Certification CE0483 3. InDenTeD uSe 16-2083 is a fluid pump featuring pressure limitation and is used in minimally invasive surgery. It can be applied in all endoscopic surgical procedures where the body cavities are not completely filled with liquid and where the volume of irrigation liquid is small compared to the volume of the anatomical cavity. The pump system is ideally suited for use in the abdomen, thorax and retroperitoneum.
ATTENTION! Contaminated devices are not accepted for check or repair. The unit consists of: Pos. Specification Article no. Qty. Membrane suction-Irrigation pump 16-2083 Test report User manual Silicone suction-irrigation tubing set, autoclavable 16-2083-102/-103 Main cable Euro 2 m 16-2083-104 Fuses...
Page 6
Ackermann MEMBRANE SUCTION-IRRIGATION PUMP not be used in the vicinity of combustible gases. • The manufacturer is responsible for the safety, reliability and performance of the system only if: - Assembly, expansions, new settings, changes and repairs are carried out by authorized persons. - Electrical installation of the operating room corresponds to the national standards for medical institutions.
Ackermann SHAVER SYSTEM 6. InSTallIng THe boTTle HolDerS There are two separate bottle holders included to support bottle with irrigation liquid and secretion container. Please install both bottle holders left and right to the bottom of pump using the four fixing screws provided (see illustration). Insert containers into the holders and connect tubing set (see chapter 9.4.). Illustration of bottle holders: 7.
Ackermann MEMBRANE SUCTION-IRRIGATION PUMP 7.2. REPROCESSING THE SPIkE Read all instructions carefully before reprocessing the spike. For pre-cleaning/cleaning we advise to use Neodisher FA (manufactured by Dr. Weigert). Pre-cleaning prevents the transmission of germs and stops residues from drying on. It therefore needs to be performed immediately after the operation.
Ackermann SHAVER SYSTEM 9. HanDlIng oF TubIng SeT 9.1. elemenTS oF TubIng SeT The tubing set consists of 4 single tubes, which are color-coded (see chapter 9.3). Tubing set IRRIGATION (blue): Tube 1 has got one male LL-connector and one female LL-connector. Single-use sterile filter is fixed to male LL-connector („to spike“). Tube 2 does not contain any components. Tubing set SUCTION (red): Tube 3 has got one male LL-connector. Single-use sterile filter is fixed to male LL-connector („to pump“).
Page 10
Ackermann MEMBRANE SUCTION-IRRIGATION PUMP 9.3. IlluSTraTIon oF TubIng SeT 24 LL-connector female 27 Disposable sterile filter 25 Tube 1 IRRIGATION (blue) 28 Tube 2 IRRIGATION (blue) 26 LL-connector male 29 Tube 3 SUCTION (red) IMPORTANT! Insert tubing set only when device is switched-on and self test is passed successfully. Do not use tubing set without disposable sterile filters! Start surgery with a new, completely filled bottle of irrigation liquid, prepare a spare bottle.
Ackermann SHAVER SYSTEM 9.4. ConneCTIng THe TubIng SeT Tubing set has to be sterile, see chapter 9.2. Step 1 - Tube 1, blue, irrigation: Connect disposable sterile filter to female LL-connector of spike. Connect male LLconnector of the tube to disposable sterile filter. Connect female LL-connector of the tube to connector IRRIGATION on pump. Step 2- Tube 2, blue, irrigation: Connect one side of the tube to tube connector on spike and the other side of the tube to irrigation connector on instrument.
Ackermann MEMBRANE SUCTION-IRRIGATION PUMP IMPORTANT! Remove tubing set after surgery. Sterilize tubing set, secretion container and instruments, always replace sterile filters! 9.6. uSIng THe DevICe In operaTIng room - IlluSTraTIon IMPORTANT! Do not use pump without sterile filters! Irrigation bottle must be qualified for the overpressure of pump system! 10.
Ackermann SHAVER SYSTEM IMPORTANT! Footswitch shall be disconnected when device is switched off! ATTENTION! Footswitch must not be connected during self test! In case of malfunction please remove adapter (C) from connector (13) on front panel for a short time and connect it again. 11. FunCTIonal buTTonS, STarT-up, FunCTIonal TeST & operaTIon 11.1. FunCTIonal buTTonS 11.1.1. POWER SWITCH The power switch (14) is used to switch the device on and off.
Ackermann MEMBRANE SUCTION-IRRIGATION PUMP WARNING! EXPLOSION HAZARD! The electrical components are not explosionproof. Do not use in an area where flammable gases are present! 11.2. STarT-up/SelF TeST 1. Connect the pump to mains. Switch on device with power switch. LED STANDBY (4) will be on. Do not connect tubing set yet! Do not connect footswitch yet! 2. Activate the device by pressing button ON / OFF (5). A short beep sounds.
Ackermann SHAVER SYSTEM If pressure and vacuum decrease below preset values they are switched on again. If only one of both (IRRIGATION or SUCTION). Is needed you can switch off refering generator: push buttons UP/DOWN to reduce preset pressure/ vacuum to zero.
Ackermann MEMBRANE SUCTION-IRRIGATION PUMP The service partner shall issue a certificate on type and extend of annual inspection and repair. Inspection shall be marked with a label „next check“ on the rear panel of device. Opening of casing, service and any other modifications of device and accessories by unauthorized persons releases manufacturer from any liability for the operating safety of the pump.
Ackermann SHAVER SYSTEM Setup of irrigation pressure and suction vaccum test: 15.2. CHeCkIng THe FooTSWITCH For this test you will need: • Membrane suction-irrigation pump • Footswitch This test will be carried out without liquid. Proceed as follows: 1. Connect pump to mains. Switch on device with power switch.
Page 20
Ackermann 0483 Ackermann Instrumente GmbH Eisenbahnstrasse 65-67 78604 Rietheim-Weilheim Germany user_manual-16-2083-revision_06/21 Copyright by Ackermann Instrumente GmbH. Specifications, design and accessories in this user manual are subject to change without any notice or obligation on the part of the manufacturer.