Introduction Introduction Biotage Flash 400 is a flash purification system based The Flash 400 system is designed and manufactured to meet ® on prepacked cartridge technology that provides users of the requirements of the EC machinery directive and the ATEX...
100 psig (6.9 bar). EPDM is resistant to polar Figure 2 . Flash 400 system. A = hoist, B = hoist control box, C = cartridge solvents but will swell in many common aliphatic and aromatic...
Page 5
4 = cartridge, 5 = O-ring “A” size 222, Chemraz, 6 = O-ring “B” size 223, Chemraz, 7 = top head assembly, 8 = cartridge seal adapter (adjustable), 9 = seal adapter housing (fixed), 10 = flow distribution plate (inside the cartridge), and 11 = frit. Flash 400 Part...
(PSV-03) that activates at pressures in excess of 125 psi (8.6 bar). Note: Biotage recommends using the same solvent for both the eluent and the radial compression fluid. Figure 5. The cartridge compression vent valve (MV-03) and the top vacuum break valve (MV-04) closed (left image) and opened (right image).
MV-01 MV-06 pressurized. PG-02 and PG-03 readings will equalize. Closes the valve. BIOTAGE FLASH 400 ® MV-08 Solvent pump valve. Turns the solvent pump on or off (O or I) by opening or closing the line between the air supply and the pump.
Ensure that the cartridge tool cannot be moved within the cartridge top. If the tool can be moved, loosen the knurled knob Biotage provides a start-up kit consisting of the following items: and repeat the tightening process to re-engage. Move the hoist arm so that the safety hook is centered over the Description cartridge tool and then fasten the hook to the tool’s eyebolt.
The bodies are stainless steel with Gauges, and Regulators glass lenses. All valves used on the Flash 400 system are fabricated from 316 For details about each component, see Figure 14 below, stainless steel with a PTFE seat and rated by the manufacturer Table 7 on page 9, and the Flash 400 flow diagram located to at least 200% of the maximum system working pressure.
Page 11
(PG-03) will be equal to the reading on the cartridge compression pressure gauge (PG-02). reservoir, which compresses the cartridge if the fluid reservoir is pressurized. PG-02 and PG-03 Table 7. Flash 400 valves, gauges, and regulators. readings will equalize. Closes the valve. MV-08 Solvent pump valve.
Biotage. » Within limits set by the system’s technical specification and Only genuine Biotage spare parts must be used in the system. in line with user standard operating procedures (SOPs). Failure to follow those instructions and operate within the limits...
» The total weight of the package including the system is up to system is used in a manner not specified by Biotage, the 1043 kg (2300 lbs) depending on the system configuration. safety features of the system may be compromised.
Page 15
If any problems are the general atmosphere, contact Biotage for instructions found, solve them before using the equipment. »...
» Recommended Tubing Solvent Pump: Connect a compressed air supply to the Biotage recommends stainless steel or brass tubing or air port on the side connection panel (see Figure 16). stainless steel overbraided PTFE tubing for the gas supply Set the pressure to a maximum of 100 psig (6.9 bar).
= Solvent inlets on the side connection panel. Figure 18. The drain tube connected to the radial compression drain valve (MV-10) on the side connection panel is located underneath the Flash 400 frame, on the right side. Fraction Outlets (1 and 2)
Only use clamps, studs, nuts, and and the vacuum break valves (MV-04 and MV-05) are open; washers supplied by Biotage. see the instructions on page 27. 5. Remove the top head clamp by loosening the nuts on the sides 2.
Page 20
11. Slowly move the head assembly while attached to the hoist toward the front of the Flash 400 frame until it is completely 7. Move the hoist arm so that the safety hook is centered over clear of the frame and the radial compression module.
Page 21
Operation 2. Move the cartridge shipping container holding the new cartridge next to the hoist. 3. Open the container and pull the plastic shipping bag away from the top of the cartridge. 4. Remove the sealing label at the top of the cartridge; see Figure 26.
Page 23
O-rings (see Figure 3 on page 3) and fulcrum arm; see Figure 30. retry, or contact Biotage 1-Point Support. 16. Inspect the head clamp. If the clamp, threaded studs, nuts, or washers show any sign of wear or damage, replace it/them.
(MV-04) on the top head assembly; compression fluid. see Figure 35. Note: Biotage recommends using the same solvent for both the eluent and the radial compression fluid. 4. Install the lid on the radial compression fluid reservoir by tilting and lowering it into the mouth of the reservoir.
Operation 4. Ensure that the bottom vacuum break and radial 12. Check for audible leaks, then turn the reservoir pressure compression drain valves (MV-05 and MV-10) on the side valve (MV-09) on the control panel to the position connection panel are closed. in-between vent and pressurize.
Flash 400 Cartridge ® d. Check the gas pressure at the outlet 1 port and compare All new cartridges shipped from Biotage are guaranteed to it with the starting pressure reading. If the pressure perform according to specifications. Testing is optional, and is within 2 psig (0.1 bar) of the starting pressure, the...
Operation Equilibrate the Cartridge Load the Sample Warning Warning » » Always equilibrate the cartridge. Take appropriate measures to protect against static discharge » by ensuring that the system and ancillary containers are Equilibrate the cartridge at a low flow rate to avoid extreme grounded (see page 14).
Operation Depressurize and Drain the System Warning » Only use waste containers that are grounded. » Ensure that the pump pressure gauge (PG-01) reads zero (0) bar/psig before disconnecting the stainless steel overbraided solvent inlet tube from the three-way injection valve. Note: Refer to Figure 14 on page 8 for component locations when performing this procedure.
Biotage safety and performance warranties. In addition 5. Inspect all O-rings. Ensure that they are not cut, flattened, this will yield a poorly performing cartridge and may cause or otherwise damaged.
Clamp before the lubricating solvent evaporates. 2. The cartridge is too long. 2. Only use cartridges supplied by Biotage and of a length that corresponds to the radial compression module in use. 3. Bottom head assembly is out 3.
Set the nitrogen gas supply pressure to between 80 and 100 psig (5.5 to 6.9 bar). 2. Safety relief valve is defective. 2. Contact Biotage 1-Point Support; see contact information on the back of this document or visit our website www.biotage.com. Gas leaks around the top of 1.
US 6,139,733, US 6,294,087, US 6,221,252, and US 6,436,284. Price List and Ordering Information Please contact your Biotage distributor for the latest price and ordering information. Download User Documentation You can download the latest version of this document and other user documentation at www.biotage.com.
Appendix Appendix Warranty and Liability Biotage Flash 400 Spare Parts ® See the “Biotage Terms & Conditions of Sale” document at Part No. Description www.biotage.com. 03020 Barrel O-Ring (D), 387 Viton Essential Tools List 03019 Barrel O-Ring (D), 387 EPDM...
Need help?
Do you have a question about the Flash 400 and is the answer not in the manual?
Questions and answers