Message section format ......................27 10.3 Module commands ..........................28 10.3.1 Module information commands ....................28 10.3.2 Module control commands ......................29 PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 3 of 48...
Page 4
How To – Clean the PID Lamp ......................43 13.3 Maintenance Intervals ........................43 Ordering information ..........................45 Spare parts and Accessories ........................46 Contact ..............................48 PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 4 of 48...
The module comes with a built-in linearization and temperature/humidity compensation. Built-in barometric pressure sensors offers later compensation and gas flow monitoring. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 5 of 48...
2 years, except for components in gas path and wearing parts The indicated values were obtained under standardized conditions with 10.6eV lamp. Test gas was Isobuten in synthetic air. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 6 of 48...
VCC level x 0.67, at load 20 µA, internal pull up typ. 10 kΩ V out low 600 mV at load 1mA PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 7 of 48...
Indicates a hazardous situation, which, if not avoided, MAY result in minor or moderate injury. NOTICE Indicates risk of property damage. Usage Indicates a helpful information, hint or recommendation. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 8 of 48...
The functionality of the module will be influenced by pollution. Checks for dust deposits must be done at regular intervals. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 10 of 48...
(including but not limited to negligence), by operation of law, or otherwise. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 11 of 48...
The module comes with an integrated gas flow monitoring. The current gas flow indicator can be readed with measurement values of UART command and I2C data registers. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 15 of 48...
Page 16
- gas flow too low, measurement result integrity loss and module may be damaged • ‘flow over’ - gas flow too high, measurement result integrity loss PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 16 of 48...
(ESD). Work on or in the unit must be done only by qualified personal and in full compliance with the appropriate instructions and pertinent regulations. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 17 of 48...
When in ERROR state the lamp is switched off. The ERROR state can only be left by switch off or reboot the module by sending the appropriate command. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 18 of 48...
Page 19
Control register set D03 In case of ERROR state the error details can be readed (Action ‘error’). PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 19 of 48...
The module result calculation is based on ISOBUTENE gas. The gas response factor is a numeric factor applied to the ISOBUTENE gas result. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 21 of 48...
The calibration will be done with currently selected calibration method in measurement config. Calibration data values stored seperately for ‘standard’ and ‘extended’ calibration. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 22 of 48...
Page 23
ZERO calibration step • no or inappropriate calibration gas applied during SPAN calibration step PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 23 of 48...
Control commands may answer with error message in case of invalid module state, e.g. ERROR state. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 29 of 48...
Mutliple state bits can be set at once e.g. the calibration method is extended and module is in measure state will result in an state value of ‘00004100’. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 30 of 48...
Page 31
Example (write) success pid.measconfig standard;75-15-0;1.200 pid. measconfig ok Example (write) error pid.measconfig standard;;1.000 here curr2 to low pid.measconfig error PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 31 of 48...
Page 32
USAGE – SAVING DATA Saving data may take about 100 ms. Consider this period of time in UART communication timeout detection. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 32 of 48...
32 Example write device.address.i2c 35 device.address.i2c 35 On change the I2C slave address is stored permanently in device. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 33 of 48...
Data Register to access ( 4 bytes ) DATA 0 … DATA N Data bytes ( max 32 ) PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 34 of 48...
Type Register Byte no. Format Module control 0x00000020 0x04 Unsigned 32-bit value Module status 0x00000024 0x04 Unsigned 32-bit value PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 35 of 48...
It is possible to poll for end of saving data by transmitting the I2C slave address in read condition. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 36 of 48...
Mutliple state bits can be set at once e.g. the calibration method is extended and module is in measure state will result in an state value of ‘00004100’. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 37 of 48...
In case of verifying calibration data fails the current valid calibration data is not changed and is set to registers. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 38 of 48...
In case of verifying measurement config data fails the current measurement config data is not changed and is set to registers. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 39 of 48...
(ESD). Work on or in the unit must be done only by qualified personal and in full compliance with the appropriate instructions and pertinent regulations. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 41 of 48...
6. Connect the lamp control connector cable to the lamp. 7. Calibrate the device after replace of front isolation or installation of a new PID lamp. PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 42 of 48...
Parts shall be exchanged no later than 2 years after continuous operation. 0003023 11,8eV Lamp 4 month 0002974 Sensor block 24 month PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 43 of 48...
Page 44
Photoionization Detector Module PIM CPN User Manual PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 44 of 48...
USB to sensor module adapter cable for power supply and communication USB memory stick with documentation and sample programs PIM_CPN - User Manual (en) - 1.08.docx Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com Page 45 of 48...