Data logging
Air
Sensor
Ref :472 089
2.5 Adjustment of 0.04 %
Note: It is possible that the displayed value is slightly different from 0.04%. In this case,
start the calibration procedure again.
2.6 Data acquisition
The sensor
purpose module. When you want to make an acquisition, slide the icon onto one of the
axes. For example, to acquire the
onto the ordinate and the time icon onto the abscissa.
To start acquisition, click on
3. Technical characteristics
Measurement
Measurement range
Precision
Heating time
Probe useful life
ENGLISH
-
Place the probe in open air.
-
Do a long press (a few seconds) on the button positioned on the sensor.
-
The LED flashes red for the time of the calibration.
-
Once the LED is continuously on in red, calibration is finished.
is represented by the
2
icon in the Atelier Scientifique general-
content as a function of time, slide the
2
.
CO
content in air
2
0-5%
at 5% of CO
: ± 0,03% of CO
2
at 0,2% of CO
: ± 0.01% of CO
2
at 0,1% of CO
: ± 0,006% of CO
2
30 seconds to 1 minute
>5 years
7
icon
2
2
2
2