After the switch is pressed, the display will instruct you to place the cal stand to the -45
degree, than to -30 degrees etc. until +60 degrees is completed. When the cal stand is placed
in the +60 degree position and the CALIBRATE switch is pressed, the program will go back to
RUN mode and the calibration procedure for that tool is finished. The accuracy of the
calibration can be checked in the RUN mode by placing the cal stand into various positions
and observing the display values. If only one tool is used, the calibration procedure is
finished. If two tools are being used, and the other tool needs calibration, repeat steps A
through D for the second tool.
* SETTING TOOL DISPLAY OFFSET*
A) Install the desired Clear Vision steering wheel set tool onto a calibration stand which can
be adjusted from +60 degrees to -60 degrees in 15 degree increments.
B) Press the CALIBRATION pushbutton switch once and the display will read.
C) DO NOT press CALIBRATION pushbutton again. The display will read.
This display shows the user what the current tool display offset value.
Press the CALIBRATION pushbutton to change the offset value. The display will read.
D) Press the CALIBRATION pushbutton to cycle from +5.00 to –5.00 degrees in 0.50 degree
increments. Stop pressing the CALIBRATION pushbutton to set value.
96-036.doc
47650 Clipper Dr. Plymouth Twp., MICHIGAN 48170 (734)-414-6500
MODEL KEI-850A READOUT BOX
CAL sw again=CAL
Timeout = offset
CAL SWITCH = OFFSET
X.XX deg offset
X.XX deg. OFFSET
TIMEOUT WILL SET
4