You will have to wait for it to fall close back to zero before giving it gas
again to set the monitor off.
Address menu walk through by hitting the menu button after each mode.
1. "ADR/001" – This is the address mode. It shows you what address your
head is set to. You never want multiple heads set to the same address on a
single location/monitor. You get to this mode by hitting the test button
until you see "ADR" show up.
2. "BRD/007" – This is the Background mode. This displays what the
background is set to. we recommend it to stay at 7 for most of our heads.
This means that when gas goes above 7ppm it constantly is transmitting
that reading to the monitor.
3. "bAt/08.7" – This is the Battery mode. This shows you your current battery
voltage for that head.
4. "GASTYPEorOFF*" – For example "H2S" would be displayed. This is the
Mode in which you select the gas type(H2S/CO/LEL/SO2) you are using or
you can turn it OFF by selecting OFF.
5. "000" – This should be the normal operating mode. In this mode if you hit it
with gas it should climb up to the correct ppm and set the monitor off.
6. "null/000" – This is the zero mode. If you have a floating value of 1 or 2
ppm this is the mode that you can Re-zero your head by clicking the up or
down buttons.
7. "CAL/000" – This is the Calibration Mode. You will give it Cal Gas in this
mode and when it quits climbing higher you will hit the up button and it will
set it to what your cal limit is. This mode will not set your monitor off. If you
call it up to 25 ppm and exit this mode it will not set the monitor off either.