Ibi SH-300 XL User Manual

Electrophoresis power supplies

Advertisement

Quick Links

User's Manual
Electrophoresis Power Supplies
Please retain all packaging materials until
the warranty period has expired.
XL
SH-300
P. 7
XL
SH-500
P. 15

Advertisement

Table of Contents
loading

Summary of Contents for Ibi SH-300 XL

  • Page 1 User’s Manual Electrophoresis Power Supplies SH-300 P. 7 SH-500 P. 15 Please retain all packaging materials until the warranty period has expired.
  • Page 3: Warranty

    All claims made under this warranty must be presented to IBI SCIENTIFIC within one year following the date of delivery of the product to the customer.
  • Page 4: Safety Precautions

    Take care as the mode of operation of the unit is continuous. If the power supply is used in a manner not specified by IBI SCIENTIFIC, then the protection systems of the equipment may be impaired. For additional information, please call IBI SCIENTIFIC or your distributor Technical Resources Department.
  • Page 5: Maintenance

    A declaration of contamination status form is attached to this manual and must be filled out prior to inspection servicing repair or return. Please request a copy from IBI SCIENTIFIC or distributor and include with the equipment prior to return.
  • Page 6 NOTES __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________...
  • Page 7 User’s Manual SH-300 Electrophoresis Power Supply Please retain all packaging materials until the warranty period has expired.
  • Page 8 Microprocessor controlled switching Power Supply a Last settings are automatically restored at power up. a IBI SCIENTIFIC Power Supplies are equipped with an automatic restart system in case of power failure. When the power returns, an audible alarm sounds for 10 seconds and the power automatically restarts with the previous set values.
  • Page 9 SH-300 XL SPECI FICATIONS Power supply voltage fluctuations not to exceed ± 10% of the normal voltage Power supply, 110V model 90 - 130V; 50 - 60Hz; T4A fuses Power supply, 220V model 180 - 260V; 50 - 60Hz; T2A fuses...
  • Page 10 1 Tactile switch for the selection of the following modes : - OUT: Display of output values when the unit is running - SET: Display of preset values 2 Tactile switch for selection of parameter to set. Operating in SET mode - Volts: Voltage - mA: Current...
  • Page 11: Operation

    (#8). 4. Press SET/OUT switch (#1) for displaying the new actual values. Maximum output values are as follow: SH-300 XL for 300 Volts - maximum current 400 mA (120 Watts) for 400 mA - maximum voltage 300 Volts (120 Watts)
  • Page 12 C – Adjusting output set values when unit is in START mode 1. While the unit is running, depress the SET/OUT switch (#1). As soon as this switch is activated, the output LEDS display last output set values. 2. The volts LED (#5) is blinking at first. Increase or decrease voltage by pressing of the two SET switches (#8).
  • Page 13 Press STOP to resume and to cease the audible alarm. a Look for the cause of the fault and correct. a Press START to run again. a If the unit goes to the fault mode, contact IBI SCIENTIFIC or your supplier technical service. AUTOMATIC RESTART...
  • Page 14: Troubleshooting Guide

    13. WARNING Never attempt to remove the outer casing or make any unit repairs. Contact IBI SCIENTIFIC if the need for repair or servicing should arise. Should the power supply fail, DO NOT remove the outer case of...
  • Page 15 User’s Manual SH-500 Electrophoresis Power Supply Please retain all packaging materials until the warranty period has expired.
  • Page 16 SH-500 XL SPECI FICATION Power supply voltage fluctuations not to exceed ± 10% of the normal voltage Power supply, 110V model 90 - 130V; 50 - 60Hz; T4A fuses Power supply, 220V model 180 - 260V; 50 - 60Hz; T2A fuses Fuse value in the mains plug (UK) Rated input power/current 150 VA...
  • Page 17 2 8 14 6 15 9 16 1 Tactile switch for the selection of the following modes: - OUT: Display of output values - SET: Display of set values 2 Tactile switch for selection of parameter to set: - V: Voltage - mA: Current...
  • Page 18 FU RTH ER I NSTRUCTION Important to read before operation ! If the unit alarms at power up, press the STOP switch. This situation is caused by a previous run stopped by the power switch instead of the STOP switch. Never use the power switch to stop separations.
  • Page 19 Press STOP to resume and to cease the audible alarm. a Look for the wrong situation. a Press START to run again. a If the unit returns to STOP mode, contact IBI SCIENTIFIC or your supplier technical service. AUTOMATIC RESTART...
  • Page 20: Unit Power Up

    OPERATION Without timer mode 1. Unit power up • Connect the AC line cord to a grounded, 3-prong wall outlet. • Connect the unit to the electrophoresis device using the safety connecting leads supplied with the electrophoresis device. • Turn on the power switch located on the rear panel. •...
  • Page 21: Timer Selection

    5. Changing the settings when unit is working It is possible to change the settings when the unit is working. Proceed as follow: • It is not necessary to press the STOP switch, leave the unit in START mode. • Press SET/OUT switch (#1) to select the SET mode. •...
  • Page 22: Changing The Settings

    Example: 150 Volts was set at the beginning, 15 volts will be supplied when gel saver function will start to work after timing mode. • When “GEL SAVER” is working, the LED (#17 ) blinks. 5 beeps are heard every minute. •...
  • Page 23: Other Settings

    Operation with timer, no “GEL SAVER “ mode 1. Timer selection • Press ON/OFF switch located in the timer area (#12). • The LED TIMER (#13) lights, showing that timer mode is selected. • Press the Gel saver switch (#16) until the LED (#17) turns off. •...
  • Page 24: Trou Bleshotti Ng Gu I De

    TROU BLESHOTTI NG GU I DE CONDITION PROBABLE CAUSE REM EDY Display fails to illuminate Fuses have blown. Contact IBI SCIENTIFIC when the POWER switch servicing. is put on. The desired MODE is not One of the other Increase the output set flashing.
  • Page 25 NOTES __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________________________________...

This manual is also suitable for:

Sh-500 xl