haTmed Q50B User Manual

haTmed Q50B User Manual

Small steam sterilizer (internal water tank)
Hide thumbs Also See for Q50B:

Advertisement

 
 
 
 
Small Steam Sterilizer 
User Manual 
               (Internal Water Tank) 
 
 
 
 
 
Q50B/Q70B
                                                                                                       
 
2460 
 
HӓTmed (Wuxi) Co., Ltd 
V01 version

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Q50B and is the answer not in the manual?

Questions and answers

Summary of Contents for haTmed Q50B

  • Page 1         Small Steam Sterilizer  User Manual                 (Internal Water Tank)            Q50B/Q70B                                                                                                           2460    HӓTmed (Wuxi) Co., Ltd  V01 version...
  • Page 2     Application    HӓTmed Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms  to the YY/T 0646‐2015 standard and is a Class B sterilizer.Process evaluation system according to EN13060.   Do  not  sterilize  any  liquid,Violating  the  use  of  this  product  to    sterilize liquids will damage the product and the loads.        Symbols description     :  Symbols  Indication  Description  Identifies  danger  points  or  conditions  and  Warning  the  correct  procedures  that  must  be   ...
  • Page 3 Name and address of authorized agent of this product in the European Union.    Service should be performed by qualified technicians only. Observe all local    and national directives related to service and safety standards.  The function and life span of this product are directly related to whether the  user follows the operating specifications and maintenance requirements in    the installation and use manual.  Manufacturer confirms that this product complies with the European Union  Medical Device Specification.    Draw user's attention to important matters.          Safety instruction    Identifies  danger  points  or  conditions  and  the  correct  procedures  that  must  be  followed  to  avoid injury.       Liquid sterilization  Do not sterilize any liquid  Professionals   The operator who use sterilizers and fabric must be got a trained professional. ...
  • Page 4  Do not use or operate this product at sites where there is a risk of explosion.   The  electrical connection  of this  product and the  connection  of  the  inlet  and  outlet  pipes should be carried out by professionals.   This product should not be in direct contact with the patient. Place this product outside  the range of at least 1.5 meters (radius) in the treatment area.   If an external printer is configured for this product, the printer should not be placed in  a location where there is a risk of contact with liquid.  Emergency door   Do not use this device to open the door under normal conditions.  opening device   Pretreatment and sterilization of instruments and fabrics should follow the regulations  of the instrument and fabric manufacturers.   Sewage pretreatment  Strictly  follow  the  relevant  standards  for  disinfection  and  sterilization,e.g.WS506‐ and sterilization  2016,WS310‐2016.  ...
  • Page 5  If the same fault message appears repeatedly in this product, please turn off the power.   malfunction  Repair must be performed by a professional authorized by the manufacturer of this  product.          Sterilization Program Introduction     Standard  Standard  Quick  Name of sterilization  Quick B  134+  Fabric  134  121  S  Class of sterilization  B  B  B  S  B  B  Dynamic pressure monitoring  √  √  √  √  √  √  Air leak  √ ...
  • Page 6: Table Of Contents

    TABLE OF CONTENTS   1   Device Description ............................ 1   1.1   Delivery list ............................ 1   1.1.1   Standard delivery list ........................ 1   1.2   Product view ............................ 1   1.2.1   Front view   ............................. 2   1.2.2   Front view   ............................. 2   1.2.3   Side view ............................ 3   1.2.4   Rear view  ............................ 3   1.2.5  ...
  • Page 7 5.3   Helix test .............................  2 0   5.4   Drainage program ..........................  2 1   6   System settings ............................  2 1   6.1   Time setting ............................  2 1   6.2   Date setting ............................  2 2   6.3   Language settings ..........................  2 2   6.4   Warm‐up setting ..........................  2 3   6.5  ...
  • Page 8: Delivery List

