CO N TE N TS Inspection Contents of Box Safety Symbols Technical Data Material Unit Instructions Initial Set Up / Function Check General Use Storage and Handling Error Messages on Control Panel Glass Replacement UV Ballast Replacement Pump Occlusion Heating Element Replacement Software Upgrades Replacement Parts / Service Kits Warranty/Replacements...
300.107 Power Cord (Euro) REF 300.154 Power Cord (UK) REF 300.155 Power Cord (AU) REF 300.156 (do not use a detachable mains power cord with inadequate ratings to the Strava cord) 8 ft Drain Hose REF 200.101 Strainer REF 300.128 Priming Bulb REF 300.127...
SYM BO LS Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. See the information below for definitions of the signal words. This symbol designates where personal injury or damage to the equipment is possible. DANGER! Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
• is not allowed and constitutes improper use. Direct application on the patient is not allowed. • T EC H N ICA L DATA Technical Specifications Strava and Rolyan Splint Bath SPLINT BATH STD Spec Operating Temperature Range - 180...
Material The outer chassis material is a thermoset composite material. The unit consists of a tempered glass lid, drainage hose, water strainer, water pump, power supply and ultraviolet (UV) light. Control Panel 1 – Targ e t te mp e ra ture di sp l ay 2 –...
GALLON OF DISTILLED. UNIT WILL NOT FUNCTION PROPERLY IF DISTILLED WATER IS EXCLUSIVELY USED. Initial Set Up/Function Check Plug the Strava supplied power cord (do not use any other power cord other than supplied by Strava) into the back of the unit and a dedicated wall receptacle. (Protective Earthing - a dedicated electrical circuit is required).
NOTE: DO NOT USE SANITATION WIPES ON THE GLASS LID OR THE CONTROL PANEL FACE PLATE. • Strava recommends a periodic descaling of the unit to remove any potential mineral build up on the heating element, UV system and internal water pump. Run a commercial descaling solution through the unit as per directions on the product (i.e Mulituse CLR, Zep, Keurig descaling solution,...
• Check the power cord and plug for excessive wear or fraying. Replace as needed. Do not use a detachable mains power cord with inadequate ratings. Only replace with a Strava power cord. It is recommended the following guidelines are used whenever this system is being stored...
Glass Replacement: If the unit suffers broken glass anywhere on the lid, the entire glass lid needs to be replaced. A glass lid assembly is needed to perform this operation. Glass Lid Assembly Kit: REF 200.303 Kit consists of 2x glass panes with center hinges, handle and 4x male sex bolts, 8x nylon washers and 4x acorn nuts, 3mm allen wrench, and 7mm nut driver.
UV Ballast Replacement: If the UV bulb has been replaced and the UV system is still not working, the next step is to replace the UV ballast. A UV Ballast Assembly Kit is needed to perform this operation. UV Ballast Assembly Kit: REF 200.301 Kit consists of UV ballast, bulb harness and two wire ties.
Page 12
UV Ballast Replacement cont. : Step 3: Install the new UV ballast • Remove new ballast assembly from box. • Feed the wire harness through the back panel. • Line up the black harness to the top of the control panel and wire tie in same two locations as removal.
Pump Occlusion: If the unit is properly primed yet continues to make a humming or screeching sound, the likelihood is that there is a clog in the pump. Check if water is coming out of the port. If there is not water circulating through the priming port, check for an occlusion.
Page 14
Pump Occlusion cont. : • Center the provided pressure-sensitive foam pad between the wire ties on the pump mounting bracket. Figure 44 & 45 • Place new pump on top of the foam pad and tighten with the wire ties, aligning them on top of the foam strips of pump (important: be certain that the wire tie buckles are acclimated to the side of the pump –...
Heating Element Replacement: If the unit no longer heats the water, and it has been determined that the water sensor is functioning properly, the heating element may need replacing. A Heating Element Assembly Kit is needed to perform this operation. Heating Element Assembly Kit: REF 200.303 Kit consists of heating element, brush, sample size Flex Seal, 11/4”...
Heating Element Replacement cont. : Step 3: Reassemble the unit • Fold the back panel into place, aligning the screw holes on either side. Figure 66 • Re-insert the 2x 3/32” button head screws into the back panel. • Re-install the base pan by aligning it with the screw in embossments and the back panel.
Under normal use and following all guidelines defined in this manual, Strava splint baths are warrantied for a period of one year from when product is received (two years outside the USA). Should replacement or repair be required, purchaser must notify dealer or manufacturer in writing.
Page 18
MANUFACTURED BY: Strava Sol ut i ons, LLC 2 73 3 Kanasi t a Dr i ve, # 125 H i xson, TN 37343 U SA TEL: 1. 42 3 .313.6157 EMA IL: i nfo@ ki sm et us.c om te chni c als ervic e@ kisme t us.