Summary of Contents for ELECTROTEK AW800TGRF4P STANDARD TWO GAS

  • Page 1 ANAEROBIC WORKSTATION AW800TGRF4P STANDARD TWO GAS VERSION MANUAL PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Burnt Mill, Elizabeth Way Harlow, Essex, CM20 2HU UK 44-45 met@munro-group.co.uk MUNRO.6.22...
  • Page 2 CONTENTS Page GENERAL ARRANGEMENT DRAWING OF ANAEROBIC WORKSTATION GENERAL SPECIFICATION GAS SUPPLIES ELECTRICAL SUPPLE$ INSTALLATION OF GAS AND ELECTRIC DRAWING ASSEMBLY INSTRUCTIONS NO. 1. DRAWING ASSEMBLY INSTRUCTIONS NO. 2. DRAWING SPECIAL FEATURES CHECKING THE INCUBATOR ATMOSPHERE FOR OXYGEN PREPARING THE ANAEROBIC INDICATOR SOLUTION COMMISSIONING THE CABINET CHECKING THE GAS FLOW TRANSFERRING ITEMS INTO THE INCUBATOR...
  • Page 3 3• • ANGLE PROJECTION In cubator operating temperature: ° ° C above ambient to 45 I nternal IE Electrical Mains Sockets fitted for powering • Laborato ry E quipment Two P a ir s of Arm Ports using MUNRO Rapid Arm En try a n d Remova l System.
  • Page 17 GENERAL MAINTENANCE The MUNRO Workstation, unlike other units, uses the energy contained in the gas bottles to do most of the work required. This means that there are no evacuation or circulation pumps to maintain. Many of the components used in the construction of the unit are available "off-the-shelf' in almost every part of the world.
  • Page 19 TEST RECORD SHEET MODEL AW800TGRF4P STD SERIAL NO CUSTOMER FLUSH CYCLE 4 SECONDS ON 2 SECONDS OFF FLUSH DURATION N2 - 3 MINUTES, MG- 2 MINUTES, TOTAL 5 MINS MIXED GAS FALLING CUT OUT PRESSURE 20_PSI MIXED GAS RISING SWITCH ON PRESSURE 30_PSI GAS FALLING CUT OUT PRESSURE 20_PSI...
  • Page 27 "Sub setting mode "SEL At Set bASE, press A to move to the second mode SEt SEL SELi =NoNE SEL2 =NoNE SEL3 =NoNE SET4 = NoNE SELS =NoNE SEL6 =NoNE SEL7 =NoNE SEL8 = NoNE (Reverts to PV/SV display by pressing on R key) "Suh setting mode "CoMM At Set bASE, press A to move to the second mode Set Co MM (Reverts to PV/SV display by pressing on R key)