    Device decription  Delivery list  1.1.1 Standard delivery list      The standard accessories sent with sterilizer are listed in the table below. Please check the list when unpacking.  1.1.1 Standard delivery list  Accessory  Accessory name  Q’ty  Description  Picture  Air level is placed above the sterilizer during  installation. Two foot pads at the front  is higher than  Air level  1  the back end. The angle between the bottom of    sterilizer and countertop is around 5‐10°.  Aluminum tray with holes at the bottom for placing  Tray  3  sterilized items.    Tray handling  1  Tool to take out the tray     Chamber rack  1  Rack to hold the tray   Fuse  2  220V/16A fuse    Air filter to ensure that the air entering the  Air filter  1  sterilization chamber is highly filtered to  reduce  bacteria and dust entry.   ...
  • Page 9: Side View

    Display                  Fig 1.1 front view    1.2.2 Side view    Door   handle  Power  switch  Fig 1.2 side view    - 2 -...
  • Page 10: Rear View

    1.2.3 Rear view    Printer interface  Test interface  USB interface  Emergency door   opening device  Safety valve  Power cable  Air filter  Cooling fan  Wastewater   drainage interface A (empty)         Fig 1.4  Rear View    Wastewater   drainage interface B (half drainage)      1.2.4 Inside View    Power Switch    Fig 1.5 inside view  - 3 -...
  • Page 11: Top View

      1.2.5 Top View  Distilled water tank  (Internal)    Fig 1.6 Top view      Technical parameters  parameter name  Q50B  Q51B  Q70B  Q71B  External size  570mm*450mm*470mm  680mm*450mm*470mm  Packaging size  688mm*576mm*688mm  806mm*576mm*688mm  Chamber size  Φ253mm*355mm  Φ253mm*475mm  Capacity of Sterilizer  18 L  23 L  Pressure vessel weight  4.8 kg  6.2 kg  Equipment weight  59.6 kg  71.6 kg  64.8 kg  66.8 kg  Built‐in clean water tank  3.8L  volume  Built‐in waste water tank  4.5L ...
  • Page 12: Loading Method

    Working environment  70 Kpa ~ 106 Kpa  pressure   o  o Working temperature  C ~ 40 C  Storage and transportation   o  o ‐20 C ~ 55 C  environment temperature  environment humidity  ≤ 85%  Maximum wastewater   o C  temperature  Maximum noise  70 dB(A)  Compliance with GB/T18268.1‐2010; Works properly in its electromagnetic  environment, but does not constitute an unacceptable electromagnetic  Electromagnetic  harassment of anything in the environment。At the same time, it has the  compatibility  ability to resist electromagnetic interference.  According to GB 4824‐2013,Electromagnetic radiation meets Group 1 Class  B requirements.  Standards compliant  YY/T 0646‐2015; GB4793.1‐2007; GB4793.4‐2001;  Note 1:The maximum operating voltage range :198V ~ 242V      Note 2: Anti‐electromagnetic interference is the ability of the sterilizer to  work properly in the presence of electromagnetic interference    Loading method  The ...
  • Page 13: Operation Buttons

    1.5.1 Operation buttons and Identification  The use of 4 buttons in different situations is as follows:  1  2  3  4  Button position (Face up from left  to right)          When the program is  Previous  return  Next option  Select  selected  option  standby mode      Open the door  Startup  Program runtime        Stop  When setting        Confirm    There is an icon in the interface to display the door lock status.  Status Identification  Description  Door lock status ...
  • Page 14: Vacuum Test

      1.7.2 Vacuum test  The purpose of the vacuum test is to test the leak rate of the sterilizer. No sterilization of the items in the  chamber during the process. Please run the vacuum test in a cold and dry chamber without loads.    Process name  Description  The air in the chamber is evacuated till the pressure in the chamber reaches  1.Vacuum  the set value of the vacuum test  After waiting for 5 minutes, the vacuum everywhere in the chamber tends to  2.Balancing time  be uniform.  3. Test holding  Test holding time is 10 minutes, The device continuously monitors the  time  pressure in the chamber during the period  After the test is over, the screen displays the test results: vacuum and leak  4.End of test  rate.      Installation and handling instructions  Unpacking precautions  Installation, setup, and commissioning should be performed by a  professional authorized by the manufacturer of this product.    This product is taken out of the box and transported must be  operated by 2 people.       Total weight:Q50B 59.6 kg;       Q51B 61.6 kg;    Q70B 64.8 kg;       Q71B 66.8 kg   - 7 -...
  • Page 15: Installation Site Requirements

