Preparation and Use

3-1 Assembly

You can connect the 9106 and 9108 with several light sources.
To use without a camera:
1.
Insert the cable into a light source and look through the eyepiece
(an adaptor may be required). When using the PENTAX Medical
Laryngeal Stroboscope, Model 9400, the 7175-4200 adaptor
provided with the rigid Laryngostroboscope should be used.
2.
Use the brightness adjustment on the front panel of the PENTAX
Medical Laryngeal Stroboscope, Model 9400, to control the amount
of light to the rigid Laryngostroboscope.
To use with a C-mount camera and attachment lens (e.g., with
PENTAX Medical's 3-CCD HD Camera, Model 9214HD, and Lens
Coupler, such as Model 9118B):
1.
Screw the attachment lens into the C-mount of the camera.
2.
Insert the rigid Laryngostroboscope eyecup into the lens coupler.
3.
Align the eyecup and adaptor, and lock them into place using the
screw.
Caution
You must loosen the screw prior to removing the adaptor from
the eyecup in order to avoid damage to the rigid
Laryngostroboscope.
Note To use the 9106 or 9108 with other light sources, adaptors
may be required. Please contact your local representative for
information.