Battery Powered Digital Pressure Gauges, Min/Max & Selectable Units
®
Ranges and Resolution
See table below for standard ranges and units
Resolution is fixed for each engineering unit
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Pressure and absolute ranges
±0.1% full scale, ±1 least significant digit
Vacuum, compound, bipolar ranges
±0.25% full scale ±1 least significant digit
Sensor hysteresis:
±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5"H and 5 character 0.25"H alphanumeric
White LED backlight active with button keypress
Batteries, Battery Life, Low Battery Indication
2 AA alkaline included
Approx. 150-1500 hours depending on backlight usage
Low battery symbol on display
Controls and Functions
Front button powers gauge on/off, activates backlighting for
1 minute, zeros gauge reference gauges, and cycles through
min./max. functions.
Passcode protected Internal buttons for calibration, engineering
unit selection, auto shutoff time, min/max setup.
Min/Max Memory
Minimum and/or maximum readings stored in memory, read-
ings cleared or stored at shutoff. User configurable.
Sensor Ranges and Engineering Units
3 psig *
Res
15 psig vac *
3PSIG
.001 100KPAVAC
6INHGG
.001 0.1MPAVAC
85INH2OG
1BARVAC
.1
50ZING
.01
1KGCMVAC
210GCMG
.1
1ATMVAC
150MMHGG
.1
15 psig
150TORRG
.1
15PSIG
200MBARG
30INHGG
.1
200CMH2OG
400INH2OG
.1
2000MMH2OG
1
240ZING
7FTH2O
.001 1000GCMG
20KPAG
.01
760MMHGG
5 psig *
Res
760TORRG
5PSIG
.001 1000MBARG
10INHGG
1000CMH2OG
.01
140INH2OG
35FTH2O
.1
80ZING
.1
100KPAG
350GCMG
.1
0.1MPAG
260MMHGG
.1
1BARG
260TORRG
.1
1KGCMG
350MBARG
1ATMG
.1
350CMH2OG
±15 psig *
.1
3500MMH2OG
±15PSIG
1
12FTH2O
.01
–30INHG/15PSIG
35KPAG
.01
±30INHGG
15 psia
Res
±400INH2OG
15PSIA
.01
±240ZING
30INHGA
±1000GCMG
.01
400INH2OA
±760MMHGG
.1
240ZINA
±760TORRG
.1
1000GCMA
1
±1000MBARG
760MMHGA
.1
±1000CMH2OG
760TORRA
.1
±100KPAG
1000MBARA
1
±0.1MPAG
1000CMH2OA
±1BARG
1
100KPAA
±1KGCMG
.1
0.1MPAA
.0001 ±1ATMG
1BARA
.001
30 psia
1KGCMA
.001 30PSIA
1ATMA
.001 60INHGA
15 psig vac *
Res
850INH2OA
15PSIVAC
480ZINA
.01
30INHGVAC
2100GCMA
.01
400INH2OVAC
.1
1600MMHGA
240ZINVAC
.1
1600TORRA
1000GCMVAC
1
2000MBARA
760MMHGVAC
.1
2100CMH2OA
760TORRVAC
200KPAA
.1
1000MBARVAC
0.2MPAA
1
1000CMH2OVAC
2BARA
1
www.mod-tronic.com | sales@mod-tronic.com | 1-800-794-5883
Calibration
Zero button for gauge reference ranges
Non-interactive zero, span, and linearity, ±10% of range
Auto Shutoff
User selectable 1 minute to 8 hours or front button on/off
Weight
Gauge:
9 ounces (approximately)
Shipping:
1 pound (approximately)
Housing Materials
Epoxy powder coated aluminum case and bezel, front and
rear rubber gaskets, polycarbonate label, rubber boot
Connection and Material
1/4" NPT male fitting
All wetted parts are 316L stainless steel
Overpressure, Burst, Vacuum Service
3000 psig sensor: 5000 psig overpressure
5000 psig sensor: 7500 psig overpressure
All others:
2 X pressure range overpressure
Burst pressure:
4 X sensor pressure rating, or 10,000 psi,
whichever is less
Vacuum service:
15 psia, ±15 psig, 15 psig, 30 psia,
100 psig, 100 psia, 200 psig sensors
Environmental Temperatures
Storage:
Operating:
Compensated range: 32 to 158°F
Dimensions
3.67"W x 3.19"H x 2"D with boot, not including fitting
*
±0.25% accuracy
Res
30 psia
Res
–15V100psig *
2KGCMA
.001 –15V100PSIG
.1
.0001 2ATMA
.001 –30INHG/100PSIG
30 psig
Res
–30V200INHGG
.001
.001 30PSIG
.01
–400V2770INH2OG
.001 60INHGG
.01
–240V1600ZING
Res
850INH2OG
1
–760V5200MMHGG 1
.01
480ZING
.1
–760V5200TORRG
2100GCMG
–100V700KPAG
.01
1
1600MMHGG
–0.1V0.7MPAG
.1
1
.1
1600TORRG
1
–1V7BARG
1
2000MBARG
1
–1V7KGCMG
.1
2100CMH2OG
1
–1V7ATMG
.1
70FTH2O
.01
100 psig
1
200KPAG
.1
100PSIG
0.2MPAG
.0001 200INHGG
1
2BARG
.001 2770INH2OG
.01
.1
2KGCMG
.001 1600ZING
.0001 2ATMG
.001 7000GCMG
.001
60 psig
Res
5200MMHGG
.001 60PSIG
.01
5200TORRG
.001 120INHGG
7000MBARG
.1
Res
1660INH2OG
7000CMH2OG
1
960ZING
230FTH2O
.01
1
.01
4200GCMG
1
700KPAG
.01
3100MMHGG
1
0.7MPAG
1
3100TORRG
1
7BARG
.1
4100MBARG
1
7KGCMG
4200CMH2OG
7ATMG
1
1
140FTH2O
–15V200 psig *
1
.1
400KPAG
–15V200PSIG
1
.1
1
0.4MPAG
.0001 –30INHG/200PSIG
1
4BARG
.001 –30V400INHGG
.1
4KGCMG
.001 –400V5500INH2OG
.0001 4ATMG
.001 –240V3200ZING
100 psia
Res
–100V1400KPAG
.001
.001 100PSIA
–0.1V1.4MPAG
.1
.001 200INHGA
.1
–1V14BARG
Res
2770INH2OA
1
–1V14KGCMG
.01
1600ZINA
1
–1V 14ATMG
.01
7000GCMA
1
200 psig
1
5200MMHGA
1
200PSIG
5200TORRA
400INHGG
.1
1
7000MBARA
5500INH2OG
1
1
1
7000CMH2OA
1
3200ZING
1
700KPAA
.1
480FTH2O
1
0.7MPAA
.0001 1400KPAG
1
7BARA
.001 1.4MPAG
7KGCMA
.001 14BARG
.1
.0001 7ATMA
.001 14KGCMG
14ATMG
.001
–40 to 203°F (–40 to 95°C)
–4 to 185°F (–20 to 85°C)
(0 to 70°C)
Consult factory for special units
Res
300 psig
Res
300PSIG
.1
.1
610INHGG
.1
.1
4800ZING
.1
1
1
700FTH2O
.1
1
2000KPAG
1
2MPAG
.001
1
20BARG
.01
20KGCMG
1
.01
.001 20ATMG
.01
500 psig
Res
.01
.01
500PSIG
.1
.01
1020INHGG
1
Res
1150FTH2O
1
.1
3500KPAG
1
3.5MPAG
.1
.001
35BARG
1
.01
1
35KGCMG
.01
1
35ATMG
.01
1
1000 psig
Res
1
1000PSIG
1
2040INHGG
1
1
2300FTH2O
1
1
7000KPAG
.1
1
.1
7MPAG
.001
.0001 70BARG
.01
.001 70KGCMG
.01
.001 70ATMG
.01
2000 psig
Res
.001
Res
2000PSIG
1
4070INHGG
.1
1
.1
4600FTH2O
1
.1
14MPAG
.01
1
140BARG
.1
1
140KGCMG
.1
140ATMG
1
.1
3000 psig
Res
.001
.01
3000PSIG
1
.01
6100INHGG
1
.01
6900FTH2O
1
Res
20MPAG
.01
.1
200BARG
.1
200KGCMG
.1
.1
200ATMG
1
.1
5000 psig
Res
1
.1
5000PSIG
1
1
35MPAG
.01
.001 350BARG
.1
350KGCMG
.01
.1
340ATMG
.01
.1
.01
O ±0.1% Test Gauge Accuracy in Most Ranges
O Ultra-Ruggedized Design
O 316L Stainless Steel Wetted Parts
O Selectable Units and Auto Shutoff Times
O White LED Display Backlight
O Min/Max Memory
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
How to Specify
Included Features
Red rubber boot
White LED backlight
CTP1B range units
All metal case
Port reinforcement
Coated circuit boards
Range and Units—See table at left
Select a range code for default units
Please specify if vacuum gauge requires a minus sign
psi = PSI
G
gauge reference pressure
inHg = INHG
VAC gauge reference vacuum
oz/in
= ZIN
A
absolute reference
2
inH
O = INH2O
2
O = FTH2O
ftH
2
mmHg = MMHG
torr = TORR
mmH
O = MMH2O
2
= KGCM
kg/cm
2
g/cm
= GCM
2
kPa = KPA
MPa = MPA
mbar = MBAR
bar = BAR
cmH
O = CMH2O
2
atm = ATM
Options—add to end of model number. See price list for details.
CD
Calibration data; 5 test points and date
NC
NIST traceability documentation, 5 points and date
Accessories—order separately
DPG-OK2, DPG-OK3, DPG-OK6
Pelican
®
brand high visibility orange heavy duty
waterproof cases. Models available for storing 2,
3, or 6 gauges.
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. For food vacuum packaging applications.
303SS body, 100 micron 304SS screen.
CON14SS
Quick connector to install or remove gauge with-
out tools. 304 stainless steel, urethane seal.
CTP1B

