Ackermann DoUbLE RoLLER PUMP foR LAPARoScoPY coNTENTS 1. front and rear panel 2. Technical specifications 3. Intended use 4. Delivery range 5. Safety Instructions, disposal, safety signs and symbols 6. Handling of tubing set 6.1. Composition 6.2. Reprocessing of reusable tubing set 6.3.
Ackermann DoUbLE RoLLER PUMP foR LAPARoScoPY 2. TEcHNIcAL SPEcIfIcATIoNS Parameter Suction-Irrigation pump Power consumption 36 W Supply voltage 100-240 V/AC - Mains fuse T 3,15 A Power frequency 50/60 Hz Classification acc. to EU directive 93/42/EEC Class IIa Type acc. to DIN EN 60601-1:2005 Protection class acc. to DIN EN 60601-1:2005 Manufactured acc.
• Model No. / Serial No. • Failure description • Error message The unit consists of: Pos. Specification Article no. Qty. Suction-Irrigation pump 16-2080 Test report User manual Silicone suction-irrigation tubing set, autoclavable 16-2080-100 Main cable Euro 2 m 15-1720-800 Fuses 15-1720-901 Suction-irrigation cannula, dia. 5 / 10 mm, autoclavable...
Ackermann DoUbLE RoLLER PUMP foR LAPARoScoPY • Liability may be voided by misuse, abuse, violation of instructions for use, omission of required maintenance, opening the equipment modifications, repairs by unauthorized persons! Use of other than original accesaccessories shall void warranty and liability. • Equipment has to be used at all times in accordance with the operating instructions! This device complies with currently valid EMV standards. However tests of EMV standard DIN EN 60601-1-2:1993-05 are carried out. Nevertheless, EMV interference may lead to malfunctioning. If malfunctioning is observed, ensure that the device has been installed and is being operating in accordance with EMV guidelines for use supplied with the device. Note that portable and mobile high-frequency communication equipment may influence the operation of this device. If the...
Ackermann DoUbLE RoLLER PUMP foR LAPARoScoPY 6.4. InSerTIg THe TubIng SeT Tubing set has to be completely assembled and sterile, see chapter 6.3. Step 1: Take the blue irrigation tube. Place the tube stopper (blue) in the upper tube holder (marked blue). Pull the pump segment taught around the roller wheel and route it through the lower tube holder until the second tube stopper (colorless) locks firmly into place.
Ackermann SHAVER SYSTEM 7. ADAPTER DUcT - PNEUMATIc coNTRoL The pneumatic control controls the pump’s suction and flow rate. The adapter duct is optionally supplied with two cannulas which are attached to the handgrip by a threaded fitting. If necessary, the cannula can even be changed during the operation.
Ackermann DoUbLE RoLLER PUMP foR LAPARoScoPY Pneumatic double footswitch: M Left footswitch- IRRIGATION N Control connector - IRRIGATION (blue) 0 Control connector - SUCTION (red) P Right footswitch - IRRIGATION 8. funCTIonal buTTonS, STarT-up, funCTIonal TeST & operaTIon 8.1. SeTTIng up 1. Plug the AC plug into an AC outlet and switch on the device using the AC power switch. The STANDBY LED light yellow.
Ackermann SHAVER SYSTEM 8.3. coNTRoL bUTToNS 8.3.1. PowER SwITcH The AC power switch is used to switch the device on and off. 8.3.2. oN/off The push button ON/OFF activates/deactivates the device. To activate device: Press the push button ON/OFF. The ON LED light green and the STANDBY LED goes out.
Ackermann DoUbLE RoLLER PUMP foR LAPARoScoPY 9. WarnIng SIngnalS, faIlure- & faulT DeSCrIpTIon • If a beep signal lasting for three seconds sounds and all the LEDs are flashing, there is a fault with the device. a. The pump stops immediately and cannot be restarted. Deactivate the device by pressing the ON/OFF button. Remove the tube set and reactivate the device after 15 seconds. The device then performs a self test.
Ackermann SHAVER SYSTEM Make sure that this certificate is handed over and that a label is attached to the rear side of the device stating the date of the next inspection. The manufacturer is relieved of any liability for the operational safety of the device if it is opened, repaired and/or modified by anyone other than the authorized service personnel. Medical devices regulations request the user of the device to carry out regular function and safety checks. As manufacturers, we recommend quarterly inspections which are to be carried out by hospital technicians or other authorized personnel.
Ackermann DoUbLE RoLLER PUMP foR LAPARoScoPY 12.2. CHeCkIng THe floW raTe For this test you will need: • The device • New tube set • 3,5 liter measuring beaker with scale (A) • Water reservoir with 3,0 liters of water (B) • Manual control/double footswitch • Adapter duct with cannula connected • A watch that displays seconds, or a stopwatch The suction and irrigation flow rate test procedure is carried out in the following steps: 1. Plug the AC plug into an AC outlet and switch on the device using the AC power switch. The STANDBY LED light yellow.
Page 18
Ackermann 0483 ackermann Instrumente gmbH Eisenbahnstrasse 65-67 78604 Rietheim-Weilheim Germany user_manual-16-2080-revision_06/21 Copyright by Ackermann Instrumente GmbH. Specifications, design and accessories in this user manual are subject to change without any notice or obligation on the part of the manufacturer.