Select measuring Unit (Gauss – mT – A/cm – kA/m) ..............13 Automatic Switch-Off ......................13 Display battery voltage ......................14 Reset ............................14 Applicable Measuring probes ........15 Checking MP-2000 with Calibration Standard ..... 16 Using the analog probe P-T4A Analog-Output .... 17 Technical Data ............. 18 List-Magnetik GmbH EN MP-2000 2022-03.doc...
Page 4
Optional ............... 19 TOP-PRINT4 Thermal Data Printer ........... 19 Available Applications ..............20 MP-2000 TRANSFER ......................20 TRANSFER ..........................20 List-Magnetik GmbH EN MP-2000 2022-03.doc...
They will be pleased to assist you. With the magnetic field meter List-Magnetik MP-2000, we offer you a first class, high-end product, with externally connectable axial and tangential field probes. With the easy to use device, you can precisely measure magnetic DC and AC fields as well as pulse fields of all kinds.
EASURING DUMMY AFETY NSTRUCTION If you are using the MP-2000 with a special measuring dummy instead of a probe for the measurement in a magnetizing system, be sure to observe the following safety instructions: The complete measuring unit with MP-2000, measuring...
TART • Connect the probe cable to the measuring probe and device. • Switch on the MP-2000 with the ON-OFF key. Automatic zero balancing takes place when the device is switched on, therefore the probe should not be in a magnetic field when doing so. The last measured value appears briefly on the display, followed by the current measured value.
Switch to DC Auto Range in order to measure static fields. The probe has an internal measuring range switch at 600 A/cm. For the user, this internal automatic switching has no effect. The DC symbol is shown behind the current measured value on the display. List-Magnetik GmbH EN MP-2000 2022-03.doc...
The peak value memory is reset via the -0- key. When measuring AC fields in peak value memory mode the peak value and not the effective value will be saved if the AC field is sinusoidal. List-Magnetik GmbH EN MP-2000 2022-03.doc...
(min. value), after which the analogue display is shown in the center on the display with the preselected start of the measuring range. The maximum measuring range value is automatically adjusted according to the size of the current measured value. List-Magnetik GmbH EN MP-2000 2022-03.doc...
Display / Select Batch Display Delete Batch >>>> Max.Batch-No. >>>> Batch Print+PC Numbers 1000 Setup After selecting a maximum number of batches, the maximum number of measured values that can be saved per batch is displayed. List-Magnetik GmbH EN MP-2000 2022-03.doc...
This function is only used to supply the data application TRANSFER. The applications TRANSFER-EXCEL and STAT6 don’t work any longer with firmware version 15.1 and up. To transfer data to PC, switch to the data transfer application MP-2000 TRANSFER at https://www.list-magnetik.com/software DC/AC/PEAK Transmit to PC >>>>...
The device will automatically turn off 2 minutes after the last measurement if no measurement value change occurs. This automatic is pre-set. It can be deactivated e.g. when the device is used in a production line. List-Magnetik GmbH EN MP-2000 2022-03.doc...
Reset returns all device settings to the factory settings. This function should be used if settings have been changed and the device does not work properly, or if the calibration of the probe does not work properly. List-Magnetik GmbH EN MP-2000 2022-03.doc...
Measuring range up to 20,000 A/cm Axial probes P-A2 and P-W2 (angular axial probe) Tangential probes P-T2 and P-Z2 (flexible reed probe) Additionally, in measuring range up to 40,000 A/cm Axial probe P-A4 Tangential probes P-T4, P-Z4 and P-T4A Analog-Output List-Magnetik GmbH EN MP-2000 2022-03.doc...
In the slot for the tangential probe, it can even be damaged. Instead, you can obtain a calibration standard with positioning point, which unfortu- nately works less accurately, but does not damage the probe. List-Magnetik GmbH EN MP-2000 2022-03.doc...
Threshold +0,1 mT Max +80 mT = +0,90 V DC High Range / Peak High Range -5140 mT = +2,70 V Threshold -1 mT 0 mT = +1,80 V Threshold +1 mT +5140 mT = +0,90 V List-Magnetik GmbH EN MP-2000 2022-03.doc...
Power supply: 3x 1.5 V AA Mignon Operating time: approx. 50 hours Dimensions: 200 x 95 x 41 mm Weight: 265 g with batteries Warranty: 12 months on the device, 3 months on the probe List-Magnetik GmbH EN MP-2000 2022-03.doc...
Page 21
2. Faulty printout: The TOP-PRINT4 must be charged up again if individual lines of the printout are not printed correctly. 3. Inserting a new paper roll: Open the cover, Insert the paper roll, pull out the end of the paper, Close the cover. List-Magnetik GmbH EN MP-2000 2022-03.doc...
Page 22
MP-2000 TRANSFER, to transfer measurement data to your PC. With MP-2000 TRANSFER, you can measure online or read the device’s memory, you can evaluate the data statistically or visualize as chart. You can print the results or hand over the data to applications like Microsoft Word and Microsoft Excel.
Page 23
We supply: • Coating Thickness Meters • Magnetic Field Meters • Magnetic Permeability Meters • Magnetizing and Demagnetizing Equipment We advise and provide tailor-made solutions for your specialized requirements in magnetizing, demagnetizing and measuring Fast calibration and repair service...
Need help?
Do you have a question about the MP-2000 and is the answer not in the manual?
Questions and answers