Page 3
ASD 220 Slide Dryer Service Manual 1、 General Provisions The Service Manual of the instrument is compiled according to its structure and characteristic. Configuration 2、 Power supply of the instrument is AC110V/220V. Slide Drier can heat up and control temperature.
Page 4
C、There is continuous ring in buzzer(the interval is 10 seconds) It means that the digital sensor is broken, and need to replace the digital sensor. ASD 220 Slide Dryer Service Manual ---2---...
Page 5
Maintain:use the Philip screwdriver to loosen the six screws in the base-board, take down the base-board and you will see the solid state relay as it showed in the picture. Take down the damaged SSR and replace a new SSR. ASD 220 Slide Dryer Service Manual ---3---...
Need help?
Do you have a question about the ASD 220 and is the answer not in the manual?
Questions and answers