TECHNICAL SUPPORT For more IAC200 support, please visit: https://www futek com/summing-junction-box Drawing Number: EM1006-A FUTEK reserves the right to modify its design and specifications without notice Please visit https://www futek com/TermsAndConditions for complete terms and conditions 10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 www.futek.com...
For example, four 100lbs sensors are adjusted to have the same 1 98mV/V Solutions are available at FUTEK to connect your IAC200 to the when100 lb is applied The final summed output will be 1 98mV/V at...
If present, for amplified IAC200, disconnect the ribbon cable Set the output of the IAC200 for S (mV/V) for direct sensor from the center amplifier to the IAC200 board on connection J6...
IAC200 Quick Start Guide Wiring Connect each sensor to the IAC200 +/- Excitation will be power Connect power supply to IAC200 for sensor excitation Refer to to the sensor and +/- Signal will be the output from each sensor sensor spec sheet online...
IAC200 Quick Start Guide Measurement Place a known load on the first sensor Load does not have to be Adjust Channel Select for next channel ON and switch OFF prior the full capacity of the sensor channel Record mV output on Voltmeter Full capacity mV output level is...
IAC200 Quick Start Guide Adjustment Observe lowest recorded output from connected sensors Set Channel Select DIP switches to first Channel ON and all other Channels OFF Serial No. Output Apply known load to the first sensor Adjust SPAN potentiometer to adjust sensor mV output down,...
FUTEK online support page of the IAA amplifier for further instructions on how to set up the attached Set the output of the IAC200 to S (mV/V) for non-amplified amplifier IAC200, V (VDC) for voltage amplified IAC200 and I (mA) for •...