP02.2176 CP02.2134 Enduro Transmission Coil Enduro Signal processing unit button assembly CP02.2148 Hear more, Live better!
Page 17
Product Name Picture Picture Product Reference Product Name Picture CP02.1038 VENUS 60C Rechargeable BTE Battery CP00.0081 Programming Cable for Enduro CP02.1039 VENUS 60C Body-worn Rechargeable Battery CP00.0080 Programming Cable for VENUS (NSP-60C) VENUS 60C Connection wire CP02.1041 CP00.0079 Fitting Cable USB for Rechargeable Body-worn Battery CP02.1046...
Page 18
CP02.1034 VENUS 60C BTE Battery Extension Wire (20cm) CP02.1035 VENUS 60C BTE Battery Extension Wire (30cm) CP02.1036 VENUS 60C BTE Battery Extension Wire (40cm) *CS-10A (SD) is also noted as CS-10A in internal documents of Nurotron. Hear more, Live better!