CAZON B26 Instructions Manual

Error messages cause and solution
Hide thumbs Also See for B26:

Advertisement

Quick Links

CAZON
UPPER ARM TYPE
BLOOD PRESSURE MONITOR
Error Messages
Model:
B26
Cause And Solution

Advertisement

loading
Need help?

Need help?

Do you have a question about the B26 and is the answer not in the manual?

Questions and answers

Merci Holcomb
March 9, 2025

Is there a way to calibrate it? Doesn’t seem to have correct reading when compared to the one used by the hospital.

1 comments:
Mr. Anderson
March 9, 2025

According to the provided information, the CAZON B26 blood pressure monitor should be checked for accuracy every year according to international standards. However, the document does not mention a specific calibration procedure. For accurate readings, ensure the cuff is properly fitted, avoid consecutive measurements without a 5-minute pause, and follow the correct measurement procedure. If calibration is required, contact the manufacturer or distributor.

This answer is automatically generated

Carol Samain
February 18, 2025

I have used the item for a couple of weeks now the machine inflates and then switches off

1 comments:
Mr. Anderson
February 18, 2025

The CAZON B26 machine may inflate and then switch off after a couple of weeks of use due to discharged batteries. When the batteries are low, the battery symbol will flash when the device is switched on, and no further measurements can be taken until the batteries are replaced. Replacing the batteries with four AAA long-life or alkaline 1.5V batteries should resolve the issue.

This answer is automatically generated

didier durand
February 6, 2025

My CAZON is not displaying properly . I need help. It ishas only beenused once!

1 comments:
Mr. Anderson
February 10, 2025

Your CAZON B26 may not be displaying properly due to weak batteries or improper battery placement. Replace both batteries with new ones and check that they are installed correctly with the proper polarities.

This answer is automatically generated

Robert Alan Clift
June 29, 2025

What is the Lo message

Summary of Contents for CAZON B26

  • Page 1 CAZON Error Messages Model: UPPER ARM TYPE BLOOD PRESSURE MONITOR Cause And Solution...
  • Page 3: Error Display

    Error Display No display appears Cause Weak battery or improper placement Solution Replace both batteries with new ones. Check the battery installation for proper placement of the battery polarities.
  • Page 4 Error Display Err 2 Cause E r r Monitor could not detect pulse wave or cannot calculate the blood pressure data. Solution Check if the air releasing is too slow or not. If it is too slow, please check if there is any dust in the tube plug and the cuff port.
  • Page 5 Error Dis play Err 4 E r r Cause Too loose cuff or air leakage. (Cannot inflate to 30mmHg within 15s) Solution Tie the cuff correctly and make sure the air plug is properly inserted in the device.
  • Page 6 Error Display Err 5 E r r Cause The air tube is crimped. Solution Correct it and make the measuring againg.
  • Page 7 Error Display Err 6 E r r Cause The sensor is sensing great fluctuation in the pressure. Solution Please keep quiet and don't move.
  • Page 8 Error Display Err 7 E r r Cause The pressure that the sensor sensing is over the limit. Solution Please send back to the distributor.
  • Page 9 Error Display Err 8 E r r Cause The demarcation is incorrect or the device has not been demarcated. Solution Please send back to the distributor.