Manuale Doc. HBI090-2 Manuale di Installazione e uso EVOSTYLE N.G. Mod. MQI006-0 EVOLUTION LINE INSTALLATION AND USER MANUAL Mod. MQI006-0 Doc. HBE090-3...
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 Index GENERAL ASPECTS...........................1 Introduction ............................1 Descriptions ............................1 Drawings, schematics, components lists, instructions for repairs: ............4 Accuracy Load Factors ..........................6 Other Data ............................6 INSTALLATION AND USE ..........................7 Use Conditions ............................
Page 3
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 Film positioning ....................66...
The unit is manufactured in accordance with the actual international standards on the protection of ionizing radiation, electrical safety, mechanical safety and electromagnetic compatibility for electro medical equipment. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 5
(mm 400-800-1100), has a maximum extension which varies between 1730 mm and 2430, depending on the extension cord used. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 6
The maximum time that can be set is 3 s (the display will read 3000) while the minimum time is 40 ms. (the display will read 0040); NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Drawings, schematics, components lists, instructions for repairs: New Life Radiology is committed to providing, upon request, drawings, circuit diagrams, component parts lists, instructions, or other information that can serve for qualified technical personnel, to perform the repair of any parts that can be repaired.
Page 8
Remote focus skin: 20 cm of the beam at the end of the spacer: 6 cm NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Not invasive method Interposition aluminium filter between rays window and Al 1 mm (AIP99,9 UNI3567) cone collimator: NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Use conditions: Equipment for continuous operation with intermittent charge Installation: Permanent e mobile NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Discard it according to local regulations for waste disposal. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Equipment Serial Number. To be used for all communication with manufacturer / service technician TUBE X-ray Tube Serial number Symbols for transportation and storage NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Please note that the radiological groups are supplied without a plug for leaving the buyer free to connect to equipment already on site. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011.
Page 14
B in the figure above. The lamp must be of value 30W 230-240 Vac. according to the actual reference standard for plants and locals for medical use (IEC 64-8- 710). NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
The system in wall version consists of the components shown in the following legend: wall plate + timer extension double pantograph arm monoblock NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 16
Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 Quotas with relative useful space – EXTENSION FROM mm 400 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 17
Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 Quotas with relative useful space – EXTENSION FROM mm 800 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 18
Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 Quotas with relative useful space – EXTENSION FROM mm 1100 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 20
, and enter it in the wall plate (A), as shown in fig. 3 fig. 3 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 21
Replace the SEGER, unhook the front cover and remove the lower strip cover strips (fig. 4) fig. 4 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 22
Connect the cable from the double pantograph arm to output of THERMOSWITCH and the ground wire IN in its slot. fig. 5 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 23
6 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 24
Unscrew the screw from the front tubular carrier of the double arm pantograph by lifting the cylinder and remove the half-moon(fig. 7) fig. 7 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 25
Mod. MQI006-0 Insert Monoblock in the double pantograph arm as shown in Fig. 8 fig. 8 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 26
Reinsert the half-moon in its slot. Lower the cylinder, securing the double arm pantograph with the screw. fig. 9 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 27
cover, making sure to replace the lower back lining strip and the front cover of the extension. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 28
This procedure is identical for the section B in which the adjusting key must be inserted in the hole 2. hole 1 hole 2 fig.11 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 29
Detail cart vertical support plate holder + timer + button with spiral cord double pantograph arm monoblock NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 30
Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 Quotas with its useful space Fig. 12 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 31
4 screws 6x20 as shown in fig. 14 fig. 14 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 32
+ timer (C) to it using four screws 3,5x6,5 as shown in Fig. 16 fig. 16 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 33
17 Connect the cable coming from the vertical support (B) to the input of the timer, NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 34
(D) to the output of the THERMOSWITCH as shown in figure 18: fig. 18 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 35
The maximum time that can be set is 3 s (the display will read 3000) while the minimum time is 40 ms. (the display will read 0040); NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 36
Following is the symbolism used on the front panel of the timer: fig.19 Button with cable Light emission radiation NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 37
Selection button: Film Type fast Selection button: Film Type slow Selection button: Dose mGy 4-digit display (the exposure time expressed in ms) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 38
At this point the timer is ready for use. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 39
1.23 [mGy] 100 [ms] 100 [ms] 100 [ms] 100 [ms] 0.77 [mGy] 0.77 [mGy] 0.77 [mGy] 0.77 [mGy] NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 40
3.86 [mGy] 250 [ms] 250 [ms] 250 [ms] 320 [ms] 1.93 [mGy] 1.93 [mGy] 1.93 [mGy] 2.47 [mGy] NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 41
During operation, we recommend the use of protective devices such as collars and aprons. To perform the exposure once positioned the cone along the monobloc on the area to be NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011.
