Operating Device In Automatic Mode; Switching To Automatic Mode; Switching Back To Manual Mode; Operating Device In Manual Mode - id LAB FOTOMETER 2008 COOL User Manual

Light attenuation meter
Table of Contents

Advertisement

Operating device in automatic mode

In automatic mode the device switch to more sensitive range, when intensity value decrease
below 8 percent of current range. When intensity rise over 100 percent of current range,
device automatically switches to less sensitive range.
During switching, the device displays SWITCHING message on the display. During
this period the intensity readout is not valid. On some conditions (slow filter selected, large
sudden change in intensity) the intensity might not be settled when SWITCHING message
disappear and additional waiting for valid value might be necessary.

Switching to automatic mode

While in manual mode, press AUTO/MANUAL button to switch to automatic mode.

Switching back to manual mode

While in automatic mode you can press AUTO/MANUAL button to switch back to manual
mode. By pressing RANGE+, RANGE- or DIAG while in automatic mode you switch back
to manual mode as well.

Operating device in manual mode

In manual mode, ranges must be switched manually by the user. When range is manually
switched, device displays SWITCHING message and goes through the switching procedure in
same manner as in automatic mode.

Switching to manual mode

While in automatic mode you can press AUTO/MANUAL button to switch to manual mode.
By pressing RANGE+, RANGE- or DIAG while in automatic mode you switch to manual
mode as well.

Switching back to automatic mode

While in manual mode, press AUTO/MANUAL button to switch to automatic mode.

Switching to less sensitive range

To switch to less sensitive range press RANGE- button.

Switching to more sensitive range

To switch to more sensitive range press RANGE+ button.

Using references

To compute light attenuation directly during measurement there are two user references
available. When reference is set it displays the ratio between reference value and current value
in a logarithmic scale in dB. The value in logarithmic scale is computed using following
formula:
=
Log
value
[
dB
]
10
Intensity
value
 
log
Re
ference
value
[
AU
]
 
[
AU
]