Turbine maintenance and spray
Uncouple turbine from hose. Remove mill or diamond bur.
Clean up the external part using 70% hydrated ethyl alcohol or any other disinfecting agent without having ammonium
or acid basis or chlorine.
Do not immerge the handpiece in disinfection baths.
Lubricate the turbine before each sterilization cycle. Handpieces which are not subject to a sterilization process must be
lubricated before each work cycle (in the morning, at noon, and evening).
Improper care and lack of maintenance may lead to a premature wear and failure of the handpiece. Sterilize the
handpiece and bur changer in autoclave up to 135 C. Autoclave for 20 minutes at 121 C and 15 minutes at 135 C;
withdraw handpiece from autoclave after the sterilization process is over.
Cleaning
Lubrication
Problem
Noisy
Light is dim or does
not emit light
Poor power
Cannot hold the bur
Leakage
Connections with
chair is not proper
If the above methods do not resolve the problem, please contact the manufacturer.
Cartridge Replacement
1. How to remove a cartridge:
Mount a dummy bur in the chuck. Turn the head cap counter clockwise with the head cap wrench, and remove the head
cap. Push up the dummy bur, and the cartridge is easily removed from the head.
2. Mounting cartridge
Align the small pin in the cartridge with the slot in the handpiece head, and insert the cartridge into the head. Mount and
screw in the head cap with fingers till tight. Further tighten with the head cap wrench.
Cautions
The threads in the head are very ne. Do not turn the head cap with the wrench from the rst turn. Use of other company's
cartridge may void the warranty of the handpiece.
Warranty
1. This product is warranted against manufacturing defects for 6 Month from the date of effective delivery of the product.
The cartridge is warranted for 90 days while bearing is excluded from warranty.
2. The warranty is limited to repair or replacement of parts having manufacturing defects duly evidenced by the
manufacturer or by the authorized technical assistance network.
3. This warranty does not include the repairing of faults due to non-observance of use and maintenance instructions,
tumbles or beats, improper storage, action of natural agents and technical assistance carried out by non- authorized
personnel.
4. This warranty would cease as per the normal course of the validity terms, if alterations are introduced in the product or
use of improper fittings and if technical assistance is carried out by non-authorized personnel.
5. The repair or replacement of parts within the warranty period will not extend the original validity terms of the warranty.
6. This warranty is limited to the repair or replacement of defective parts only.
7. The parts replaced during the warranty period belong to the company and users must not claim to retain the damage /
defective parts.
8. Charges resulting from transport of parts of the product will be the customer's responsibility.
9. This warranty is valid in the country where the sales documents were permitted.
TM
Clean the spray hole
with a wire
2-3 Sec.
Exceptional issues and their solution
Reason
1. Bearing damaged
2. Curved bur
3. Cap without cushion
1. The generator or LED is damaged
2. Un-used for a long time
1. Foreign body blocking the hole in the head
2. Chair exhaust pipe plug is not suitable
1. Bur is not standardized
2. Spindle is damaged
1. Connection is not well tightened
2. O- ring is damaged
1. The chair joint is damaged
Clean the spray hole with
3-way syringe
Sterilization
Autoclave Pouch
1. Change bearing or bur
2. Put new washer
1. Change generator or LED
2. Will work if adjusted to 0.4MPa operation
1. Use a file into the air hole to clean
2. Vent the air to the vent hole
1. Change the bur
2. Change the spindle
1. Proper connection should be established
2. Change plug O-ring
1. Change chair connector
Waldent Innovations Private Limited
Disinfection
Use a soft cloth soaked in
disinfectant
Solution