      an authorized dealer or logistics company immediately.  Please check the packing box and the random accessories inside  the sterilizer. A list of standard accessories can be found in section    1.1 of this manual.    Installation site requirements  This product should be placed in a dust‐free and dry environment。Surrounding humidity is less than 85%,   o  o Temperature between 5 C  and  40 C.   Do not use or operate this product at a site where there is a risk of  explosion.     Countertop must be able to support the weight of this product :  Q50B 59.6 kg;       Q51B 61.6 kg;      Q70B 64.8 kg;       Q71B 66.8 kg      Each surface of the device (except the bottom) must have a gap  of 10 cm outward. Too small a gap will result in heat buildup,  shorten the life of the equipment, reduce the performance of the  equipment and extend the running time of the program.    Device height adjustment  Take out air level and put it in the horizontal position at the top of the product and adjust the two foot pads at  the front end to make sure the front is higher than the rear side. The angle between the bottom of sterilizer and  countertop is around 5‐10°. It is beneficial to realize the drying of the loads and saves distilled water.    Rear end  Front end                Fig 2.1 Device height adjustment ...
  • Page 16: Connection Requirements

    Connection requirements  2.4.1 Electrical connections   220V/50Hz power supply. The maximum power supply  fluctuation range is 198V ~ 242V.   Use a separate 16A power circuit. Provides 30mA leakage  protection.   Make sure the power cable is not knotted or pressed.   Do not use force to pull the power cable,   Do not place the power cable near a heat source.     Do not touch the power cable with a sharp object.   Do  not  replace  the  power  cable  or  power  plug.  Do  not  use  a  damaged power cable or power plug   Do  not  use  this  device  if  the  power  cable  or  power  plug  is  damaged. The power cable or power plug must be replaced by a ...
  • Page 17: Wastewater Connection

     This  product  has  water  quality  monitoring  device,it  will  not  work  properly when unqualified water and will give warning that water  quality  cannot  reach  the  standard.When  you  see  this  message,  please replace high quality distilled water in time.    If  you  need  to  use  the  water  purifier,  please  consult  the  manufacturer of this device first.   It is best to tilt and secure all the water pipes connected to the back  of the device are not bent or knotted.    2.4.3 Wastewater connection ...
  • Page 18: First Time Operation

      First time operation  Startup                   Turn on the power switch, the power switch position is in the lower right corner of the device inside view  (Figure 1.4).                     Make sure to remove the latex plug from the air filter mounting    hole  at  the  back  of  the  device  and  install  the  air  filter  from  the  accessories included.    After Startup  After startup at the first time, please set up language and carry out pressure test. After the pressure verification  is successful, the initial setting is completed. The date and time can also be set as required. See Chapter 6 system  settings in this manual.  Steps of pressure verification is as follows:     Door  open  to  complete  pressure verification      press button  and open up the door  ...
  • Page 19: Distilled Water Injection

    Door open up operation:Please press  button at standby mode,you will see door lock icon will change  its status on the display to indicted the door is opened. At this time, open the door handle on the right side of the  device upwards by about 90 C and then open the door outward.  Closing operation:Close the door of sterilizer,The door handle is turned down to the end。   When closing the door, please press the door handle down to the  end.  The  strength  required  after  using  for a  period of  time  may  increase     Don’t slam the door at any time.   Failure  to  comply  with  the  above  requirements  may  affect  the  function and reliability of the door lock device   Open the door only when loading and unloading, Keeping the door  closed at other times is beneficial to energy saving      Distilled water injection  Inject qualified distilled water into the distilled water tank and make sure the water level is between the lowest ...
  • Page 20: Drainage