Advertisement

loading

Summary of Contents for Cecomp CTP1B

  • Page 1 Red rubber boot 6INHGG .001 0.1MPAVAC .0001 2ATMA .001 –30INHG/100PSIG 610INHGG White LED backlight 85INH2OG 1BARVAC 30 psig –30V200INHGG 4800ZING .001 CTP1B range units All metal case 50ZING 1KGCMVAC .001 30PSIG –400V2770INH2OG 700FTH2O Port reinforcement 210GCMG 1ATMVAC .001 60INHGG –240V1600ZING 2000KPAG...
  • Page 2 Lo reading Press/hold MIN > min. reading & units Cecomp maintains a constant effort to upgrade and improve its products. indicated, the display is tested again, then the actual pressure Exit Hi/Lo Press/hold SSSS >...
  • Page 3 Instructions CTP1B Basic Configuration Advanced Configuration Advanced Configuration-continued Accessing the Internal Buttons Gauge Type Configuration Advanced configuration requires a passcode for access and allows more features to be configured. This will only appear with 15, 100, or 200 psig ranges that 1.
  • Page 4 Instructions CTP1B Calibration Preparation Calibration—continued User-Defined Passcode Configuration Gauges are calibrated at the factory using equipment traceable to release the front button when the appropriate engineering units are Remove the rear cover to access the buttons located near the lower NIST.