Start-up and Operation
Item
Step
No.
01
Switching the
appliance on
02
Opening the door
03
Loading a
compartment
04
Closing the door
05
Confirming loading
06
Start
07
Thawing /
heating
®
2000
Quick Reference Guide
Activity
Use the ON / OFF power switch to
switch the appliance on.
Use the door handle to open the
door.
To load a blood bag into any vacant
compartment,
•
prepare the bag, as shown in
the pictograms and described in
the instructions provided;
•
raise the spring plate of the
compartment;
•
place the bag in the middle of
the compartment;
•
make sure that tube ends /
connectors are located on the
right-hand side and on top of
the bag;
•
release the spring plate.
•
IMPORTANT
Bag filling volume:
100 ml to 600 ml
Use the door handle to close the
door.
When prompted by the yellow
flashing LED and the text appearing
in the display, press the confirmation
button of the compartment just
loaded.
When prompted by the yellow
flashing LED, press the start button.
A star symbol will appear in the
Universal model and when fresh
frozen plasma (FFP) bags are
inserted. This star will be
extinguished when approximately
+20° C are reached, with an acoustic
signal indicating that the bag has
thawed and may be removed, as
required. If the bag is not removed,
heating will continue until the target
temperature is reached.
Display
Attention to the Comments Below.
SELF-TEST; ON
Appliance carries out a self-test
VACANT
(approx. 3 seconds).
ON
ON
Place the bag in the middle of the
----------
compartment, with tube ends
located on the right-hand side and
on top of the bag.
Always make sure that
not protrude from the
compartment.
When it comes to bags containing
plasma / FFP, tube connectors
frozen in a bent position must be
located on the right-hand side and
on top of the bag. If the
connectors are frozen in a straight
position, the bag is to be inserted
with the connectors located on the
right-hand side.
It is possible to thaw a blood bags
inserted into protective bags,
provided the above instructions
are observed.
ON
ON
CONFIRM
LOADING
ON / START
LOADED
ON / HEATING
The appliance will automatically
LOADED
recognize bags that are already
heated. If this is the case, the
Star symbol if the
appliance will shut down. An
bag contains
intermittent alarm tone will be
FFP
sounded. A read flashing signal
will appear on the compartment
concerned, and the display will
show REMOVE / CHECK BAG.
Remove the bag. Check the
temperature of the bag and/or
reinsert the bag. Once the loading
of the bag has been confirmed
and the start button pressed, the
heating process will continue.
Important: Be Sure to Pay
tubes do
Status: June 2013