Chapter 3 Features and Functions
Calibration Overview
To Unsecure for Calibration You can unsecure the multimeter
for calibration either from the front panel or remote interface.
The multimeter is secured when shipped from the factory, and the
security code is set to "
Front-Panel Operation:
1: SECURED
If the multimeter is secured, you will see the above command when
you go into the
CAL MENU
in the menu, you will notice that the "
"hidden" if the multimeter is secured.) To unsecure the multimeter,
select the "parameter" level of the
security code, then press Menu Enter.
000000 CODE
When you go to the "commands" level in the
you will notice that the multimeter is unsecured. Notice also that the
"
" command is no longer hidden and you can perform
2: CALIBRATE
a calibration.
1: UNSECURED
Remote Interface Operation:
CALibration:SECure:STATe {OFF|ON},<
To unsecure the multimeter, send the above command with the same
code used to secure. For example,
"CAL:SEC:STAT OFF,HP034401"
96
".
HP034401
(CAL MENU)
. (If you move across the "commands" level
2: CALIBRATE
SECURED
" command is
command, enter the
again,
CAL MENU
code
>