. The components of the Flash 150 radial compression module. aluminum barrel with an instant-tight fitting for connection to The Flash 75 module has the same features but instead of being delivered the inert gas outlet on the AM-190 manifold (see page 3).
MV-03 MV-02 BIOTAGE FLASH 75/150 ® assembly is fitted with a three-way valve that connects to the CHROMATOGRAPHY MV-03 AM-190 MANIFOLD solvent reservoir and the outlet of the Biotage sample injection PG-01 PG-02 PRR-01 MV-01 PRR-02 MV-03 MV-02 module (Biotage SIM).
316 stainless steel tube with a sanitary clamp at the top and a compatibility information, see Table 3 below. stainless steel 1/8" (Flash 75) or 1/4" (Flash 150) tube fitting at the bottom. The top lid comes with a three-way valve. One port...
» One (1) user manual (P/N 02140). the Flash 75 system and a larger one for the Flash 150 system, For detailed information on how tubing and fittings are attached see Figure 9). For instructions on how to install the grounding and assembled, refer to the section “Installation”...
If returning equipment to Biotage, this should be done in accordance with the material return procedures supplied Biotage Flash 75 and Flash 150 systems are intended solely separately by Biotage. for purification. The systems must be operated in a laboratory environment with an ambient temperature between 4°C and...
Page 10
If the of any chemical or hazardous substance. system is used in a manner not specified by Biotage, the » Personnel working with or near the system must safety features of the system may be compromised.
4. Screw the lower half of the frit holder back in place. Finger- 4. If using a Flash 75 radial compression module: tight is normally adequate. a. Screw the two mounting rods into the mounting pads 5.
Solvent Lines schematic in Figure 13). Warning Note: The Flash 75 start-up kit contains both 1/4" and 1/8" » clear PTFE tubing. The 1/8" tubing is to be used between the Always connect inert gas such as nitrogen to the gas inlet on the AM-190 manifold.
Page 13
Installation BIOTAGE FLASH 75/150 ® CHROMATOGRAPHY AM-190 MANIFOLD PG-01 PG-02 PRR-01 MV-01 PRR-02 MV-03 MV-02 Inert gas supply Figure 12 . Schematics of a system without the sample injection module (SIM). BIOTAGE FLASH 75/150 BIOTAGE FLASH 75/150 ® ® CHROMATOGRAPHY...
The one with the smaller nozzle and a tube with and a pair of fraction collection assemblies (see Figure 15). an outer diameter of 1/8" is for Flash 75 and the one with the larger nozzle and a tube with an outer diameter of 1/4" is for Flash 150;...
Page 15
Installation BIOTAGE FLASH 75/150 ® CHROMATOGRAPHY AM-190 MANIFOLD PG-01 PG-02 PRR-01 MV-01 PRR-02 MV-03 MV-02 Connect to earth ground Figure 16 . System grounding schematic. Note that all the alligator clips must be connected to metal fittings/objects, not to the plastic tubing.
How to Use the System 6. Set the radial compression pressure to 100 psi (6.9 bar) by adjusting the PRR-01 regulator on the AM-190 manifold. 7. Confirm that the cartridge is compressed; the red pressure indicator on the front of the radial compression module sticks out (see Figure 2 on page 2).
Page 19
For flash cartridges packed with KP-Sil silica, Biotage recommends a solvent flow rate between 100 and 250 mL/min for Flash 75 and between 400 and 1000 mL/min for Flash 150. Note: Ensure that the solvent pressure is at least 20 psi...
13. Figure 26. The O-ring at the top of the radial compression barrel. 12. If the Flash 75/150 system is not used for a longer period of time or you want to change solvent before the next run, please empty the solvent tube that is connected to the solvent reservoir.
(5.5 to 6.9 bar). 2. Safety relief valve is defective. 2. Replace the safety relief valve (P/N 00790 for Flash 75 and P/N 02196 for Flash 150). 3. Safety relief valve is not properly 3. Remove and re-apply PTFE tape, and then re-install using the installed.
The following section is a summary of work completed by The following sections address the two parameters used to Biotage application chemists in an effort to reduce the amount optimize ΔCV; solvent strength and solvent selectivity. of trial and error that is typically associated with optimizing flash separations.
Optimizing Flash Separations Vary Solvent Selectivity to Maintain In Figure 29, the R of the more polar component was changed by replacing ethyl acetate with dichloromethane and resulted in Values Between 0.15 and 0.40 the upper (i.e. less polar) spot being essentially unchanged. Frequently, the R of one component can be changed relative In some instances, varying the polar solvent component can...
Load is Dependent on ΔCV Table 8 shows recommended sample loads for different length Flash 75 and 150 prepacked cartridges. In general, the more difficult the separation (i.e. a low ΔCV), the less material can be loaded onto the cartridge for a successful/high resolution separation.
The Biotage Flash Chromatography Workbook has been designed to document the separation steps taken when using a Biotage Flash system to purify synthetic compounds. Upon the completion of a separation using one of the prepacked cartridges from Biotage, you'll be able to evaluate the performance of the Biotage approach versus flash separations using traditional glass columns.
Page 29
Flash Chromatography Workbook TLC Method Optimization Solvents Used: Concentration: Elution Calculations: Distance from the origin to the center of the spot CV= 1/R Distance from the origin to the solvent front...
General Information General Information Price List and Ordering Information Please contact your Biotage distributor for the latest price and ordering information. Download User Documentation You can download the latest version of this document and other user documentation at www.biotage.com. If you have problems downloading, please contact your local Biotage representative.
Need help?
Do you have a question about the Flash 75 and is the answer not in the manual?
Questions and answers