Example: 150 Volts was set at the beginning, 15 volts will be supplied when
gel saver function will start to work after timing mode.
• When "GEL SAVER" is working, the LED (#17 ) blinks. 5 beeps are heard
every minute.
• To stop the "GEL SAVER" mode, press the stop switch (#9).
• It is possible to cancel the "GEL SAVER" option during the run by simply
pressing the switch (#16). The "GEL SAVER" LED (#17) turns off.
• It is possible to select the "GEL SAVER" mode during the run by simply
pressing the "GEL SAVER" switch (#16). The LED (#17) turns on.
4. Adjusting Output set values. Constant Voltage
• Follow the same instructions described on page 20, paragraph 2.
5. Adjusting Output set values. Constant Current
• Follow the same instructions described on page 20, paragraph 3.
6. START Mode
• Press START, the LED (#10) lights and the actual values are displayed
on each display, left for voltage, right for current or minutes.
22
• To read minutes instead of mA or vice versa, press the TIME switch (#15)
or VOLTS/mA switch (#2).
• The actual constant mode is indicated by one of the LED (#5 or #6).
• When set time has elapsed, the unit will turn on "GEL SAVER" option
if this option has been selected, or in STOP mode with an audible alarm.
• If this option has been selected, and when set time has elapsed, the unit
turns automatically to constant voltage and supplies 10% of the voltage
set at the beginning of the run,
- The mA/minute display (#3) shows the elapsed time,
- The voltage display (#4) shows a value, indicating that 10% of the set
voltage is supplied into the electrophoresis tank.

7. Changing the settings

• Follow the same instructions described on previous page 21, paragraph 5.