HemoCue 201 DM Reference Manual page 189

Analyzer, docking station
Hide thumbs Also See for 201 DM:
Table of Contents

Advertisement

Symptom
Explanation
No transfer of
1) No network communication
data
2) Communication error.
3) No communication.
4) No communication between
5) No communication between
6) No communication with any
with OR or PC.
the first and second Docking
Station.
the Secondary Docking
Stations.
Docking Station.
Action
1a) Check that the LED on the
Docking Station does not show a
flashing red light.
1b) Check the network configuration
of PDS/PDS
1c) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
SDS
+
Stations.
1d) Check the external set-up and
settings (OR/PC, firewall, network
infrastructure).
2a) More than 5 Docking Stations are
docked together.
2b) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
SDS
+
Stations.
3a) Check that a USB connection is
not used for communication in a
Primary Docking Station.
3b) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
SDS
+
Stations.
4) Check that two Primary Stations
are not docked together.
5a) Check the electrical connections.
The LED on the docking station
should give either a stable green
light or a flashing green light.
5b) If multiple Docking Stations are
used, make sure not to mix
Docking Stations labelled PDS
SDS
+
Stations.
6) Check the electrical connection.
The LED on the Docking Station
should give either a stable green
light or a flashing green light.
If problems remain contact
distributor.