    Packaging  materials  for  packaged  sterilization  products  shall  comply with GB/T 19974 standard.     Sterilized items packed in paper and plastic bags, The paper must  be faced up and laid flat in the tray.   For mixed load for sterilization, fabric is on the top and the metal  is on the lower layer.   There must be a gap between the sterilization bags.   Metal bags and dressing bags cannot be attached together.   The load capacity must not exceed 90% of the chamber capacity.    Select sterilization program  This  product  has  7  preset  sterilization  programs.  Please  select  the  appropriate  sterilization  program  according to the load type and packaging material. The parameter setting of the sterilization program and the typical  load description are as follows. The maximum load of Q50B/Q51B is 6kg instruments or 2kg fabric.The maximum  load of Q70/Q71B is 7kg instruments or 2.5kg fabric.      - 13 -...
  • Page 21: Auxiliary Option

                Table 4‐1 Sterilization program and parameter settings  Drying  Holding  Name of  T  time  Sterilization load  time  Sterilization  (℃)  (min)  (Min)  , 121  a  small  amount  of  fabric 121  >15  1‐99      program Heat‐resistant instrument. 134   Mixed  load,long  and  hollow  >3  134  1‐99   ...
  • Page 22: Automatic Printing

    The  chamber  will  automatically  heat  up  to  110  °C  when  the  automatic  preheat setting is turned on. Please be careful of burns!      4.3.2 Automatic printing  Turn on the automatic printing function and the sterilizer is properly connected to the printer to  automatically print the sterilization record after the program ends. See section 6.5 for details      4.3.3 Extra drying  Additional drying settings based on user preferences, extend the drying time of the preset sterilization  program based on the drying time in Table 4‐1 to achieve good drying results for loads, especially cotton fabrics.  See section 6.6 for details.      Running a program  4.4.1 Start sterilization  After selecting sterilization program,pleaes press  to start up program.Sterilizer will ensure that the door  is locked and the water quality of the distilled water is monitored after the start of the program.              07/25/2017       09:23:58  ...
  • Page 23: Manual Stop

    Manual stop  A running program can be manually stopped at any stage. If the sterilization process is stopped before  the start of drying, the loaded items are not sterilized. If the sterilization procedure is manually stopped after the  start  of  drying,  the  loaded  items  have  been  effectively  sterilized,  but  not  fully dried. In  the  case of fully  loaded  packages, the possibility of wet bags is very high and these items cannot be stored because Because aseptic storage  places high demands on the dryness of the packaged items.   After  the  program  is  manually  stopped,  there  may  be  high  temperature  steam coming out and causing burns when the door is opened!   Do not touch the product, loaded items, chamber or door cover    with bare hands.   Be sure to use the tray handle to take the loaded items.  Manually stopping the sterilization process at an earlier stage may cause an  infection! ...
  • Page 24   Fig 4.3 Manual stop    Sterilization procedure completed  When the sterilization process ends normally, the program end interface and program running success  information are displayed on the screen. If the automatic printing function is set and the printer is connected, the  product will automatically print the sterilization record. Please press  to open the door.      Careful burns!   Do  not  force  the  door  to  open.  Failure  to  do  so  may  result  in  damage  to  the  product  and/or  hot  steam  coming  out  of  the  chamber.   After the sterilization is finished, wait for a while until the sterilizer  is fully cooled before opening the door. ...
  • Page 25 the  cycle  is  134  °  C,  the  holding  time  should  be  not  less  than  3  minutes.   Chemical  monitoring  should  be  performed  for  each  sterilization  cycle,by observing the color change of the chemical indicator to  determine  whether  the  sterilization  requirement  is  met  compliance  with  the  information  and  requirements  provided  by  the manufacturer of the chemical indicator. ...
  • Page 26: Functional Testing And Other Procedures

        Functional testing and other procedures  Vacuum test  When first installed, unused for a long time, moved, or periodically tested,vacuum test is recommended to  detect leakage in the system of this sterilizer:    1 –Sterilization program       2 – Test program    3 – System setting    4 – File record  press to select    Vacuum test program     B&D test program    HELIX test program    press to enter       07/25/2017       09:23:58     Vacuum test program       15Min    T:  25.0℃    P: 0.00 Bar    Door status    press to start up ...
  • Page 27: B&D Test

