  • Page 1 Veri-Q PREP M16 G2-16TU User Manual Apr. 2020 Cat No. 9S101...
  • Page 2 Model name : G2-16TU For safe use, please use the device after reading the manual thoroughly. Please furnish this document at a designated location so it can be used at all times. MiCo BioMed Co., Ltd. and 4 Floor , 54 Changeop-ro, Sujeong-gu,...
  • Page 3: Table Of Contents

    Veri-Q PREP M16 G2-16TU User Manual Table of Contents 1. Before You Begin 5. Recommendations 1.1 About the manual 6. Trouble Shooting 1.2 General Requirements for Installation 6.1 Error Messages 1.3 How to Obtain More Information 1.4 About Customer Support 6.2 Additional Error Situation and Required Action...
  • Page 4: Before You Begin

    Warning Servicing on G2-16TU is performed by MiCo BioMed. 2.1.3 The user must use only components of MiCo BioMed in order to receive guarantee of the product. Please perform regular maintenance according to the user manual. MiCo BioMed charges the repair cost regarding problems arising due to incorrect maintenance.
  • Page 5: Working Environment And Installation Method

    Veri-Q PREP M16 G2-16TU User Manual 2.1.5 If the device does not meet the performance conditions, make an inquiry to the 2.2.5 Do not use a loose power cable or connector. If a power cable is overheated, it may be manufacturer.
  • Page 6: Operation And Maintenance

    2.3.4 In case of product malfunction, error message must be displayed on the LCD. If error Ventilation for fumes and disposal of wastes must follow all national, and local message is indicated, do not operate the product by force and make an inquiry to MiCo BioMed health and safety regulations and laws.
  • Page 7: Transport And Storage

    Veri-Q PREP M16 G2-16TU User Manual 2.3.11 Waste disposal - Used lab ware such as sample tubes and filter-tips may contain harmful Symbol Location Explanation chemical substances or infectious substances occurring from the extraction process. Such wastes must be appropriately collected and disposed of according to local safety regulations.
  • Page 8: Product Components And Specifications

    Veri-Q PREP M16 G2-16TU User Manual 3. Product Components and Specifications by using Varistor on operating parts such as valve, pump, fan, etc. We made sure the electric shock inside the board does not affect the element by using diode in the motor operating part.
  • Page 9: Equipment Part Names And Functions

    Veri-Q PREP M16 G2-16TU User Manual Contents Explanation 3.6 Equipment Part Names and Functions NST (Sample Tube) Plastic tube with nucleic acid binding membrane. Nucleic acid will be separated with this tube. * Disposable, re-use prohibited. NET (Elution Tube) Extracted nucleic acid is collected with Elution Tube.
  • Page 10: Product Features And Explanation

    Veri-Q PREP M16 G2-16TU User Manual 4. Product Features and Explanation 4.1 Product Features and Principle 4.1.1 Product Features Fluid control part for washing, air pumping, and extraction Main body Moving part to perform pre-treatment per channel 4.1.2 Operation Principle This device has the function to rapidly extract nucleic acid from the samples that have gone through the pre-treatment process.
  • Page 11: Transporting The G2-16Tu

    “Motor Initialize (RUN-INIT)” appears on the bottom of the LCD. In this case, do not operate 4.4 How to Use the equipment by force. And please make an inquiry to the AS team of Mico Biomed Co., Ltd. Check the G2-16TU device with the required parts according to installation protocol. You must regarding such malfunction.
  • Page 12 Veri-Q PREP M16 G2-16TU User Manual 4.5.2 Pre-running 4.5.3 Placing Elution Tube and Cartridge 1) Please check “PreRUN Required” message from the configuration screen. 1) Open the lower tray and place a Waste cartridge. 2) Insert the Elution Cartridge: Place the Elution Cartridge so it fits in the location for insertion.
  • Page 13 Veri-Q PREP M16 G2-16TU User Manual 4.5.6 Load eluted sample 4) Place the lid on the Elution Cartridge. This lid is for preventing the Elution tube from being contaminated. Afterwards, close by pushing the lower tray again. 1) Carefully place the samples into the Sample tubes and close the upper tray.
  • Page 14 Veri-Q PREP M16 G2-16TU User Manual 4.5.9 Start operation 1) Press “START” button or “▶Start” image. The equipment automatically operates. Please refer to the following for operating procedures. A. Checking & Binding When the user operates the equipment, the operation screen of the equipment changes or is displayed.
  • Page 15: Recommendations

    4.5.10 Finish 1) If the device operation is finished, you will hear a ‘beep’ sound. And the following image will We recommend that you use only MiCo BioMed reagents optimized for the equipment. For more appear. information, please visit http://micobiomed.com/, or send an e-mail to salesbio@micobiomed.com 2) Next, follow the following procedure.
  • Page 16: Trouble Shooting

    Veri-Q PREP M16 G2-16TU User Manual 6. Trouble Shooting 6.1 Error Messages Message Explanation This error message occurs in the following situation. Elution Door Open a. When lower tray door is opened This error message occurs in the following situation.
  • Page 17: Warranty Policy

    According to this warranty, all replaced materials are guaranteed only for the original warranty period, and shall not exceed the original expiration date unless there is written consent from MiCo BioMed. 10. EMC & Safety Test Results The device, interface device, and related software is guaranteed only during the period originally provided by the manufacturer.
  • Page 18: Cleaning

    Veri-Q PREP M16 G2-16TU User Manual 12. Cleaning 12.1 Cleaning of Air Rubber 12.1.1 Open the upper tray. 12.1.2 Prepare a clean cotton or cloth wetted with 70% ethanol. 12.1.3 Scrub the air rubber, especially the bottom side, with cotton or cloth wetted with 70% ethanol.
  • Page 19: Maintenance Checklist

    Veri-Q PREP M16 G2-16TU User Manual 13. Maintenance Checklist Equipment Checklist 4. Semiannual Checklist (through MiCo BioMed staff) 1. Daily Checklist Checklist Pass Fail Checklist Pass Fail The air pump pressure has been checked and calibrated if necessary. The equipment is in the designated location.
  • Page 20 This warranty is subject to the following exceptions and limitations : 1. Mico BioMed co., Ltd. Shall not be required to replace any unit which is damaged or malfunctions due to abuse, accidents, alteration, neglect, misuse, maintenance by someone other than Mico BioMed co., Ltd.
  • Page 21 Veri-Q PREP M16 G2-16TU User Manual Apr. 2020 MiCo BioMed Co., Ltd. and 4 Floor , 54 Changeop-ro, Sujeong-gu, Seongnam-si, Gyeonggi-do, Korea, 13449 TEL: +82-70-5227-6000 FAX: +82-70-5227-6001~2...