Page 2
Safety Instruction For safety operation please follow the safety instruction. WARNING Settings, Calibration, Inspection and Maintenance should only be done by a trained person or staff. WARNING Please make sure the indicator has sufficient grounding. WARNING The indicator is a static sensitive piece of equipment.
Page 3
Instruction This indicator is designed for basic weighing function. Main function Weighing function: Zero, tare, G.W, N.W, accumulation, printing, animal -weighing. kg/lb conversion. Print format: S.N. G.W N.W Tare. Date, Time Options: Pinter RS232/RS485 serial interface or second display Technical parameter Accuracy class 6000 e Resolution...
Installation and calibration Power supply connection The indicator is powered by A/C adapter that plugs directly into the “DC” pin at the bottom of the indicator. Connection of load cell and indicator The indicator can be connected up to 6 load cell of 350Ω, 4 wire or 6 wire load cell both ok.
Communication interface RS232 : DB9 Pin DB9 definition Pin function and definition as bellows: DB9 joint Definition Function Sending data Receiving data Ground interface Note: if RS485, The connection pin is 2 and 5 pin. Basic operation Power on & off Press 2 seconds to power on or off the indicator.
Page 6
Tare operation During “TARE” operation when gross weight is tared the indicator will show the Net weight. The “Net” “tared” status light is on. In tare mode, Press” TARE” key to clear the tare weight and the indicator will show the gross weight.
Page 7
Print If the weight is stable, press” PRINT” you can print the weight. COUNT 1.In weighing mode, load the Pieces on the platform scales, Press” Count” the indicator show” PCS 0” press” Zero” key in the quantity, press” Print” to confirm it 2.
Page 8
The key functions in setting: Enter Down Left Power switch, exit setting without saving, “Count” exit with saving changes. Calibration operation: According to the second method which can enter setting menu, C01-C39 display Remark step Method of operation [C01 After enter calibration mode, it display [C01...
Page 9
press [C03 Set graduation press [C03 option:1/2/5/10/20/50 [C03 press Select required graduation example: graduation 5: [C03 5] press [C04 Max capacity press [0100.00] [0100.00] press example:max weighing 100kg: [0100.00] press [C05 Zero calibration press [C05 Option [C05 0=no need zero calibration press press [CAL...
Page 10
[0080.00] As bellows: [CAL End] Input the 0080.00, count down , then indicator shows 0080.00 , calibration is over. If you want to set application function parameter. Press “PRINT” if you want to exit press “Count” press [C07 Default parameters setting press option:0=non-restore default parameters...
Page 11
Hold Instruction: function Hold mode Peak-hold: it shows the max. data, mainly application for materials testing, SBI-505 such as tension and pulling force. Date-hold: it shows current weight value. Mainly application for animal weighing. Kg/lb Kg/lb C12=0 stop kg/lb conversion...
Page 12
option: 0= Close serial interface data output 1=Continuous sending, connect big display 2=Print method, connect printer. Serial interface 3=Command request method , data output method Communication connect computer. setting 4=PC continues sending format, connect computer. 5=PC/ display continuous sending format. option:...
Page 13
Options: 0= close zero tracking 0.5=±0.5d 1.0=±1.0d 2.0=±2.0d Automatically zero 3.0=±3.0d tracking range 4.0=±4.0d 5.0=±5.0d Zero tracking Note: 1. d = division 2. the zero tracking range can not bigger than manual zero range. Options: 0= close zero tracking time Automatically zero 1= 1 second tracking time...
Page 14
Noise filter 1=1 digital filter strength 2=2 digital filter strength 3=3 digital filter strength C30=0 yy.mm.dd Print time and date C30=1 mm.dd.yy C30=2 dd.mm.yy C30=3 yy.mm.dd Analog output C31 output type C31=0 0~5Vouput setting C31=1 4~20mA output NA SBI-505...
Page 15
C33=0 close relay output setting C33=1 Open relay output function 1 NA SBI-505 C3=2 Open relay output function2 C33=3 Preserved menu Muti C34= 0~99 Add. Code communication Communication add. NA SBI-505 add. Wireless C35=0~99 signal communication NA SBI-505 Gravity C36=9.7000~9.9999 calibration location Gravity C37=9.7000~9.9999...
Output format Big display continuous sending format Output continuous format State A Bits0,1,2 Decimal point position XXXXXX0 XXXXXXX XXXXX.X XXXX.XX XXX.XXX Bits3,4 Division State B BitsS function Bits0 gross=0, net=1 Bits1 Symbol: positive =0,negative =1 Bits2 Overload(or under zero)=1 Bits3 dynamic=1...
Page 17
Bits4 unit:lb=0, kg=1 Bits5 Constant 1 Bits6 Constant 0 State C Bit2 Bit1 Bit0 unit Kg or lb Bit 3 printing=1 Extend Bit 4 display=1 Bit 5 Constant 1 Bit 6 Constant 0 Computer continuous sending format Data weight status, ST= standstill, US= not standstill, OL= overload weight mode, GS=gross mode, NT=net mode weight of positive and negative, “+”...
Page 18
RS232COM serial interface can receive simple ASCII command. Command word and role as follows: Command NAME Function TARE Save and clear tare ZERO Zero gross weight PRINT Print the weight G.W/N.W Read gross weight or net weight Kg/lb Kg/lb conversion Check gross weight at net weight mode R command receive data format Print format...
6. Maintenance 6.1 Regular error and solution ERROR REASON SOLUTION 1. Overload 1. reduce the weight 2. wrong connection 2. check load cell connection with load cell 3. Inspect load cell. Check the UUUUUU Load cell input and output voltage 0R problem.
Page 21
1. check the connection is During calibration, the correct ERR3 input signal 2. check load cell negative During calibration, the After the platform is stable, start ERR4 signal is unstable calibration ERR5 Change PCB Default parameters Parameter instruction Default Calibration Decimal digits Resolution Max.
Page 22
Date setting Time setting Serial interface data output Serial interface Baud rate 3(9600) Zero manually Initial zero Zero tracking range 0.5 Zero tracking time Overload range Negative range Standstill time Standstill range Dynamic filter Noisy filter Print format Analog signal options 4~20mA testing Relay output setting Muti PC communication add.