    After the vacuum test program is started, when the pressure in the chamber reaches the standard required  relative vacuum value, the vacuum system stops, and the equilibration time of 5 minutes is turned on. After the  balancing time elapses, the vacuum test is started for 10 minutes. After the test is completed, the system returns  to normal pressure.  After the vacuum test is finished, the system will prompt the result on the program end interface and save the  record in the memory. If the automatic printing function is set and the printer is connected, the test result will be  printed at the same time.      When the vacuum test is passed, the message is “vacuum test success”. Indicates that the system's leak rate  meets the design requirements. Print the information as follows:  ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐           PROGRAM PROPERLY EXECUTED!        Vacuum Test Success!              Leakage rate: <0.13Kpa/min         Leak rate:<0.13Kpa/min  Your system is steady!           ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐   When the vacuum test fails, the message is “Vacuum Test Failed”. Print the information as follows:           ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐   PROGRAM ERROR EXECUTED  Vacuum Test Failed!                     Vacuum Test Fail!                  ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐ ‐‐   ...
  • Page 28: Helix Test

    Helix  Test  The Helix test is used to verify the steam penetration and saturation for A type hollow instruments, such as  dental handpieces. Select the Helix test in the test program interface (Figure 5.1).   Test methods and test results are judged in accordance with the  information  and  requirements  provided  by  the  test  package  manufacturer.      Drainage program  This product allows users to perform drainage procedures when needed. This procedure is used to drain water  remaining in a small steam sterilizer.   When the water in the chamber is obvious after opening the door,  it is recommended to use the drainage program before loading the  item for sterilization. Especially when the program is abnormally  stopped, this situation is easy to occur. Please open the door for    inspection.   Please  carry  out  the  drainage  program  before  handling  this  product.  System setting  Enter  the  system  settings  interface  (Figure  6.1)  for  time,  date,  language,  warm‐up,  automatic  printing,  additional drying settings, and system reset. ...
  • Page 29: Date Setting

          Completion     Fig 6.4  Time Setting            Date Setting  Date setting operation is as follows:    Time setting  Press to  select  date    Date setting    setting.Press to  enter    Language setting date setting.    Warm‐up setting       Press and  to  set  up    Time setting ...
  • Page 30     Completion      Fig 6.2 Language Setting  Warm‐up settings  Warm up setting operation is as follows:                Time setting    Press to select warm‐up  Date setting    setting.Press to  enter  Language setting  into warm‐up setting.    Warm‐up setting        Warm‐up setting    Press  to  adjust.        Automatic warm‐up:  Press  to exit warm‐up setting        Close      after completion.  Back                         Confirm     ...
  • Page 31: Warm-Up Setting

    Fig 6.6 Automatic print setting   Additional drying setting  Additional drying setting operation is as follows:    Press to  select  additional         Warm‐up setting           Printer setting  drying  setting.  Press to    enter  to  set  up  additional  Additional drying setting   drying.          System reset          Additional drying setting      Press and  to  enter  into  additional drying setting.  Additional drying:close      Press to exit additional drying ...
  • Page 32: System Reset

              Press and  to  select    System rest   whether  to  perform  a  system      reset. Press  to exit after the              Yes            No    setting is completed.      Completion    Fig 6.9 System reset interface     System Information  System information operation is as follows:    Water supply setting    System reset  Press to  select  system  Administrator setting.
  • Page 33: Log Export

    Log export  For  long‐term  storage  of  sterilization  records,  sterilization  records  can  be  exported  via  USB  interface  to  a  computer or portable storage device. Specific operation: Insert the USB flash drive into the USB interface and export  the record as follows. After the export is completed, the interface displays the export completion prompt. Please  remove the USB flash drive.    1 –Sterilization program     2 –Test program     3 –System setting  按 键 进 行 记 4 – File record  录导出。       File export    Export all records?     ...
  • Page 34: Log Interpretation

        Log interpretation  The format of the sterilization record for printer output and internal memory storage is as follows.      - - - - - - - - - - - - - - - -   Quantus Sterilizer Product Number    - - - - - - - - - - - - - - - -  ...
  • Page 35: Small Steam Sterilizr Maintenance

