Biorep Technologies PER-5.2 User Manual

Biorep Technologies PER-5.2 User Manual

Perfusion apparatus

Advertisement

Quick Links

PERFUSION APPARATUS
MODEL No: PER-5.2
______________________________________________________
USER MANUAL
______________________________________________________
Page 1 of 21

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PER-5.2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Biorep Technologies PER-5.2

  • Page 1 PERFUSION APPARATUS MODEL No: PER-5.2 ______________________________________________________ USER MANUAL ______________________________________________________ Page 1 of 21...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS GETTING STARTED .......................... 3 ....................3 YMBOLS SED IN THIS ANUAL ......................... 3 ACHINE NFORMATION ......................... 3 ONTACT NFORMATION ........................4 AFETY NFORMATION .......................... 5 REPARATION ......................5 NPACKING NSTRUCTIONS ..........................5 ACKING INTRODUCTION ..........................6 MACHINE PARTS IDENTIFICATION ................... 7 ...........................
  • Page 3: Getting Started

    In the spaces provided below, record the Model and Serial No. located on the rear panel of this machine. Model No. __________________ Serial No.______________________ Retain this information for future reference. Contact Information Biorep Technologies, Inc. 3225 NW 112 Street Miami, FL 33167 info@biorep.com www.biorep.com...
  • Page 4: Safety Information

    If the provided plug does not fit into your outlet, consult an electrician for replacement of the obsolete outlet or contact Biorep Technologies, Inc. at (305) 687-8074. 4. Protect the power cord and the plug from being damaged. Avoid walking on or pinching the cord or plugs.
  • Page 5: Site Preparation

    Site Preparation This machine is intended to be used in a controlled laboratory environment. Please refer to the figure below for proper installation of the machine. The machine has one ventilation opening in the rear. A 15 inches clearance is recommended between the back of the machine and the wall. 15 Inches Figure 1: Dimensions of the Perfusion Apparatus 5 Unpacking Instructions...
  • Page 6: Introduction

    Introduction This manual is intended to be used by doctors, researchers, and technicians who have already been trained in islet isolation procedures and technologies. This document contains the necessary information for unpacking, installing, and operating the machine. The Perfusion Apparatus has been developed to assist in and allow automation of the pancreas perfusion step of the islet isolation process.
  • Page 7: Machine Parts Identification

    Machine Parts Identification Front View Automatic Heater Pressure sensor (A) Pressure sensor (B) Touch Screen Tray t/c Input Aux t/c Input Infrared Transducer Protector Pump B Transducer Protector Pump A Thermocouple Holder Holder Figure 3: Front View Page 7 of 21...
  • Page 8: Rear View

    Rear View Vent Plate Figure 4: Rear View Right View LAN Port 1 LAN Port 2 USB Port 1 USB Port 2 Power Entry Module Figure 5: Right Side View Page 8 of 21...
  • Page 9: Accessories

    Accessories The following disposable accessories are included with the Perfusion Apparatus. For ordering information, please consult our webpage www.biorep.com or contact one of our sales representatives at (305) 687-8074. Perfusion Tray (PERTRAY-MABS-01) The Perfusion tray has a removable mesh. Below the mesh, the tray has a funnel shape to direct all the fluid to the Enzyme Port.
  • Page 10: Disposable Tubing Set With Coil (Per05-Tubhc-01)

    Disposable Tubing Set with Coil (PER05-TUBHC-01) For information about how install the tubing set please refer to Section 5.1. Luer Connectors Transducer Protector B Heating Coil Transducer Protector A Figure 7: Perfusion Apparatus Tubing Set (PER05-TUBHC-01) Important: If silicone tubing remains inside a closed peristaltic pump head for extended periods, there may be permanent tubing profile deformation.
  • Page 11: Perfusion Machine Set-Up

    Figure 8: Biorep Thermoprobe Perfusion Machine Set-up Tubing Set-up Luer Connectors (To Organ) Transducer Protector B Transducer Protector A Heating Coil Figure 9: Tubing Set Installation Tubing Set-up Directions (also see photograph on the following page): 1) Place the heating coil inside the automatic heater. Make sure there is good contact between the stainless steel coil and the aluminum arms.
  • Page 12 3) Place tube sections “D” and “E” in the appropriate peristaltic pump. Make sure the tubing is not being pinched on the inlet or outlet. 4) Connect Tubes “A” and “B” to the appropriate pressure sensor fittings. 5) Secure both “Transducer Protector A” and “Transducer Protector B” to their respective holders.
  • Page 13: Heater Set-Up

    Heater Set-up A heating element is a coiled stainless steel tube. It has a large heating surface area to provide precise and non-localized heat to the fluid that circulates through it. The temperature of the fluid is constantly being monitored and controlled by the Infrared Thermocouple immediately after the heater’s outlet.
  • Page 14: Perfusion Apparatus Operation

    IMPORTANT: At least one peristaltic pump has to be ON (flow rate > 15ml/min), AND the heater cover needs to be closed for the heater to operate. Perfusion Apparatus Operation The main operation of the Perfusion Apparatus is done through the touch-screen interface. This type of interface eliminates the need for an external keyboard or mouse device and provides a visual and intuitive interaction with the user.
  • Page 15: Power-Up Page

    Power-up Page This will be the first page that appears after the system has completed the boot-up process. In this page you will be prompted to touch anywhere to begin. Do not touch the screen or enter any commands via keyboard or mouse until you see this screen. Login Page The Login Page allows you to Log into your password protected user account.
  • Page 16: Main Menu

    Main Menu Once logged in, you will arrive at the Main Menu page. On this page you can select the operation mode of the machine. “Manual Mode” will take you to a control panel interface where you can manually manipulate the machine parameters during an experiment session.
  • Page 17: Control Panel

    6.5.2 Control Panel All the controls required to operate the machine during a Manual Mode experiment are located in this single interface. In this control panel you will be able to change the Temperature parameters, the Flow/Pressure parameters and control the recording of your experiment for use in the Data Viewer mode.
  • Page 18 6.5.2.2 Flow control There are two independent flow loops labeled LOOP A and LOOP B. Both loops operate the same way, so we will describe the controls in LOOP A only. The “drop” icon on the top left is the ON/OFF button for the peristaltic pump.
  • Page 19: Data-Viewer Mode

    Data-Viewer Mode The Data Viewer screen is where all recorded experiments can be analyzed, replayed or deleted. Each experiment recording is called a Report. A list of all reports can be found in the first tab of the Data-Viewer page. They are listed in chronological order, so the last experiment will be found last on the list.
  • Page 20: Experiment Playback

    You can delete a report from the list by highlighting it and clicking on Delete. All deleted reports can be found on the last tab “Deleted Reports” and can be recovered from there. The most valuable feature of the Data-Viewer is the ability to replay an experiment. To do this, select the experiment you wish to replay and click on Playback.
  • Page 21: Troubleshooting

    Check that the heater is in good contact with the electrode arms and that it they are screwed properly. If problems persist or you encounter any other problems or have additional questions regarding the operation of the perfusion apparatus, please contact customer support at: Biorep Technologies, Inc. info@biorep.com; Tel: 305-687-8074 Page 21 of 21...