Before use Dear User: Thanks for your trust. This operating manual is designed to familiarize you with the operation of our units and their possible applications, please read the operating manual carefully prior your use. And handle the device with care during transportation, unpack, testing, operation, maintains, storage, and scrape.
Warning Warning ! Danger of electric shock. Have all service and repair work carried out by authorized specialists only. Switch off the unit and pull the plug before carrying out any service or repair work. Warning ! Do not operate the device in explosive atmospheres, with hazardous substances. Warning! Unit parts may develop high surface temperatures.
Page 4
The device can only run without any malfunction or damage. As the operator of the device, any malfunction or damage found, please: Switch off the device. Pull the plug of the device. Contact us for trouble shooting or repairing. Note: Place the unit on an even, stable, clean, nonslip, dry and fireproof surface.
Unpack and check Unpack the device and check the device an accessories carefully。 Please FILL the warranty card if you found any damage of the device and accessories,notify the distributor and inform us immediately. Setting up Functional Module Control Panel Liquid level Pump Inlet...
Setting up Place the unit on an even, stable, clean, nonslip, dry and fireproof surface. Keep at least 10 cm of open space on the front and rear side. Do not set up the device close to heat sources and do not expose to sun light. Ensure the pump plug is security fixed without any external circulation。...
Operation Control panel: Item Button/Indicator Function Menu/Pump button press for 1 seconds,to turn on/off the circulation pump. press for 5 seconds, to entry setting mode (Menu). press Run / Stop button for start/stop circulation and temperature control. press Temperature setting (up) for turn up the setting temperature.
Page 8
Menu: Press button (A) for 5 seconds to enter the setting mode. To set the maximum temperature limitation (High Temperature —— Hi T): Press button (B),to jump to the next setting item. Press button (A),to exit the setting mode, and display the working temperature. Press button (C) or (D) to enter the max.
Page 9
Working mode (Mode): Press button (B),to jump to the next setting item. Press button (A),to exit the setting mode, and display the operation temperature, Press button (C) or (D) to enter the working mode setting interface. Press button (C) or button (D) to adjust the working mode setting. Press button (B) to save the ...
Page 10
Temperature calibration Reset(Calibration —— rCAL): Press button (B),to jump to the next setting item. Press button (A),to exit the setting mode, and display the operation temperature. Press button (C) or (D) to enter the temperature calibration mode setting interface. ...
Page 11
Examples: The actual measurement data of the third-party sensor is 37.00℃, and the value before CLP1 correction is 37.05℃, then the COP1 deviation value should be entered -0.05 (deviation value=actual measurement data - pre-correction value), press button (B), to save the deviation value. CLP1 Before correction ,...
Error codes Any error happen, the device will stop circulation and heating,beeper on and show the error message on the display. Proceed as followings in such cases: Turn off the device and pull the plug. Carry out corrective measures. restart the device.
Accessories: PVC tube (DN 8) Temperature range:-20…60° C PVC tube (DN 12) Temperature range:-20…60° C NBR tube (DN 8) Temperature range:-30…80° C NBR tube (DN 12) Temperature range:-30…80° C Silicone tube (DN 8) Temperature range:-40…180° C Silicone tube (DN 12) Temperature range:-40…180°...