Type 'Bf' Applied Parts; Frequently Used Functions; Carry Bar - Prism CP User Manual

Portable hoist
Hide thumbs Also See for CP:
Table of Contents

Advertisement

4.0 Type 'BF' Applied parts

Below shows the two parts of the hoisting system, which are classed as Body Floating (BF) applied parts. The carry bar is a
complete assembled unit which allows approved Prism slings to be attached, to lift and assist patient. See section 5.1 for
instructions to attach carry bar to Hoist system and section 5.2 to attach an approved sling to the carry bar. To see Prism
Medical UK approved sling list see table 5 and 6.

5.0 Frequently Used Functions

5.1 Carry Bar

Attach the carry bar (type 'BF applied part) into the hook on the lift tape, located at the end opposite to the hoist, in the
following way:
1. On the hook, move the locking mechanism into the hook by pressing down on the tab (Figure 6 and 7).
2. With the carry bar positioned sideways along the length of the bar, move the pin at the top of the carry bar into the
hook (Figure 8).
3. Alternatively, with the carry bar positioned sideways, the pin at the top of the carry bar can gently push the locking
mechanism out of the way as the pin is carefully moved into the hook.
4. Once the pin is in the hook, rotate the carry bar 90° down so the carry bar is hanging below the hook on the pin in the
hook. Move the locking mechanism into place by pushing up on the tab, securing the carry bar on the hook and lifting
tape. (Figure 10 and 11).
Figure 6
Figure 9
999096 – Revision A
Black Carry Bar
White Carry Bar
Figure 7
Figure 10
Slings
(Reference image)
Figure 8
Figure 11
P a g e