    Small Steam Sterilizer Maintenance  Cleaning of loaded items prior to sterilization     Before sterilization, the load of handpiece must be thoroughly cleared.   Dust and contaminants separated from the load will cause blockage of filters, valves and  piping, therefore, the sterilization load without cleaning is not allowed to enter the device  for sterilization.   Follow all regulations for cleaning, disinfection, and sterilization provided by the device  manufacturer.  Small steam sterilizer maintenance  8.2.1 Cleaning and maintenance  Regular cleaning and maintenance can effectively extend the life of the sterilizer and ensure a good working  condition.  Maintenance item  Maintenance cycle  Special tool  Maintenance operation  Replace the seal  Sterilize 1000 times    Open the door and check whether the seal is  deformed or wrinkle, etc. If it is deformed or  wrinkled, it is recommended to replace it.  Seal maintenance  2 weeks    Remove  the  seal.Please  rinse  it  with  water  and  wipe  it  dry  with  a  soft  cloth.After  that  replace the lid. ...
  • Page 36: Regular Maintenance

    Clean the filter at   Wrench  Unscrew the filter with a wrench and wash it  the  bottom  of  the  Sterilize 200 times  with clean water.  sterilization chamber    Rinse it with clean water, if stained, clean it  Clean trays  Sterilize 100 times  with toothpaste    Wipe the sterilization chamber with a clean  and wet medical cotton cloth,then wipe the  Sterilization chamber  Every week  chamber with a clean and dry medical cotton  cloth      8.2.2 Regular maintenance   WS310‐2016  and  WS506‐2016  require  a  comprehensive  maintenance  of  the  small  steam  sterilizer  every  year  to  ensure   ...
  • Page 37: No Distilled Water

    Solution:   Please check if the foot pad adjustment is correct. Under normal circumstances, the front side is  slightly higher than the back side.   Open the door and check if the filter at the bottom of the sterilization chamber is blocked.   None of the above can help reduce the consumption of distilled water, please contact professional  distributor authorized by HӓTmed.    The drying effect is not good  Faulty description: Drying cannot achieve satisfactory results  Solution:   Please check if the foot pad adjustment is correct. Under normal circumstances, the front side is  slightly higher than the back side.   Open the door and check if the filter at the bottom of the sterilization chamber is blocked.   Please check if the load exceeds the maximum loading capacity(especially textiles) and if the  load   is placed correctly in the chamber(do not touch the wall of the pressure chamber directly) and if the  standard trays and tray rack are used for loading.   Activate automatic warm‐up.   Open the left side cover and check if the filter is clogged   None of the above can help improve the effect of drying, please contact professional distributor  authorized by HӓTmed.    No distilled water  Faulty description:No distilled water  Solution:   Please check if there is water in distilled water tank.    Please check if the distilled water pipe is bent or knotted and inserted into place.  Unable to open the door  Faulty description: The door cannot be opened even if the emergency button is activated.  Solution:  ...
  • Page 38: Leakage At The Door Or Dripping Underneath

     Return the silicone tube to the original position and clamp. Turn on the vacuum test to check for  leakage.  Leakage at the door or dripping underneath  Faulty description: Air leakage and water dripping are obvious.  Solution:     Please check if the seal is installed correctly(the long side with the word "H" is exposed outside).  Please check if the seal is damaged.    Appendix 1 Safety Device Description  Preheating  system  over  temperature  protection  switch:  The  preheating  system  has  an  over‐temperature  protection switch that automatically disconnects the preheating system power supply circuit when the temperature  exceeds  the  limitation  of  protection  switch  threshold  temperature  to  realize  the  protection  function  of  the  preheating system. ...
  • Page 39: Appendix 3 Electrical Schematic Diagram

          Piping system schematic                        Appendix 3 Electrical schematic diagram    - 32 -...
  • Page 40 Small steam sterilizer electrical schematic diagram ...

This manual is also suitable for:

Q70bQ51bQ71b

Table of Contents

Save PDF