Do you have a question about the B26 and is the answer not in the manual?
Questions and answers
Merci Holcomb
March 9, 2025
Is there a way to calibrate it? Doesn’t seem to have correct reading when compared to the one used by the hospital.
1 comments:
Mr. Anderson
March 9, 2025
According to the provided information, the CAZON B26 blood pressure monitor should be checked for accuracy every year according to international standards. However, the document does not mention a specific calibration procedure. For accurate readings, ensure the cuff is properly fitted, avoid consecutive measurements without a 5-minute pause, and follow the correct measurement procedure. If calibration is required, contact the manufacturer or distributor.
This answer is automatically generated
Carol Samain
February 18, 2025
I have used the item for a couple of weeks now the machine inflates and then switches off
1 comments:
Mr. Anderson
February 18, 2025
The CAZON B26 machine may inflate and then switch off after a couple of weeks of use due to discharged batteries. When the batteries are low, the battery symbol will flash when the device is switched on, and no further measurements can be taken until the batteries are replaced. Replacing the batteries with four AAA long-life or alkaline 1.5V batteries should resolve the issue.
This answer is automatically generated
didier durand
February 6, 2025
My CAZON is not displaying properly . I need help. It ishas only beenused once!
1 comments:
Mr. Anderson
February 10, 2025
Your CAZON B26 may not be displaying properly due to weak batteries or improper battery placement. Replace both batteries with new ones and check that they are installed correctly with the proper polarities.
Error Display No display appears Cause Weak battery or improper placement Solution Replace both batteries with new ones. Check the battery installation for proper placement of the battery polarities.
Page 4
Error Display Err 2 Cause E r r Monitor could not detect pulse wave or cannot calculate the blood pressure data. Solution Check if the air releasing is too slow or not. If it is too slow, please check if there is any dust in the tube plug and the cuff port.
Page 5
Error Dis play Err 4 E r r Cause Too loose cuff or air leakage. (Cannot inflate to 30mmHg within 15s) Solution Tie the cuff correctly and make sure the air plug is properly inserted in the device.
Page 6
Error Display Err 5 E r r Cause The air tube is crimped. Solution Correct it and make the measuring againg.
Page 7
Error Display Err 6 E r r Cause The sensor is sensing great fluctuation in the pressure. Solution Please keep quiet and don't move.
Page 8
Error Display Err 7 E r r Cause The pressure that the sensor sensing is over the limit. Solution Please send back to the distributor.
Page 9
Error Display Err 8 E r r Cause The demarcation is incorrect or the device has not been demarcated. Solution Please send back to the distributor.
Need help?
Do you have a question about the B26 and is the answer not in the manual?
Questions and answers
Is there a way to calibrate it? Doesn’t seem to have correct reading when compared to the one used by the hospital.
According to the provided information, the CAZON B26 blood pressure monitor should be checked for accuracy every year according to international standards. However, the document does not mention a specific calibration procedure. For accurate readings, ensure the cuff is properly fitted, avoid consecutive measurements without a 5-minute pause, and follow the correct measurement procedure. If calibration is required, contact the manufacturer or distributor.
This answer is automatically generated
I have used the item for a couple of weeks now the machine inflates and then switches off
The CAZON B26 machine may inflate and then switch off after a couple of weeks of use due to discharged batteries. When the batteries are low, the battery symbol will flash when the device is switched on, and no further measurements can be taken until the batteries are replaced. Replacing the batteries with four AAA long-life or alkaline 1.5V batteries should resolve the issue.
This answer is automatically generated
My CAZON is not displaying properly . I need help. It ishas only beenused once!
Your CAZON B26 may not be displaying properly due to weak batteries or improper battery placement. Replace both batteries with new ones and check that they are installed correctly with the proper polarities.
This answer is automatically generated
What is the Lo message