Performance Characteristics - Palmdoc II Owner's Manual

Table of Contents

Advertisement

Ashwood,E.R.(ed.),Tietz Textbook of Clinical
Chemistry. Philadelphia:W.B. Saunders Company
(1994),959.

Performance Characteristics

• Precision:
Standard deviation (SD) for each glucose
concentration < 100 mg/dL (5.55 mmol/L) and
coefficient of variation (CV) for each glucose
concentration ≥ 100 mg/dL (5.55 mmol/L)
is ≤ 5.0 mg/dL (0.278 mmol/L) and ≤ 5.0%,
respectively.
Intermediate precision
Control Solution Level
(mmol/L)
Pooled Mean (mmol/L) 2.5
Repeatability
Blood Glucose
(mmol/L)
Pooled Mean
60
Palmdoc 2 User Man.indd 60
SD
CV (%)
1.7~
2.8
2.2
(mmol/L)
SD
2.7
CV (%)
6.9%
Owner's Manual
Low
Normal
(1.7~2.8)
(5.3~8.0)
5.6
2.8
3.1
-
3.1%
2.9~
6.2~
8.4~
6.1
8.3
13.9
4.5
7.2
11.0
3.5
4.6
6.5
4.3%
3.5%
3.3%
High
(15.6~23.3)
19.6
8.6
2.4%
14.0~
22.2
18.0
8.3
2.6%