Page 42
3 seconds, after which the display will be displayed again the last successful exposure time, in this conditions the timer is ready for a new exposure. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011.
Page 43
Err3 an exposure and key; subsequent cooling action NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 44
If the timer’s display does not turn on, first check that the fuses are intact as shown in fig. 21 FUSE fig.21 (wall / column version) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 45
In this way the system is reset and ready for use. fig. 22 (wall / column version) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 46
If the monoblock does not emit X-rays, as first check the continuity of the THERMOSWITCH; if this test is negative, you must bypass it as shown in fig.23 fig. 23 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 47
Using a tester to check the continuity between the points A and B as shown in figure 24. fig.24 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 48
“END” will appear on the display. At this point, turning off and rebooting the system is necessary. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 49
After removing the back cover, referring to Fig.25 Unsolder the wire from the connector at point A and connect in series a resistance from 1KΩ fig. 25 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 50
1 sec). The value that is expected to read is like in the previous case of about 8 V (to each V corresponds 1 mA) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011.
Page 51
specifications; The tests are stored in digital format and made available for those who request them; NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 52
Mod. MQI006-0 Plates positioning Below, in fig.27 it shows the layout plates positioning on EVOSTYLE N.G. fig.27 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 53
4. S/N XX.XY.XYZ (It represents the device serial number) XX It indicates the year; XY It indicates the month; XYZ indicates the serial sequential number; NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 54
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 (“A” wall version) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 55
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 (“B” column version) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 56
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 (“B”) (“C”) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 57
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 (“D”) NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
The extraordinary maintenance interventions are not carried out by the operator Extraordinary maintenance CARRIED OUT BY THE NEW LIFE RADIOLOGY SRL TECHNICAL ASSISTANCE IN CASE OF NEED AND IN ANY CASE AFTER RELEVANT REPAIRS . We provide a guideline on controls, on minimum checks that are to be performed in any technical intervention on the plant: monoblock, support and pantograph stability;...
Page 60
The accessories and connection cables must be disinfected only with a cloth soaked in disinfectant solution. Do not use solvents or corrosive disinfectants. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 61
In case of monoblock's damage for impact or crushing with a consequent oil leaking, prevent dispersal into the environment, proceeding to disposal as indicated above. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 62
Focal spot value for the reference axis comply with IEC 0,8 mm Kailong KL2-0.8-70G; 0,7 mm 60336: Toshiba DG-073B-AC NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Page 63
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 X-ray tube type kailong KL2-0.8-70G fig.28 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 64
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 Curve di riferimento fig.29 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 65
MANUAL Doc. HBE090-3 Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 fig.30 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 66
Installation and User Manual EVOSTYLE N.G. Mod. MQI006-0 X-ray tube type Toshiba DG-073B-AC Reference curves fig.31 fig.32 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Investigate and make sure of this condition before starting the exposure. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
Film accidental exposure to heat sources Film accidental exposure to daylight or safety lamp darkroom no longer suitable NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it...
During normal operation, the film must be positioned at 90 ° with respect to the cone collimator, as shown in fig.33 Upper jaw Lower jaw ig.33 NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...
Page 70
New Life Radiology does not assume responsibility for mistakes, omissions inaccuracies. NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095 Grugliasco (TO) Italy Phone and fax + 39.011. 781994 info@newliferadiology.it www.newliferadiology.it...