Summary of Contents for Cellbox Flight CDI

  • Page 1 TECHNICAL MANUAL Cellbox Flight CDI...
  • Page 2 Holtenauer Str. 74 Englische Planke 8 24105 Kiel I Germany 20459 Hamburg I Germany T +49 40 226316410 cellbox-solutions.com office@cellbox-solutions.com S O LU T I O N S G M B H Cellbox is a registered trademark of Cellbox Solutions.
  • Page 3 For Research Use Only (RUO) DISCLAIMER The manual is intended as a guideline and that Cellbox Solution GmbH shall in no event be held liable for any direct, or incidental damages that arises out of, or are related to the use of this manual.
  • Page 4: Table Of Contents

    Safe Removal of the Live Cells from the Device Cleaning the Device LABELS ON THE DEVICE Type Plate Labeling requirements for ground transport of the Cellbox Packaging of biological materials ENVIRONMENTAL CONDITIONS FOR THE USE OF THE DEVICE STORAGE AND DISPOSAL...
  • Page 5: Safekeeping Of The Manual

    In order to achieve the optimal conditions for the It is strictly forbidden to use the device for the transport transport of living cells, the Cellbox is able to control the of infectious substances as classified within Class 6.2 ambient temperature around the cell culture vessels with of the ADR (European Agreement concerning the Inter- precision.
  • Page 6: Safety Instructions

    Features The use of the device in potentially The Cellbox Flight CDI has the following features: explosive areas (containing flammable • Protective environment for the temporary stor- substances, gases, fumes or mixtures of gas and air) is strictly forbidden.
  • Page 7: Device Overview

    • Cellbox App for data export and graphical presenta- tion of logged environmental conditions. Device Overview The outer case of the Cellbox is a robust transport box made to withstand mechanical impact. It is therefore the ideal device for the transport of biological material by road, sea, rail and air.
  • Page 8 Tap “ ” to return to the Homescreen. of the incubation chamber is performed with the Cellbox connected to a wall power socket to re- The remaining dry ice level is displayed by a serve the fully charged batteries for transport. Depending timer that counts backwards from 24 hours.
  • Page 9 Tap “ ” to return to the Homescreen. Setup - Energy Saving Mode The Cellbox can record up to 50 logs. The remaining log counter is displayed in the upper right corner. Once 50 log files have been recorded, the Cell- box will be unable to record any further logs and the user will be notified by a pop-up window.
  • Page 10 Mode when the device is already pre-heated and the Cellbox is about to be transported. If a Cellbox is left to To learn more about the function of the Cellbox App, heat up the incubation chamber while Energy Saving...
  • Page 11: Operation Of The Device

    Service OPERATION OF THE DEVICE Charging the Device The Cellbox is fitted with 4 Lithium-Ion battery packs that each have a capacity of 99.4Wh. These battery packs provide an uninterrupted power source for the operation of the device during transport for more than 24 hours.
  • Page 12: Refilling Co2

    Please be aware that more time is required to fully charge the Cellbox when it is in operation. Open the Cellbox and gain access to the dry ice con- The available current is preferentially used to tainer by unscrewing the lid in an anti-clockwise motion...
  • Page 13: Preparing The Device For Transport

    To protect the contents of the Cellbox, the device should only be transported in an upright position. It is strongly recommended that the Cellbox, and more so the incubation chamber, remains closed during transport.
  • Page 14: Safe Removal Of The Live Cells From The Device

    Cellbox. Detailed information about each device can be found on Should the Cellbox not fit, do not force it, simply the type plate located at the back of the Cellbox. rotate the Cellbox until it slides into the polystyrene housing.
  • Page 15 Labels on Cardboard Overpack / Cellbox during Ground Transport with dry ice Cellbox Flight CDI with dry ice Road & Air Labels on cardboard overpack UN 3481 label indicating the use of Lithium-Ion batteries. Please add Cellbox Solutions‘ phone number in the indi- cated field: +49 40 226316410.
  • Page 16 (if transported with/without the overpack) UN 3481 label indicating the use of Lithium-Ion bat- teries. Please add Cellbox Solutions‘ phone number in the indicated field: +49 40 226316410. Information in transport documents (optional) Please add amount of dry ice in kg as well as ship- per‘s and consignee‘s information on UN 1845 Dry...
  • Page 17: Packaging Of Biological Materials

    The Cellbox provides a fourth sturdy outer layer packag- to the manufacturer or distributor. ing. Affix the appropriate label(s) to the outer package, either on the surface of the Cellbox, or on the carton TECHNICAL SPECIFICATIONS overpack. • Dimensions of the Cellbox: Cellbox Solutions GmbH shall in no event be held liable 535 ×...
  • Page 18 CLEANING INSTRUCTIONS General Advice Clean the Cellbox before every use. Wear clean gloves for cleaning. Use an appropriate detergent (e.g. 70% ethanol). Avoid spray disinfection. Immediate Measure Please remove any spills immediately. Regular Measure The device must be switched off and dis- connected from the external power supply before any attempts at cleaning are made.
  • Page 19 EC DECLARATION OF CONFORMITY We, Cellbox Solutions GmbH, Holtenauer Str. 74, 24105 Kiel, Germany, declare under our sole responsibility that this electrical device, Cellbox Flight CDI, is in conformity with the following relevant Union harmonization legislation: Directive 2014/35/EU:2014-02-26 of the European Parliament and of the Council of 26 February 2014 on...
  • Page 20 S O LU T I O N S G M B H Holtenauer Str. 74 24105 Kiel | Germany Englische Planke 8 20459 Hamburg | Germany T +49 40 226316410 office@cellbox-solutions.com cellbox-solutions.com...